• Title/Summary/Keyword: 고관절 대치술

Search Result 5, Processing Time 0.023 seconds

The Distal Filling Effects on Hip Jont Function in Cementless Total Hip Replacement (인공 고관절 대치술에서 무시멘트형 스템의 원위부 압박이 고관철 성능에 미치는 영향)

  • 채수원;박상석;박재원
    • Transactions of the Korean Society of Mechanical Engineers A
    • /
    • v.24 no.11
    • /
    • pp.2777-2785
    • /
    • 2000
  • In cementless total hip replacement(THR), an initial stability of the femoral component is important to long term fixation of femoral stem. The intial stability has close relationship with the relative displacement of prosthessis and sponge bone at the proximal of femur. After implantation of the proshesis, the surrounding bone is partially shielded from load carrying and starts to resorb. Stress shielding is the cause of the loss of proximal bone. Assessing stress distribution of femur is important to predict stress shielding. The initial stability and the stress shielding were investigated for two loading conditions approximating a single leg stance and a stair climbing. Three types of stems were studied by the finite element method to analyze the biomechanical effects of distal filling of cementless femoral stems. Three types of stems empolyed are a distal filling stem, a distal flexible stem, and a distal tapered stem.

A Study on the Contact Stress Analysis of the Artificial Hip Joint Using Finite Element Method (유한요소법을 이용한 인공 고관절의 접촉응력 해석에 관한 연구)

  • 김청균;윤종덕
    • Proceedings of the Korean Society of Tribologists and Lubrication Engineers Conference
    • /
    • 1996.10a
    • /
    • pp.189-194
    • /
    • 1996
  • The modern orthopaedics frequently uses the total hip replacement in the artificial hip joint. The wear in this joint requires the re-replacement of the hip joint beacuse it is under the severe load and friction conditions. To solve these problems the previous studies have been mainly focused on the development of the new materials. The research of new material, however, needs much time and effort since it should be experimented for its bio-compatibility, friction, and wear characteristics. To reduce the work, in this study, the finite element analysis is applied to find the new combination of the materials in the total hip replacement which has the excellent contact characteristics. The finite element uses MARC and the 5-node aximetric element. The results show that in case of acetabulum UHMWP has good characteristics, in femoral head, the $Al_2O_3$, and in stem, Ti6Al4V.

  • PDF

Biomechanical characteristics of the distal filling effects in cementless femoral stem (무시멘트형 대퇴스템에서 원위부 압박 정도에 따른 생체역학적 특성)

  • Park, Sang-eok;Park, Jae-Won;Chae, Soo-Won
    • Proceedings of the KSME Conference
    • /
    • 2000.04a
    • /
    • pp.387-392
    • /
    • 2000
  • In cementless total hip replacement(THR), an initial stability of the femoral component is important to long term fixation of femoral stem. The initial stability has close relationship with the relative displacement of prosthesis and spongy bone at the proximal of femur. After implantation of the prosthesis. the surrounding bone is partially shielded from load carrying and starts to resort. Stress shielding is the cause of the loss of proximal bone. Assessing stress distribution of femur is important to predict stress shielding. The initial stability and the stress shielding were investigated for two loading conditions approximating a single leg stance and a stair climbing. Three types of stems were studied by the finite element method to analyze the biomechanical effects of distal filling of cementless femoral stems, Three types of stems employed are a distal filling stem, a distal flexible stem, and a distal tapered stem.

  • PDF

Surgical Treatment of Metastatic Tumor in Pelvis (골반부 전이성 종양의 수술적 치료)

  • Kim, Jae-Do;Park, Woong;Jo, Myung-Rae;Son, Jung-Whan;Lee, Young-Gu
    • The Journal of the Korean bone and joint tumor society
    • /
    • v.10 no.2
    • /
    • pp.61-70
    • /
    • 2004
  • Purpose: We studied to decide the operative indication of the metastatic tumor in pelvis according to the oncologic results, the Eastern Cooperative Oncologic Group (ECOG) performance status and complication. Materials and methods: From May 1994 to May 2003, 9 patients who were performed on palliative treatment and 10 paitents on operative treatment due to metastatic tumor of pelvic bone were investigated. On palliative/operative group, the mean age of patients was 57.6/48.0 years old and the ratio of male to female was 5:4/7:3. Primary origins were 3 cases from kidney, 3 from cervix and 2 of lung, 2 of myeloma, 2 of Non-Hodgkin's Lymphoma, and 1 from breast, bladder, testis, prostate, stomach, liver and retroperitoneal leimyosarcoma respectively. The palliative treatment was performed in 5 cases with radiotherapy, 1 with chemotherapy, 2 with combined chemo-radiotherapy and 1 with percutaneous cementation. The operative methods were 1 case of bone cement insertion after curettage, 2 of Girdlestone with internal hemipelvectomy and 7 of reconstruction after wide excision. Reconstructions were done.: 1 case of bone cementation, 5 of autograft prosthesis composite with irradiation or pastuerization and 1 of saddle prosthesis. We have observed the oncologic results, the ECOG performance status and complication. Results: The oncologic results of palliative/operative groups are NED 0/1, AWD 2/6, DOC 1/2 and DOD 6/1. The ECOG performance status was changed from 1.5 into 4.3 in palliative group and from 2.6 into 2.2 in operative group. The complications were 3 cases of the prosthesis failure and 2 of infection. Conclusion: The indication of operation of metastatic pelvic tumor is decided in consideration of the patient's condition, the grade of malignancy in primary tumor and the life expectancy.

  • PDF