DOI QR코드

DOI QR Code

Alpha spectrometry: Avoiding recoil contamination of solid state alpha detectors

  • C. Olondo (Department of Energy Engineering, University of the Basque Country (UPV/EHU)) ;
  • R. Idoeta (Department of Energy Engineering, University of the Basque Country (UPV/EHU)) ;
  • M. Herranz (Department of Energy Engineering, University of the Basque Country (UPV/EHU))
  • 투고 : 2023.10.02
  • 심사 : 2024.06.18
  • 발행 : 2024.11.25

초록

Due to its low background, alpha spectrometry is capable of determining very low activity concentrations of alpha emitters in all types of samples, including environmental samples. This low background can be lost due to detector contamination caused by recoil nuclei, which can even render the detector useless. In this paper, the thickness of a thin film used to cover the source is calculated analytically. This film, capable of retaining these nuclei and avoiding this contamination, allows all the alpha particles emitted by the source to reach the detector. The energy loss of the alpha particles passing through it was also estimated in order to identify the peaks in the spectrum, including the unexpected ones. The validity of the calculations carried out and the effectiveness of the film in retaining the recoil nuclei were verified using thorium samples. Finally, the effect of applying a negative potential to the source on the retention of these nuclei was studied. The study was carried out using a PIPS detector under vacuum conditions (0.1 mm Hg) with the samples placed at a distance of 5 mm from the detector. The recoil suppression systems of some commonly used commercial instruments were also tested.

키워드

참고문헌

  1. C.W. Sill, D.G. Olson, Sources and prevention of recoil contamination of solid-state alpha detectors, Anal. Chem. 42 (13) (1970) 1596-1607. https://doi.org/10.1021/ac60295a016
  2. K.G.W. Inn, E. Hall, J.T. Woodward, B. Stewart, R. Pollanen, L. Selvig, S. Turner, I. Outola, S. Nour, H. Kurosaki, J. LaRosa, M. Schultz, Z. Lin, Z. Yu, C. McMahon, Use of thin collodion films to prevent recoil-ion contamination of alpha-spectrometry detectors, J. Radioanal. Nucl. Chem. 276 (2) (2008) 385-390. https://doi.org/10.1007/s10967-008-0516-y
  3. K.G.W. Inn, Thin film to prevent recoil contamination of solid-state alpha detectors, in: Proceedings of the Sixth International Conference on Methods and Applications of Radioanalytical Chemistry, Kailua-Kona, Hawaii, 2003. April 7-11.
  4. J.L. Gels, J.T. Elliston, J.W. Talnagi, S.E. Glover, H.B. Spitz, A pre-concentration neutron activation analysis method for the determination of 232Th utilizing a dry-tube irradiation facility, J. Radioanal. Nucl. Chem. 277 (1) (2007) 217-221.
  5. N. Vainblat, O. Pelled, U. German, G. Haquin, A. Tshuva, Z.B. Alfassi, Determination of parameters relevant to alpha spectrometry. De termination of parameters relevant to alpha spectrometry when employing source coating, Appl. Radiat. Isot. 61 (2004) 307-311. https://doi.org/10.1016/j.apradiso.2004.03.026
  6. J. Janda, P. Sladek, D. Sas, Electrodeposition of selected alpha-emitting radionuclides from oxalate-ammonium sulfate electrolyte and measured by means of solid-state alpha spectrometry, J. Radioanal. Nucl. Chem. 286 (2010) 687-691. https://doi.org/10.1007/s10967-010-0756-5
  7. C. Cosma, O.A. Rusu, V. Cosma, D. Nita, R. Cs Begy, A. Timar Gabor, A. Astilean, Protection of alpha spectrometry detectors using thin formvar films and influence on detection characteristics, IEEE Trans. Nucl. Sci. 59 (2012) 1175-1179. https://doi.org/10.1109/TNS.2012.2184802
  8. A. Vander der Wijk, L. Venema, S.P. Steendam, The use of thin film plastic foils in low-level alpha spectrometry, International Journal of Radiation Applications and Instrumentation, Part A, Appl. Radiat. Isot. 38 (1987) 1061-1065. https://doi.org/10.1016/0883-2889(87)90071-2
  9. N. Tsoulfadinis, S. Landsberger, Measurement and Detection of Radiation, third ed., Taylor and Francis Group, New York, 2011.
  10. Nuclide Lara, 2022. http://www.nucleide.org/Laraweb/index.php.
  11. G.F. Knoll, Radiation Detection and Measurement, second ed., John Wiley and Sons, Inc., New York, 1989.
  12. A. Stoklosa, J. Zajecki, S.S. Kurek, Effective nuclear charge of an ion, Materials Science-Poland 22 (I) (2004) 35-45.
  13. R.D. Evans, The Atomic Nucleus, first ed., McGraw-Hill Inc., New York, 1955.
  14. GUM, Evaluation of measurement data - guide to the expression of uncertainty in measurement, JCGM 100 (2008) 2008.