• Title/Summary/Keyword: 배후

Search Result 681, Processing Time 0.016 seconds

Evaluation of Image Quality Based on Time of Flight in PET/CT (PET/CT에서 재구성 프로그램의 성능 평가)

  • Lim, Jung Jin;Yoon, Seok Hwan;Kim, Jong Pil;Nam Koong, Sik;Shin, Seong Hwa;Yoon, Sang Hyeok;Kim, Yeong Seok;Lee, Hyeong Jin;Lee, Hong Jae;Kim, Jin Eui;Woo, Jae Ryong
    • The Korean Journal of Nuclear Medicine Technology
    • /
    • v.16 no.2
    • /
    • pp.110-114
    • /
    • 2012
  • Purpose : PET/CT is widely used for early checking up of cancer and following up of pre and post operation. Image reconstruction method is advanced with mechanical function. We want to evaluate image quality of each reconstruction program based on time of flight (TOF). Materials and Methods : After acquiring phantom images during 2 minutes with Gemini TF (Philips, USA), Biograph mCT (Siemens, USA) and Discovery 690 (GE, USA), we reconstructed image applied to Astonish TF (Philips, USA), ultraHD PET (Siemens, USA), Sharp IR (GE, USA) and not applied. inside of Flangeless Esser PET phantom (Data Spectrum corp., USA) was filled with $^{18}F$-FDG 1.11 kBq/ml (30 Ci/ml) and 4 hot inserts (8. 12. 16. 25 mm) were filled with 8.88 kBq/ml (240 ${\mu}Ci/ml$) the ratio of background activity and hot inserts activity was 1 : 8. Inside of triple line phantom (Data Spectrum corp., USA) was filled with $^{18}F$-FDG 37 MBq/ml (1 mCi). Three of lines were filled with 0.37 MBq (100 ${\mu}Ci$). Contrast ratio and background variability were acquired from reconstruction image used Flangeless Esser PET phantom and resolution was acquired from reconstruction image used triple line phantom. Results : The contrast ratio of image which was not applied to Astonish TF was 8.69, 12.28, 19.31, 25.80% in phantom lid of which size was 8, 12, 16, 25 mm and it which was applied to Astonish TF was 6.24, 13.24, 19.55, 27.60%. It which was not applied to ultraHD PET was 4.94, 12.68, 22.09, 30.14%, it which was applied to ultraHD PET was 4.76, 13.23, 23.72, 31.65%. It which was not applied to SharpIR was 13.18, 17.44, 28.76, 34.67%, it which was applied to SharpIR was 13.15, 18.32, 30.33, 35.73%. The background variability of image which was not applied to Astonish TF was 5.51, 5.42, 7.13, 6.28%. it which was applied to Astonish TF was 7.81, 7.94, 6.40 6.28%. It which was not applied to ultraHD PET was 6.46, 6.63, 5.33, 5.21%, it which was applied to ultraHD PET was 6.08, 6.08, 4.45, 4.58%. It which was not applied to SharpIR was 5.93, 4.82, 4.45, 5.09%, it which was applied to SharpIR was 4.80, 3.92, 3.63, 4.50%. The resolution of phantom line of which location was upper, center, right, which was not applied to Astonish TF was 10.77, 11.54, 9.34 mm it which was applied to Astonish TF was 9.54, 8.90, 8.88 mm. It which was not applied to ultraHD PET was 7.84, 6.95, 8.32 mm, it which was applied to ultraHD PET was 7.51, 6.66, 8.27 mm. It which was not applied to SharpIR was 9.35, 8.69, 8.99, it which was applied to SharpIR was 9.88, 9.18, 9.00 mm. Conclusion : Image quality was advanced generally while reconstruction program which is based on time of flight was used. Futhermore difference of result compared each manufacture reconstruction program showed up, however this is caused by specification of instrument of each manufacture and difference of reconstruction algorithm. Therefore we need further examination to find out appropriate reconstruction condition while using reconstruction program used for advance of image quality.

  • PDF