Browse > Article
http://dx.doi.org/10.7742/jksr.2018.12.1.47

Comparison of Lens Dose in accordance with Bismuth shielding and Patient position in Brain perfusion CT  

Gang, Eun Bo (Department of Radiological Science, Dong-Eui Institute of Technology)
Publication Information
Journal of the Korean Society of Radiology / v.12, no.1, 2018 , pp. 47-52 More about this Journal
Abstract
Brain perfusion CT scanning is often employed usefully in clinical conditions as it accurately and promptly provides information about the perfusion state of patients having acute ischemic stroke with a lot of time constraints and allows them to receive proper treatment. Despite those strengths of it, it also has a serious weakness that Lens may be exposed to a lot of dose of radiation in it. In this study, as a way to reduce the dose of radiation to Lens in brain perfusion CT scanning, this researcher conducted an experiment with Bismuth shielding and change of patients' position. TLD (TLD-100) was placed on both lens using the phantom (PBU-50), and then, in total 4 positions, parallel to IOML, parallel to IOML (Bismuth shielding), parallel to SOML, and parallel to SOML (Bismuth shielding), brain perfusion scanning was done 5 times for each position, and dose to Lens were measured. Also, to examine how the picture quality changed in different positions, 4 areas of interest were designated in 4 spots, and then, CT number and noise changes were measured and compared. According to the results of conducting one-way ANOVA on the doses measured, as the significance probability was found to be 0.000, so there was difference found in the doses of radiation to crystalline lenses. According to the results of Duncan's post-hoc test, with the scanning of being parallel to IOML as the reference, the reduction of 89.16% and 89.66% was observed in the scanning of being parallel to SOML and that of being parallel to SOML (Bismuth shielding) respectively, so the doses to Lens reduced significantly. Next, in the scanning of being parallel to IOML (Bismuth shielding), the reduction of 37.12% was found. According to the results, reduction in the doses of radiation was found the most significantly both in the scanning of being parallel to SOML and that of being parallel to SOML (Bismuth shielding). With the limit of the equivalent dose to Lens as the reference, this researcher conducted comparison with the dose to occupational exposure and dose to Public exposure in the scanning of being parallel to IOML and found 39.47% and 394.73% respectively; however in the scanning of being parallel to SOML (Bismuth shielding), considerable reduction was found as 4.08% and 40.8% respectively. According to the results of evaluation on picture quality, every image was found to meet the evaluative standards of phantom scanning in terms of the measurement of CT numbers and noise. In conclusion, it would be the most useful way to reduce the dose of radiation to Lens to use shields in brain perfusion CT scanning and adjust patients' position so that their lens will not be in the field of radiation.
Keywords
Brain Perfusion CT; IOML; SOML; CT Number; Noise;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 http://www.mfds.go.kr
2 http://www.nhis.or.kr
3 M. S. Ju, "Reduction in Radiation Dose With Reconstruction System in the Brain Perfusion CT," Department of Medical Engineering Graduate School of Soonchunhyang University Asan, Korea, 2014
4 ICRP Publication 26. Radiation protection Recommendations of the International Commission on Radiological Protection. Oxford: Pergamon Press, 1977, cited from reference7
5 Y. Imanishi, A. Fukui, H. Niimi, etal., "Radiation-induced temporary hair loss as a radiation damage only occurring in patients who had the combination of MDCT and DSA," Eur .Radiol. 15, pp 41-46, 2005.   DOI
6 I. C. Hwang, W. J. Shin, E. B. Gang,"A Comparative Study on the Lens Dose According to the Change of Shielding Material Used in Brain Computed Tomography," Journal of the Korean society of Radiology, Vol. 9, No. 1, pp.31-37, 2015.   DOI
7 S. H. Doh, W. G. Lee, M. J. Yoo, H. D. Kang, D. S. Kim, "Thermal Annealing and Fading Characteristics of LiF:Mg,Cu,P TLD" Journal of the Korean se nsors society, Vol. 7 No. 5, pp.313-318. 1998.
8 ICRP: 2007 Recommendations of the International Commission on Radiological Protection. Publication 103, Annals of the ICRP Vol. 137, No. 2-4, Pergamon Press, Oxford, 2007.
9 National Research Council (US); Committee to Assess Health Risks from Exposure to Low Levels of Ionizing Radiation. Health risks from exposure to low levels of ionizing radiation: BEIR VII phase 2. Washington DC: National Academies Press; 2006.
10 The Korea Society of Medical Imaging Technology, Textbook of Computed Tomography, Chung-Ku Publishing co., pp. 19-39, pp. 496-500, 2009.