Comparison of Center Error or X-ray Field and Light Field Size of Diagnostic Digital X-ray Unit according to the Hospital Grade  

Lee, Won-Jeong (Department of Radiological Technology, Daejeon Health Institute of Technology)
Song, Gyu-Ri (Department of Radiological Technology, Daejeon Health Institute of Technology)
Shin, Hyun-yi (Department of Radiological Technology, Daejeon Health Institute of Technology)
Publication Information
Journal of the Korean Society of Radiology / v.14, no.3, 2020 , pp. 245-252 More about this Journal
Abstract
The purpose of this study was intended to recognize the importance of quality control (QC) in order to reduce exposure and improve image quality by comparing the center-point (CP) of according to hospital grade and the difference between X-ray field (XF) and light field (LF) in diagnostic digital X-ray devices. XF and LF size, CP were measured in 12 digital X-ray devices at 10 hospitals located in 00 metropolitan cities. Phantom was made in different width respectively, using 0.8 mm wire after attaching to the standardized graph paper on transparent plastic plate and marked as cross wire in the center of the phantom. After placing the phantom on the table of the digital X-ray device, the images were obtained by shooting it vertically each field of survey. All images were acquired under the same conditions of exposure at distance of 100cm between the focus-detector. XF and LF size, CP error were measured using the picture archiving communication system. data were expressed as mean with standard error and then analyzed using SPSS ver. 22.0. The difference in field between the XF and LF size was the smallest in clinic, followed by university hospitals, hospitals and general hospitals. Based on the university hospitals with the least CP error, there was a statistically significant difference in CP error between university hospitals and clinics (p=0.024). Group less than 36-month after QC had fewer statistical errors than 36-month group (0.26 vs. 0.88, p=0.036). The difference between the XF and LF size was the lowest in clinic and CP error was the lowest in university hospital. Moreover, hospitals with short period of time after QC have fewer CP error and it means that introduction of timely QC according to the QC items is essential.
Keywords
Quality control; X-ray field; Light field; Center-point error; Digital X-ray unit;
Citations & Related Records
Times Cited By KSCI : 4  (Citation Analysis)
연도 인용수 순위
1 S. Hashimoto, Y. Fujita, T. Katayose, H. Mizuno, H. Saitoh, K. Karasawa, "Field-size correction factors of a radiophotoluminescent glass dosimeter for small-field and intensity-modulated radiation therapy beams", Journal of Medical Physics, Vol. 45, No. 1, pp. 382-390, 2018. http://dx.doi.org/10.1002/mp.12665   DOI
2 S. K. Choi, "Change of the Scattered Dose by Field Size in X-ray Radiography", Journal of The Korea Contents Association, Vol. 13, No. 3, pp. 198-203, 2013. http://dx.doi.org/10.5392/JKCA.2013.13.03.198   DOI
3 A. Groenewald, W. A. Groenewald, "A universal phantom suitable for quality assurance on X-ray imaging modalities", Acta Radiologica, Vol. 60, No. 11, pp. 1523-1531, 2019. http://dx.doi.org/10.1177/0284185119831685   DOI
4 S. Green, J. E. Palethorpe, D. Peach, D. A. Bradley, "Performance assessment of patient dosimetry services and X-ray quality assurance instruments used in diagnostic radiology", Applied Radiation and Isotopes, Vol. 50, No. 1, pp. 137-52, 1998. http://dx.doi.org/10.1016/S0969-8043(98)00030-X   DOI
5 C. J. Hourdakis, "Determination of the diagnostic x-ray tube practical peak voltage (PPV) from average or average peak voltage measurements", Physics in Medicine and Biology, Vol. 56, No. 7, pp. 2199-217, 2011. http://dx.doi.org/10.1088/0031-9155/56/7/019   DOI
6 I. C. Im, S. H. Lee, "Evaluation to X-ray Tube Variable Beam Limiting Device Ability Test, Colimation and Beam Alignment Test of Diagnostic X-ray Unit", Journal of The Korea Contents Association, Vol. 9, No. 3, pp. 250-255, 2009. http://dx.doi.org/10.5392/JKCA.2009.9.3.250   DOI
7 K. R. Dong, S. J. Lee, D. C. Kweon, E. H. Goo, J. E. Jung, K. S. Lee, "Actual Condition of Quality Control of X-ray Imaging System in Primary Care Institution: focused on Gwangju Metropolitan City", Journal of Radiation Protection, Vol. 35, No. 1, pp. 34-42, 2010.
8 J. H. Park, I. C. Im, K. R Dong, S. S. Kang, "A Performance Evaluation of Diagnostic X-ray Unit Depends on the Hospitals Size", Journal of Radiation Protection, Vol. 34, No. 1, pp. 31-36, 2009.
9 J. K. Park, "Medical Clinics' quality Management of X-ray Units in Gyeongbuk Area", Journal of The Korea Contents Association, Vol. 10, No. 9, pp. 267-275, 2010. http://dx.doi.org/10.5392/JKCA.2010.10.9.267   DOI
10 http://www.law.go.kr/lsInfoP.do?lsiSeq=197782&efYd=20170929#0000
11 Y. J. Heo, K. T. Kim, S. C. Noh, S. H. Nam, J. K. Park, "A Base Study on the Constancy Quality Control Test and Clause of Diagnosis Radiation Equipment", Journal of the Korean Society of Radiology, Vol 8, No. 3, pp. 105-110, 2014. http://dx.doi.org/10.7742/jksr.2014.8.3.105   DOI
12 T. L. Fauber, M. C. Dempsey, "X-ray Field Size and Patient Dosimetry", Radiologic Technology, Vol. 85, No. 2, pp. 155-161, 2013.