• Title/Summary/Keyword: 물흡수선량 교정정수

Search Result 1, Processing Time 0.015 seconds

Evaluation of the Long-Term Stability for the Cylindrical Ionization Chambers (교정정수 변화에 의한 원통형이온함의 안정성 평가)

  • Rah Jeong-Eun;Hong Ju-Young;Kim Gwe-Ya;Lim Chun-Il;Jeong Hee-Kyo;Shin Dong-Oh;Suh Tea-Suk
    • Radiation Oncology Journal
    • /
    • v.24 no.2
    • /
    • pp.138-143
    • /
    • 2006
  • Purpose: To analyze the long-term stability of Farmer-type cylindrical ionization chambers by calibration factor provided from the KFDA (Korea Food Drug Administration) Materials and Methods: The cylindrical ionization chambers used in this study were the PTW 30001 (30006), 30013, 30002, 30004, 23333, the Capintec PR06C, the WE 2571, the Exradin A12 and the Wellhofer FC65G (IC70). We were analyzed that the $N_k$ and $N_{D.W}$ calibration factor for the cylindrical chambers and compared between the measured $N_{D.W}$ and calculated $N_{D.W}$ calibration factor. Results: We have observed that the long-term stability of the PTW 30013 (30006), the Wellhofer FC65G (IC70) and the NE 2571 has varied within 0.2%. The measured $N_{D,W}$ calibration factor was about 1.0% higher than the calculated $N_{D,W}$ that determined by the $N_k$ calibration factor. Conclusion: The study has evaluated that the long-term stability of the cylindrical chambers through analysis for the $N_k\;and\;N_{D,W}$ calibration factor. It has contributed to the improvement of clinical electron dosimetry in radiotherapy centers.