• Title/Summary/Keyword: 움직임인공물

Search Result 1, Processing Time 0.012 seconds

Quantitative Comparison of Motion Artifacts in PET Images using Data-Based Gating (데이터 기반 게이팅을 이용한 PET 영상의 움직임 인공물의 정량적 비교)

  • Jin Young, Kim;Gye Hwan, Jin
    • Journal of the Korean Society of Radiology
    • /
    • v.17 no.1
    • /
    • pp.91-98
    • /
    • 2023
  • PET is used effectively for biochemical or pathological phenomena, disease diagnosis, prognosis determination after treatment, and treatment planning because it can quantify physiological indicators in the human body by imaging the distribution of various biochemical substances. However, since respiratory motion artifacts may occur due to the movement of the diaphragm due to breathing, we would like to evaluate the practical effect by using the a device-less data-driven gated (DDG) technique called MotionFree with the phase-based gating correction method called Q.static scan mode. In this study, images of changes in moving distance (0 cm, 1 cm, 2 cm, 3 cm) are acquired using a breathing-simulated moving phantom. The diameters of the six spheres in the phantom are 10 mm, 13 mm, 17 mm, 22 mm, 28 mm, and 37 mm, respectively. According to maximum standardized uptake value (SUVmax) measurements, when DDG was applied based on the moving distance, the average SUVmax of the correction effect by the moving distance was improved by 1.92, 2.48, 3.23 and 3.00, respectively. When DDG was applied based on the diameter of the phantom spheres, the average SUVmax of the correction effect by the moving distance was improved by 2.37, 2.02, 1.44, 1.20, 0.42 and 0.52 respectively.