• Title/Summary/Keyword: 전자선 측정

Search Result 952, Processing Time 0.038 seconds

The Calculation of Energy Distributions for Clinical Electron Beams from Mono Energetic Depth dose Data (단일에너지 깊이선량률 자료에 의한 치료용 전자선의 에너지분포 계산)

  • 이정옥;정동혁
    • Progress in Medical Physics
    • /
    • v.15 no.1
    • /
    • pp.39-44
    • /
    • 2004
  • The energy distributions for clinically used electron beams from measured and calculated mono energetic depth dose values were calculated. The energy distributions having the minimum difference between the measured and reduced values of depth dose are determined by iterations based on least square method. The nominal energies of 6, 9, 12, 15 MeV clinical electron beams were examined. The Monte Carlo depth dose calculations with determined energy distributions were peformed to evaluate those distributions. In a comparison of the calculated and measured depth dose data, the standard errors are estimated within $\pm$ 3% from surface to R$_{80}$ depth and within $\pm$4% from the surface to near the range for all electron beams. This can be practically applied to determine the energy distributions for clinically used electron beams.

  • PDF

Dose distribution at junctional area abutting X-ray and electron fields (X-선과 전자선의 인접조사에서 접합 조사면에서의 선량분포)

  • Yang, Kwang-Mo
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.16 no.1
    • /
    • pp.91-99
    • /
    • 2004
  • Purpose : For the head and neck radiotherapy, abutting photon field with electron field is frequently used for the irradiation of posterior neck when tolerable dose on spinal cord has been reached. Materials and methods : Using 6 MV X-ray and 9 MeV electron beams of Clinac1800(Varian, USA) linear accelerator, we performed film dosimetry by the X-OMAT V film of Kodak in solid water phantom according to depths(0 cm, 1.5 cm, 3 cm, 5 cm). 6 MV X-ray and 9 MeV electron(1Gy) were exposes to 8cm depth and surface(SSD 100cm) of phantom. The dose distribution to the junction line between photon($10cm{\times}10cm$ field with block) and electron($15cm{\times}15cm$ field with block) fields was also measured according to depths(0 cm, 0.5 1.5 cm, 3 cm, 5 cm). Results : At the junction line between photon and electron fields, the hot spot was developed on the side of the photon field and a cold spot was developed on that of the electron field. The hot spot in the photon side was developed at depth 1.5 cm with 7 mm width. The maximum dose of hot spot was increased to $6\%$ of reference doses in the photon field. The cold spot in the electron side was developed at all measured depths(0.5 cm-3 cm) with 1-12.5 mm widths. The decreased dose in the cold spot was $4.5-30\%$ of reference dose in the electron field. Conclusion : When we make use of abutting photon field with electron field for the treatment of head and neck cancer we should consider the hot and cold dose area in the junction of photon and electron field according to location of tumor.

  • PDF

High Energy Electron Dosimetry by Alanine/ESR Spectroscopy (Alanine/ESR Spectroscopy에 의한 고에너지 전자선의 선량측정)

  • Chu, Sung-Sil
    • Radiation Oncology Journal
    • /
    • v.7 no.1
    • /
    • pp.85-92
    • /
    • 1989
  • Dosimerty based on electron spin resonance (ESR) analysis of radiation induced free radicals in amino acids is relevant to biological dosimetry applications. Alanine detectors are without walls and are tissue equivalent. Therefore, alanine ESR dosimetry looks promising for use in the therapy level. The dose range of the alanine/ESR dosimetry system can be extended down to 1 Gy. In water phantom the absorbed dose of electrons generated by a medical linear accelerator of different initial energies $(6\~21MeV)$ and therapeutic dose levels (1~60 Gy) was measured. Furthermore, depth dose measurements carried out with alanine dosimeters were compared with ionization chamber measurements. As the results, the measured absorbed doses for shallow depth of initial electron energies above 15 MeV were higher by$2\~5\%$ than those calculated by nominal energy $C_E$ factors. This seems to be caused by low energy scattered beams generated from the scattering foil and electron cones of beam projecting device in medical linear accelerator.

  • PDF

Comparison of Electron Beam Dosimetries by Means of Several Kinds of Dosimeters (수종의 측정기에 의한 전자선의 선량 측정의 비교)

