• Title/Summary/Keyword: 갑상샘 신티그래피

Search Result 1, Processing Time 0.014 seconds

Evaluation of Validity Thyroid Scintigraphy Using Parallel Hole Collimator (갑상샘 신티그래피 검사 시 평행다공형 조준기 적용의 유효성 평가)

  • Su-Young Park;Ji-Youn Kim;Sung-Min Ahn
    • Journal of the Korean Society of Radiology
    • /
    • v.18 no.1
    • /
    • pp.27-36
    • /
    • 2024
  • In this study, When acquisition thyroid scintigraphy images, a parallel hole collimator was applied, and the difference from the pinhole collimator was quantitatively analyzed under each image acquisition condition. Visual size, resolution, sensitivity, signal to noise ratio (SNR), and contrast to noise ratio (CNR) were evaluated using thyroid phantom and point source. When comparing visual size, it was confirmed that an image similar to the size of the pinhole collimator could be obtained only when a magnification ratio of about 2.00 to 2.09 times when applying a parallel hole collimator. There was no tendency in FWHM(mm) measurement using a point source, and sensitivity was high in the parallel hole collimator. SNR and CNR were high when using a low magnification ratio, matrix size of 128×128, and a parallel hole collimator. In images of similar size to the naked eye, when the matrix size was the same, both SNR and CNR were high in the pinhole collimator. Therefore, when performing a thyroid scintigraphy test, if appropriate conditions are set according to the situation of each hospital and a parallel hole collimator is applied, it can be a good option in terms of equipment utilization and work efficiency.