An Animal Study on Electrohydraulic Type Ventricular Assist Device (전기 유압식 심실보조장치의 동물실험 연구)

  • 백완기;심상석
    • Journal of Chest Surgery
    • /
    • v.29 no.7
    • /
    • pp.689-699
    • /
    • 1996
  • An animal experiment was designed for the evaluation of in vivo performance of the newly developed electrohydraulic type ventricular assist device and its influence on the left ventricular function during pal- satile left ventricular assist. Eight adult sheep were incorporated into the study and data were collected from seven sheep. Total as- sist time ranged from 69 minutes to 7 days. The performance of the device was satisfactory both in asyn- chr nous and synchronous mode within the range of given native heart rate. More than 4 liters of device output could be reached within the range of normal left atral pressure without development of negative pressure in the left atrium. Moderate to severe degree of hemolysis was noted as evidenced by significant increase of plasma free hemoglobin level after 3 days of left ventricular support along with the presence of the small amount of thrombi around the floating disc type polymer valve apparatus reflecting that further study and refinement of the device need to be done in regard of biocompatibility and thromboresistance. The hemodynamics showed increase in heart rate (p < 0.05), cardiac output and left ventricular minute work (p < 0.05) after placement of the device at the flow rate of 2.0∼2.5 Llmin. The left atrial pressure, left ventricular pressure and LV dpldt were decreased after the device placement(p < 0.05). The endocardial viability ratio and oxygen contents of the mixed ven us blood and coronary venous blood were all increased (p < 0.05) after the device placement suggesting effective unloading of the left ventricle was accomplished. The myocardial perfusion was thought improved in synchronous counterpulsation as suggested by sig- nificant increase in endocardial viability ratio and coronary venous blood oxygen content in synchronous assist mode comparing with asynchronous mode.

  • PDF