• Title/Summary/Keyword: 축/측 방향 해상도

Search Result 2, Processing Time 0.018 seconds

Performance Evaluation of Medical Ultrasonic Equipment in Gwangju Using Multipurpose Ultrasonic Phantom (다목적 초음파 팬텀을 이용한 광주 지역 의료용 초음파장비 성능평가)

  • Jong-Gil Kwak;Cheol-Min Jeon;Joo-Ah Lee
    • Journal of the Korean Society of Radiology
    • /
    • v.18 no.5
    • /
    • pp.473-481
    • /
    • 2024
  • As a result of measuring and evaluating 24 units of 10 types of medical ultrasonic equipment in Gwangju using the multipurpose ultrasonic phantom (ATS-539), all 24 units were 100% passed in the non-inspection area, vertical distance measurement, horizontal distance measurement, axial/side resolution, gray tone and dynamic range items. As a result of analyzing the ultrasonic phantom pass rate according to the number of years of use of ultrasonic equipment in Gwangju, this study derived that the service life of ultrasonic equipment is related to the equipment performance. As the coverage for ultrasonic inspection is gradually expanded starting with the upper abdominal ultrasound, it is time to further manage the quality of ultrasonic inspection. Therefore, in addition to the standard guidelines for performance measurement for abdominal inspection, research on equipment phantom measurement methods and standards that can be used in various ultrasonic inspection areas such as the neck and chest, which have recently become increasingly important in clinical practice, should be conducted at the same time. In addition, quality control standards should be legally established to maintain the optimal performance of ultrasonic imaging devices and to secure the appropriateness of images. In addition, ultrasound imaging devices should be classified as special medical equipment, and it is considered that the level of equipment should be consistently maintained through continuous management so that images beneficial to treatment can be obtained.

Performance Testing of Medical US Equipment Using US Phantom(ATS-539) (Focusing on Daegu Region) (초음파팬텀(ATS-539)을 이용한 의료용 초음파장비의 성능검사(대구지역을 중심으로))

  • Kim, Do-Hyung;Kwon, Deok-Moon
    • Journal of radiological science and technology
    • /
    • v.37 no.4
    • /
    • pp.295-305
    • /
    • 2014
  • This study is to provide accurate information as medical imaging equipment to check for the presence of body disease US equipment. We investigated the status of medical US equipment performance in Daegu and criteria US phantom (ATS-539) for US equipment performance measurements. The results in this study, 1. US phantom measurement results: The test passed rate were 88.6% and the failed rate were 11.4%. 2. The difference between the group of mean and the pass/failed groups were statistically significant. Focal zone and 4 mm functional resolution in the two items that are not present the passing standard. 3. The difference was statistically significant number of years and used equipment and pass the failed equipment (4.13 vs 7.25 years). We investigated the performance status of US equipment used in the clinical area in Daegu. The basis for the two items are not present this proposed passing standard. Equipment performance was associated with the number of years of using US equipment. It is necessary to maintain the best performance of the equipment phantom measurements for performance testing of US equipment.