A study of $Smartpeg^{TM}'s$ lifetime according to sterilization for implant stability (임플랜트 안정성을 위한 자기공명막대의 소독방법에 따른 수명에 관한 연구)

  • Won, Ho-Yeon;Cho, In-Ho;Lee, Joon-Seok
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.46 no.1
    • /
    • pp.42-52
    • /
    • 2008
  • Purpose: Resonance Frequency Analysis(RFA) technique can be used as an effective method in measuring the implant stability and documenting the clinical results. This technique also determines how stable the implant is before performing a prosthetic practice. Having become one the guidelines of the implant therapy whose final objective is the immediate loading, the $Osstell^{TM}$ mentor is giving a lot of information to the clinicians recently. In this communication, experiments were performed to investigate how reliable the measured ISQ values by $Osstell^{TM}$ mentor are, and to see if those are also stable even after sterilization. As five objectives: 1) How stable measured ISQ values after fixation $Smartpeg^{TM}s$ for 400 times. 2) How stable measured ISQ values after 'attach-detach'$Smartpeg^{TM}'s$ for 400 times. 3) How stable measured ISQ values after clinical sterilization methods. 4) How stable measured ISQ values after repeatedly sterilization in autoclave for 10 times. 5) What is the critical temperature which is lost the magnetism of $Smartpeg^{TM}$. Materials and Methods: Clinical sterilization methods(Autoclave sterilization, Dentistar sterilization, Ultra violet sterilization, Vacuum dry unit sterilization, Boiling water sterilization, combined $H_{2}O_{2}$ and Alcohol sterilization).$Smartpeg^{TM}s$. D3 Block bone($3{\times}9{\times}2cm$). Osstem implant(${\emptyset}4.1$-10mm).$Osstell^{TM}$ mentor. Individual experiment was used 8 number of $Smartpeg^{TM}s$ and they had measured to ISQ values of before experiment and after experiment. Results: 1. The measured ISQ values did not change after fixation $Smartpeg^{TM}s$ for 400 times. 2. There was no significant changes in the measured ISQ values of 'attach-detach $Smartpeg^{TM}s'$ for 400 times. 3. The measured ISQ values did not change after the usual clinical sterilization methods. 4. The measured ISQ values did not change after sterilization in autoclave for 10 times. 5. It was impossible to exactly measure the critical temperature which is lost the magnetism of $Smartpeg^{TM}s$. But, the results was resulted to lost its magnetism in higher temperature than $150^{\circ}C$/10 minute. Conclusion: The measured ISQ values showed insignificant differences in case of no changes in the magnetism of the $Smartpeg^{TM}s$. It seems that the $Smartpeg^{TM}s$ can be used repeatedly in every measurement if the original magnetisms of the $Smartpeg^{TM}s$ can be recognized. There seems to be no significant changes in the measured ISQ values of 'attach-detach $Smartpeg^{TM}s'$ only if the screw pitches were unimpaired. The clinical sterilization methods seems acceptable because the result was resulted to lost its magnetism in higher temperature than $150^{\circ}C$/10minute.

THE EFFECT OF AUTOCLAVE STERILIZATION AND REUSE OF $SMARTPEG^{TM}$ ON THE IMPLANT STABILITY QUOTIENT (ISQ) MEASUREMENT ($Smartpeg^{TM}$의 고압멸균소독 및 재사용이 임플랜트 안정성 지수(ISQ) 측정에 미치는 영향)

  • Kang, In-Ho;Kim, Myung-Joo;Lim, Young-Jun;Kim, Chang-Whe
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.45 no.5
    • /
    • pp.644-652
    • /
    • 2007
  • Statement of problem: Resonance frequency analysis is evaluated as the non-invasive and objective method for the evaluation of implant stability and has been increasingly used. It is necessary to evaluate the factors affect the ISQ measurement stability. Purpose: This study was performed to evaluate the effect of the autoclave sterilization and reuse of $Smartpeg^{TM}$ on ISQ measurement. Material and methods: $SmartPeg^{TM}$ (Integration Diagnostics Ltd., $G\ddot{o}teborg$, Sweden) of autoclave group (A) was autoclave sterilized 9 times and $Smartpeg^{TM}$ of reuse group (R) was reused 9 times. Ten $SmartPeg^{TM}s$ were allocated to each group and after each autoclave sterilization and reuse, implant stability quotient (ISQ) values were measured 3 times from the two directions a and b at a right angle. $Osstell^{TM}$ mentor (Integration Diagnostics Ltd. $G\ddot{o}teborg$, Sweden) was used and type 1 (article no. 100353) $Smartpeg^{TM}$ was selected according to $Smartpeg^{TM}$ reference list. Osstem Implant US II future (Osstem Co., Seoul, Korea) in $4.0mm{\times}11.5mm$ was embedded in the self-curing acrylic resin ($Orthojet^{(R)}$, Lang Dental, U.S.A.). Data was statistically analyzed by one-way ANOVA $({\alpha}=.05)$ and scheffe test was done where a significant difference exist. Correlation test was also done between ISQ value and the number of autoclave sterilization or reuse. Results: 1. In autoclave group, the means and sd. of ISQ value before autoclave sterilization were $84.97{\pm}0.41,\;84.93{\pm}0.74$ at direction a and b. There was significant differences between autoclave groups at direction a and b (P=.000). 2. In reuse group, the means and sd. of ISQ value before reuse were $85.40{\pm}0.62,\;85.50{\pm}0.57$ at direction a and b. There was no significant difference between reuse groups at direction a and b (P>.05). 3. There was a weak positive correlation between the number of reuse and ISQ value at direction a and b (${\gamma}=.207$ and .246, P<.01). Conclusion: Within the limitations of this study, the following conclusions were drawn. Till ninth reuse of $Smartpeg^{TM}$, the ISQ measurement stability did not be affected. After twice autoclave sterilization of $Smartpeg^{TM}$ the ISQ measurement stability was affected.