• 제목/요약/키워드: 엑스/감마서베이미터

검색결과 1건 처리시간 0.011초

군사용 방사선 계측기 PDR-1K 반응도 평가 : 방향 및 에너지 의존성을 중심으로 (Evaluation for Performance of a Military Radiation Detecter PDR-1K : Focused on Dependence of Angular and Energy)

  • 박원석;최준혁;정도영;김장오;민병인
    • 한국군사과학기술학회지
    • /
    • 제20권2호
    • /
    • pp.300-305
    • /
    • 2017
  • In this paper, we performed a evaluation for angular and energy dependence of military radiation detector PDR-1K. Its measuring range is divided into two section, low and high, and each range has a GM tube separately owing to broad scale. We observed a change in relative angular reactivity within 0.928 ~ 1.188 in low range and within 0.743 ~ 1.000 in high range from $-90^{\circ}$ to $+90^{\circ}$. The evaluation for energy dependence appeared a relative reactivity within 0.892 ~ 1.000 above 83 keV. This means PDR-1K isn't possible to use below 83 keV of radiation energy. It is possible to provide response information to user and to increase reliability of radiation measurement through this paper.