• Title/Summary/Keyword: 관자뼈 컴퓨터단층촬영

Search Result 1, Processing Time 0.019 seconds

A Study on the Dose Reduction Method for Temporal Bone HRCT Scan (관자뼈 HRCT 스캔 시 선량감소 방법에 관한 연구)

  • Joon Yoon;Hyeon-Ju Kim
    • Journal of the Korean Society of Radiology
    • /
    • v.17 no.7
    • /
    • pp.1041-1047
    • /
    • 2023
  • Temporal bone CT, which is a high-resolution CT, uses a high tube voltage and a thin section thickness, so the scan dose is higher than that of adjacent areas. Accordingly, we applied changes to the reconstruction algorithm among the test conditions to find an algorithm with excellent sensitivity to lesions while reducing the test dose, and investigated its significance and the possibility of providing basic clinical data. As a result, when the tube voltage was lowered to 100 kVp and applied, the dose was reduced by about 35.6%, and when the definition algorithm was applied to the raw data acquired at 100 kVp, the SNR and CNR were excellent, and a statistically significant difference was shown when compared to other algorithms(p<0.05). And as a result of comparing structural similarity, the SSIM index was analyzed as 0.776, 0.813, and 0.741 for each ROI. Therefore, we believe that applying algorithm changes to temporal bone CT scans can partially reduce the dose generated from CT scans and are very meaningful in terms of basic clinical data.