  • Kang Wee-Saing
    • Radiation Oncology Journal
    • /
    • v.7 no.1
    • /
    • pp.93-100
    • /
    • 1989
  • Several combinations of measuring devices and phantoms were studied to measure electron beams. Silicon Pmt junction diode was used to find the dependence of depth dose profile on field size on axis of electron beam Depths of 50, 80 and $90\%$ doses increased with the field size for small fields. For some larger fields, they were nearly constant. The smallest of field sizes over which the parameters were constant was enlarged with increase of the energy of electron beams. Depth dose distributions on axis of electron beam of $10\times10cm^2$ field were studied with several combinations of measuring devices and phantoms. Cylindrical ion chamber could not be used for measurement of surface dose, and was not convenient for measurement of near surface region of 6MeV electron. With some exceptions, parameters agreed well with those studied by different devices and phantoms. Surface dose in some energies showed $4\%$ difference between maximum and minimum. For 18MeV, depths of 80 and $90\%$ doses were considerably shallower by film than by others. Parallel-plate ion chamber with polystyrene phamtom and silicon PN junction would be recommended for measurement of central axis depth dose of electron beams with considerably large field size. It is desirable not to use cylindrical ion chamber for the purpose of measurement of surface dose or near surface region for lower energy electron beam. It is questionable that film would be recommended for measurement of dose distribution of electron with high energy like as 18MeV.

  • PDF

평행평판형이온함을 이용한 전자선 흡수선량결정에 대한 연구

  • Ra, Jeong-Eun;Seo, Tae-Seok;Sin, Dong-O
    • Proceedings of the Korean Society of Medical Physics Conference
    • /
    • 2004.11a
    • /
    • pp.149-152
    • /
    • 2004
  • 물 흡수선량 표준에 토대를 두고 있는 프로토콜에서는 저에너지 전자선의 경우 평행평판형이온함의 사용과 기준 선질 $^{60}$CO 감마선의 물 흡수선량 교정정수를 받은 원통형이온함을 사용하여 고에너지 전자선에서 평행평판형이온함을 교차교정하도록 권고하고 있다. 따라서 본 연구에서는 국제원자력기구의 프로토콜(IAEA TRS-398)에서 권고하고 있는 절차에 따라 저에너지 전자선에 대한 원통형이온함의 선질보정정수를 계산하고, 원통형이온함과 평행평판형이온함의 교정방법에 따른 흡수선량을 상호 비교하였다. 그 결과 전자선에너지 10 MeV 이상에서는 두 이온함간의 선량이 잘 일치하였으나 전자선에너지 6, 9 MeV에서 최대 3.3%까지 선량 차이를 보여 저에너지 전자선에서는 반드시 평형판판형이온함의 사용하여 선량측정 할 것을 권고한다. 교정방법 차이에 의한 평행평판형이온함의 선량은 서로 잘 일치하는 것으로 나타나 표준기관에서 직접 교정받은 $^{60}$Co 감마선의 물 흡수선량교정정수를 사용하여 전자선 물 흡수선량을 결정해도 큰 영향은 없을 듯하다. 또한 평행평판형이온함을 교차 교정하기 위한 전자선 에너지에 따른 흡수선량을 상호 비교한 결과 20MeV이외 12, 16 MeV의 전자선 에너지에서도 잘 일치하여 교차교정을 위한 전자선의 기준 선질에 대한 연구가 더 진행되어야 한다고 사료된다.

  • PDF

Fabrication of Backscatter Electron Cones for Radiation Therapy (산란전자선을 이용한 강내측방조사기구의 제작과 특성)

  • Chu, Sung-Sil;Suh, Chang-Ok;Kim, Gwi-Eon
    • Radiation Oncology Journal
    • /
    • v.19 no.1
    • /
    • pp.74-80
    • /
    • 2001
  • Purpose : Irradiation cones by using backscatter electrons are made for the treatment of superficial small lesions of skin, oral cavity, and rectum where a significant dose gradient and maximum surface dose is desired. Methods and Materials : Backscatter electrons are produced from the primary electron beams from the linear accelerators. The design consists of a cylindrical cone that has a thick circular plate of high atomic number medium (Pb or Cu) attached to the distal end, and the plate can be adjusted the reflected angle. Primary electrons strike the metal plate perpendicularly and produce backscatter electrons that reflect through the lateral hole for treatment. Using film and a parallel plate ion chamber, backscatter electron dose characteristics are measured. Results : The depth dose characteristic of the backscatter electron is very similar to that of the hard x-ray beam that is commonly used for the intracavitary and superficial lesions. The basckscatter electron energy is nearly constant and effectively about 1.5 MeV from the clinical megavoltage beams. The backscatter electron dose rate of $35\~85\;cGy/min$ could be achieved from modern accelerators without any modification. and the depth in water of $50\%$ depth dose from backscatter electron located at 6mm for $45^{\circ}$ angled lead scatter. The beam flatness is dependent on the slit size and the depth of treatment, but is satisfactory to treat small lesions. Conclusions : The measured data for backscatter electron energy, depth dose flatness dose rate and absolute dose indicates that the backscatter electrons are suitable for clinical use.

