Browse > Article

Application for Measuring the Glucose, Ammonia nitrogen, and Tylosin Concentration using Near Infrared Spectroscopy  

Kim, Jong-Soo (Division of Chemical Engineering, College of Engineering, Chonbuk national university)
Cho, Hoon (Department of Polymer Science & Engineering, Chosun University)
Publication Information
Journal of environmental and Sanitary engineering / v.23, no.2, 2008 , pp. 19-25 More about this Journal
Abstract
For measurement of tylosin, ammonia nitrogen, and glucose concentration during the culture of Streptomyces fradiae using Near Infrared Spectroscopy, the calibration using various mathematical models was performed and then, based on the linear model, the validation was carried out. In the case of sucrose concentration using the MLR method, the Standard Error of Prediction and Multiple correlation coefficient were 1.97, and 0.991, respectively. In the case of ammonia nitrogen concentration using the PLSR method, the Standard Error of Prediction and Multiple correlation coefficient were 0.13, and 0.990, respectively. In the case of tylosin concentration using the PLSR method, the standard Error of Prediction and Multiple correlation coefficient were 0.54, and 0.984, respectively.
Keywords
Glucose; Ammonia nitrogen; Tylosin; Near Infrared Spectroscopy;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Cavinato, A.G., Mayes, D.M., Ge, Z., Callis, J. B.: noninvasive method for monitoring ethanol in fermentation processes using fiber-optic near-infrared spectroscopy, Anal. Chem., 62, 1977-1982, 1990   DOI   ScienceOn
2 Riley, M. R., Rhiel, M., Zhou, X., Arnold, M. A., Murhammer, D. W.: simultaneous monitoring of glucose and glutamine in insect cell culture media by NIR spectroscopy. Biotechnol. Bioeng., 55, 11-15, 1997   DOI   ScienceOn
3 Crowley, J., McCarthy, B., Nunn, N. S., Harvey, L. M., and McNeil, B.: Monitoring a recombinant Pichia pastoris fed batch process using fourier transform mid-infrared spectroscopy, Biotechnology letters, 22, 1907-1912, 2000   DOI   ScienceOn
4 Vaccari, G., Dosi, E., Campi, A. L, Gonzalez-varay R, A., Matteuzzi, D., Mantovani, G.: A near-infrared spectroscopy technique for the control of fermentation processes: application to lactic acid fermentation, Biotechnol. Bioeng., 43, 913-917, 1994   DOI   ScienceOn
5 Macaloney, G., Draper, I., Preston, J., Anderson, K. B., Rollins, M. J., Thompson, B. G., Hall, J. W., and McNeil, B.,: At-line controls and fault analysis in an industrial high cell density Escherichia coli fermentation, using NIR spectroscopy, Food bioprod. Test., 74, 212-220, 1997
6 Osborne, G. B. and Fearn. T.: Near Infrared Spectroscopy in food analysis, Longman Scientific and technical, Lonman Group UK Limited, 1986
7 Ge, Z., Cavinato, A. G., Calis, J. B.: Noninvasive spectroscopy for monitoring cell density in a fermentation process, Anal. Chem., 66, 1354-1362, 1994   DOI   ScienceOn
8 Hall, J. W., Mcneil, B., Rollins, M. J., Draper, I., Thompson, B. G., Macaloney, G.: Near-infrared spectroscopic determination of acetate, ammonium, biomass, and glycerol in an industrial Escherichia coli fermentation, Appl. Spectrosc, 50, 102-108 (1996)   DOI
9 Fayolle, P., Picque, D., Perret, B., Latrille, E., corrieu, G.: Determination of major compounds of alcoholic fermentation by middle-infrared spectroscopy: study of temperture effects and calibration methods, Appl. Spectrosc, 50:1325-1330, 1996   DOI   ScienceOn
10 Yano, T. and Harata, M.: Prediction of the concentration of several constituents in a mouse-mouse hybridoma culture by near-infrared spectroscopy, J. Ferment. Bioeng., 77:659-662, 1994   DOI   ScienceOn