극한 온도환경에서의 안전조치 검증장비 사용을 위한 리트로핏 제작 및 현장 평가

Retrofit Production and Field Evaluation for Use of Safeguards Equipment in Extreme Temperature Environments

  • 투고 : 2024.02.01
  • 심사 : 2024.03.26
  • 발행 : 2024.03.31

초록

In a previous study, the suitability for use of inspection equipment was evaluated at temperatures outside the quality assurance range. The quality assurance operating temperature of the safeguards equipment is 0~+40℃, and previous studies have confirmed the performance of the safeguards equipment for temperatures ranging from -40~+70℃. The scintillator-based verification equipment showed a shift in the centroid channel and a change in the count rate in all temperature ranges, and the semiconductor-based safeguards equipment generated Leakage Current and equipment failure. In this study, a retrofit was performed applying a vacuum housing to the safeguards equipment (Inspector-2000-based inspection equipment), and performance evaluation was performed at a low temperature and snowy site, and it was confirmed that the same performance was observed as the measurement results at room temperature.

키워드

과제정보

This work was supported by the Nuclear Safety Research Program through the Korea Foundation Of Nuclear Safety (KoFONS) using the financial resource granted by the Nuclear Safety and Security Commission (NSSC) of the Republic of Korea (No. 2103090).

참고문헌

  1. Lee BC et al. 2023. Implications of changes in North Korea's nuclear strategy for South Korea." Kyungnam University Institute for Far Eastern Studies North Korea Studies Series 63
  2. Baek HK et al. 2023. Evaluation of usability in extreme temperature environments of safeguards equipment. Journal of Radiation Industry. In publication
  3. Kumar Dileep et al. 2020. "Comparative analysis of building insulation material properties and performance." Renewable and Sustainable Energy Reviews 131:110038. https://doi.org/10.1016/j.rser.2020.110038
  4. Ianakiev KD et al. 2009. "Temperature behavior of NaI(Tl) scintillation detectors." Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 607.2:432-438. https://doi.org/10.1016/j.nima.2009.02.019
  5. Park SH et al. 2010. "Effect of temperature on the performance of a CZT radiation detector." Journal of the Korean Physical Society 56.4:1079-1082. https://doi.org/10.3938/jkps.56.1079