Analysis of the cause by Pre Exposure Tube Voltage and Actual Exposure Tube Voltage deviation in Mammography Examination  

Cho, Ji-Hwan (Department of Health Care Clinic, Inje University, Busan Paik Hospital)
Lee, Hyo-Yeong (Department of Radiological Science, Dongeui University)
Im, In-Chul (Department of Radiological Science, Dongeui University)
Publication Information
Journal of the Korean Society of Radiology / v.11, no.2, 2017 , pp. 79-85 More about this Journal
Abstract
The purpose of this study is to investigate the cause analysis according to the difference between the pre exposure tube voltage and actual exposure tube voltage in mammography in connection with breast pressure thickness, breast size and body mass index and to find the improvement. The study tracked 377 women age 40 and older among the mammography examiners conducted by the National Health Insurance Corporation. It was analyzed that breast pressure thickness, breast size and body mass index according to the difference between the pre exposure tube voltage and actual exposure tube voltage among the parameters of dose report been sent to the picture archiving communication system with reference to the image with cranio-caudal projection in mammography. As are result, it shows that the thicker the breast thickness, smaller the breast size and lower body mass index, the higher the difference of tube voltage. In conclusion, the minimum tube voltage of mammography machine should be reset in order to set the tube voltage according to breast pressure thickness and breast size that are suitable for our country in mammography, in addition, it was considered that radiologist should make an effort to reduce radiation exposure and make a good quality image with reducing the difference of mammography condition by making a correct exposure condition in case of examining the patients with thin breast pressure and small breast size.
Keywords
Mammography; Pre exposure tube voltage; Actual exposure tube voltage; Breast pressure thickness; Breast size; Body mass index;
Citations & Related Records
Times Cited By KSCI : 3  (Citation Analysis)
연도 인용수 순위
1 "Patient dose recommendations guidelines in general radiology" Korea Food & Drug Administration, 2012.
2 J. G. Elmore, K. Amstrong, C. D. Lhman, “Screening forbreastcancer,” The Journal of the American Medical Association, Vol. 293, No. 10, pp. 1245-1256, 2005.   DOI
3 S. Y. Choi, C. S. Kim, "Development of Automatic Cluster Algorithm for Microcalcification in Digital Mammography" The Journal of Radiological Science and Technology, Vol. 32, No. 1, pp. 45-52, 2009.
4 Ministry of Health and Welfare, Korea Food & Drug Administration, The Korean Society of Radiology, The Korean Radiological Technologists Association, "Patient dose recommendations guidelines in mammography" Radiation safety management Series No. 16, pp. 1, 2008.
5 D. R. Dance, C. L. Skinner, K. C. Young, J. R. Beckett, C. J. Kotre, "Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol" The Joural of Physics Medicine Biology, Vol. 45, No. 11, pp. 3225-40, 2000.   DOI
6 F. Bouzarjomehri, A. Mostaar, A. Ghasemi, M. H. Ehramposh, H. Khosravi, "The study of mean glandular dose in mammography in yazd and the factors affecting it" The Journal of Iran Radiology, Vol. 4 pp. 29-35, 2006.
7 R. Klein, H. Aichinger, J. Dierker, J. T. Jansen, S. Joite-Barfuss, M. Säbel, R. Schulz-Wendtland, J. Zoetelief "Determination of average glandular dose with modern mammography units for two large groups of patients" The Journal of Physics Medicine Biology, Vol. 42, No. 4, pp. 651-71, 1997.   DOI
8 Supawitoo Sookpeng, Potjana Ketted, "Mean glandular dose from routine mammography" The Journal of Naresuan University, Vol. 14, pp. 19-26, 2006.
9 Inspection criteria for mammography devices and test method explanation: Korea Food & Drug Administration, 2001.
10 G. S. Sin, J. H. Choi, Y. H. Kim, J. M. Kim, C. K. Kim, J. H. Yang, "Patient Dose in Mammography" The Journal of Radiological Science and Technology, Vol. 28, No. 4, pp. 293, 2005.
11 H. J. Yang, S. K. Ko, M. H. Joo, "Evaluation of MTF Image by Target/Filter Combined of X-ray Tube Using Mammography", The Journal of Radiological Science and Technology, Vol. 30, No. 2, pp. 113-119, 2007.
12 GE Healthcare Senographe DS User Manual 5307907-6-S-1KO, pp. 247, 2001.