  • PDF

Effect of UV-C and Electron Beam Irradiation of on the Quality of Rice Wine (Makgeolli) (UV-C 및 전자선 조사가 막걸리의 품질에 미치는 영향)

  • Kim, Hyun-Joo;Lee, Kyung Haeng;Yong, Hae In;Jo, Cheorun
    • Food Science and Preservation
    • /
    • v.20 no.1
    • /
    • pp.45-51
    • /
    • 2013
  • This study was conducted to evaluate the changes in the quality of rice wine (Makgeolli) treated with UV-C and electron beam (EB) irradiation during its storage at $4^{\circ}C$ for 15 days. The EB irradiation was found to be more effective than the UV-C irradiation for microbiological control. The pH tended to be increased by UV-C, EB irradiation, and storage period. Acidity was significantly higher in rice wine treated with EB irradiation than those of control and UV-C irradiation during storage period. The $L^*$-value of the samples treated with UV-C and EB irradiation tended to be proportionately higher than that of the control, but this was not consistent during storage period. The $a^*$-value of the sample treated with EB was higher than that of the others, but this was not consistent during the storage period. The reducing sugar content was higher in the control at day 0, but it rapidly decreased compared with the control during the storage. The results indicate that EB irradiation was more effective than UV-C irradiation in controlling microorganisms of rice wine. However, a further study is needed to minimize the rice wine quality deterioration caused by UV-C or EB irradiation during storage.

A Study of Polarity Effect of Parallel Plate Type ionization Chamber with Different Volume (평행평판형 전리함의 두 전극간의 거리에 따른 극성효과 연구)

  • 윤형근;신교철
    • Progress in Medical Physics
    • /
    • v.13 no.2
    • /
    • pp.69-73
    • /
    • 2002
  • Exposure measurement data with parallel plate ionization chambers were known to depend on the polarity ($k_{pol}$) effect. In this work, the polarity effect were investigated for three parallel plate ionization chambers with different volume. The ionization chamber was fabricated using acrylic plate for the chamber medium and printed circuit board for electrical configuration. The various sizes of the sensitive volumes designed so far were 0.9, 1.9, and 3.1 co. High voltage generator was fabricated using the conventional 9 V batteries to apply the high voltage (300-500 V) to the electrode of the parallel plate ionization chamber. The gap between two electrodes ranged from 3, 6, and 10mm. As the result of our experiment, the polarity effect was within 0.5% in photon beam and 1% to 3.5% in the electron beams. Among electron beams, 16 MeV beam, which had highest energy, showed less polarity effect than electron beams with other energies.

  • PDF

Evaluation of Lead Oxide Dosimeter for Quality Assurance of Electron Beam in Radiotherapy (방사선치료 전자선의 정도관리를 위한 Lead Oxide 선량계 평가)

  • Yang, Seungwoo;Han, Moojae;Park, Sungkwang
    • Journal of the Korean Society of Radiology
    • /
    • v.15 no.1
    • /
    • pp.79-83
    • /
    • 2021
  • In radiation therapy, electron beam is often used in the treatment of superficial lesion. Accurate measurements are required because electron beam interacts with them in the beam path and affects dose measurements. However, no research has been conducted on electron beam quality assurance. in this study, PbO-based dosimeter was fabricated as a basic study for electron beam quality assurance. Thus, the reproducibility and linearity of the energy of 6, 9, and 12 MeV were analyzed to evaluate measurement accuracy and precision. Reproducibility measurements show RSD value of 1.024%, 1.019% and 0.890%, respectively, at 6, 9, and 12 MeV. linearity measurements show 0.9999 R2 at 6, 9, and 12 MeV altogether. Both evaluations show that the PbO dosimeter has very good measurement accuracy and precision with excellent results.

Dosimetry by Using EBT2 Film for Total Skin Electron Beam Therapy (TSET) (전신 피부 전자선 치료(TSET)에서 EBT2 필름을 사용한 선량측정)

  • Hwang, Ui-Jung;Rah, Jeong-Eun;Jeong, Ho-Jin;Ahn, Sung-Hwan;Kim, Dong-Wook;Lee, Sang-Yeob;Lim, Young-Gyung;Yoon, Myong-Geun;Shin, Dong-Ho;Lee, Se-Byeong;Park, Sung-Young;Pyo, Hong-Ryull;Chung, Weon-Kuu
    • Progress in Medical Physics
    • /
    • v.21 no.1
    • /
    • pp.60-69
    • /
    • 2010
  • For treatment of Total Skin Electron beam Therapy (TSET), measurement of dose at various conditions is need on the contrary to usual radiotherapy. When treating TSET with modified Stanford technique based on linear accelerator, the energy of treatment electron beam, the spatial dose distribution and the actual doses deposited on the surface of the patient were measured by using EBT2. The measured energy of the electron beam was agreed with the value that measured by ionization chamber, and the spatial dose distribution at the patient position and the doses at several point on the patient's skin could be easily measured by EBT2 film. The dose on the patient that was measured by EBT2 film showed good agreement with the data measured simultaneously by TLD. With the results of this study, it was proven that the EBT2 film can be one of the useful dosimeter for TSET.