• Title/Summary/Keyword: 손목관절 CT

Search Result 1, Processing Time 0.014 seconds

MTF Evaluation according to change in posture and channel during CT examination for wrist Joint : X-axis and Z-axis changes around Isocenter (손목관절 CT 검사 시 자세 변화와 채널 변경에 따른 MTF 평가 : Isocenter를 중심으로 X-축, Z-축 변화)

  • Seo, Min Jae;Lim, Jong Chon;Jung, Dabin;Han, Dong Kyoon
    • Journal of the Korean Society of Radiology
    • /
    • v.14 no.6
    • /
    • pp.811-817
    • /
    • 2020
  • This study aims to evaluate the Modulation Transfer Function (MTF) according to the change in the number of channels of the CT examination device by changing the posture of the patient to the X-axis and Y-axis in the wrist joint CT examination. Using a CT device and a wrist phantom, the test was performed by moving 0 (matched), 5, 10, and 15 cm in the X-axis around the isocenter, and the Z-axis was rotated by -20° and -40°. For the test, 16, -40 and 64 channels were used to check whether there was a difference for each number of channels. The examined images were compared by measuring the MTF values of the ulna and left and right sides of the radius. In the experiment where the isocenter was moved along the X-axis, the MTF value decreased with an increase in the moving distance, and the MTF value was found to be unaffected by the number of channels. In the experiment in which the wrist joint was rotated by -20° and -40° on the Z-axis, the degree of deviation and MTF were found to be irrelevant. It was not related to the number of channels either. In conclusion, the movement of the wrist along the X-axis should be restrained as much as possible for a wrist joint CT scan, whereas deviation around the Z-axis depending on the environment for the patient would not affect the MTF of the image.