DOI QR코드

DOI QR Code

Evaluation of Coefficients of Variation for Clinical Chemistry Tests Based on Internal Quality Control Data Across 5,425 Laboratories in China From 2013 to 2022

  • Wei Wang (National Center for Clinical Laboratories, Beijing Engineering Research Center of Laboratory Medicine, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing Hospital, National Center of Gerontology) ;
  • Zhixin Zhang (National Center for Clinical Laboratories, Beijing Engineering Research Center of Laboratory Medicine, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing Hospital, National Center of Gerontology) ;
  • Chuanbao Zhang (National Center for Clinical Laboratories, Beijing Engineering Research Center of Laboratory Medicine, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing Hospital, National Center of Gerontology) ;
  • Haijian Zhao (National Center for Clinical Laboratories, Beijing Engineering Research Center of Laboratory Medicine, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing Hospital, National Center of Gerontology) ;
  • Shuai Yuan (National Center for Clinical Laboratories, Beijing Engineering Research Center of Laboratory Medicine, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing Hospital, National Center of Gerontology) ;
  • Jiali Liu (National Center for Clinical Laboratories, Beijing Engineering Research Center of Laboratory Medicine, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing Hospital, National Center of Gerontology) ;
  • Na Dong (National Center for Clinical Laboratories, Beijing Engineering Research Center of Laboratory Medicine, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing Hospital, National Center of Gerontology) ;
  • Zhiguo Wang (National Center for Clinical Laboratories, Beijing Engineering Research Center of Laboratory Medicine, Institute of Geriatric Medicine, Chinese Academy of Medical Sciences, Beijing Hospital, National Center of Gerontology) ;
  • Fengfeng Kang (Laboratory Medicine Center, Zhejiang Center for Clinical Laboratory, Zhejiang Provincial People's Hospital (Affiliated People's Hospital), Hangzhou Medical College)
  • Received : 2023.06.05
  • Accepted : 2023.11.07
  • Published : 2024.05.01

Abstract

Background: Clinical chemistry tests are most widely used in clinical laboratories, and diverse measurement systems for these analyses are available in China. We evaluated the imprecision of clinical chemistry measurement systems based on internal QC (IQC) data. Methods: IQC data for 27 general chemistry analytes were collected in February each year from 2013 to 2022. Four performance specifications were used to calculate pass rates for CVs of IQC data in 2022. Boxplots were drawn to analyze trends of CVs, and differences in CVs among different groups were assessed using the Mann-Whitney U-test or Kruskal-Wallis test. Results: The number of participating laboratories increased significantly from 1,777 in 2013 to 5,425 in 2022. CVs significantly decreased for all 27 analytes, except creatine kinase and lipase. Triglycerides, total bilirubin, direct bilirubin, iron, and γ-glutamyl transferase achieved pass rates >80% for all goals. Nine analytes with pass rates <80% based on 1/3 allowable total error were further analyzed; the results indicated that closed systems exhibited lower CVs than open systems for all analytes, except total protein. For all nine analytes, differences were significant between tertiary hospitals and non-tertiary hospitals and between accredited and non-accredited laboratories. Conclusions: The CVs of IQC data for clinical chemistry have seen a continuous overall improvement in China. However, there is ample room for imprecision improvement for several analytes, with stricter performance specifications.

Keywords

Acknowledgement

This work was supported by the National Natural Science Foundation of China (Grant No. 81871737), Project for Science Technology Department of Zhejiang Province (Grant No. 2020C35057), and Zhejiang Provincial Project for Medical and Health Science and Technology (Grant No. 2022KY526).

References

  1. Rosenbaum MW, Flood JG, Melanson SEF, Baumann NA, Marzinke MA, Rai AJ, et al. Quality control practices for chemistry and immunochemistry in a cohort of 21 large academic medical centers. Am J Clin Pathol 2018;150:96-104. https://doi.org/10.1093/ajcp/aqy033
  2. Fei Y, Wang W, He F, Zhong K, Wang Z. Imprecision investigation and analysis of routine chemistry in China. J Clin Lab Anal 2016;30:444-50. https://doi.org/10.1002/jcla.21877
  3. Kang F, Zhang C, Wang W, Wang Z. Sigma metric analysis for performance of creatinine with fresh frozen serum. Scand J Clin Lab Invest 2016;76:40-4. https://doi.org/10.3109/00365513.2015.1091494
  4. ISO 15189:2022. Medical laboratories—Requirements for quality and competence. 4th ed. Geneva, Switzerland: International Organization for Standards, 2022.
  5. Duan M, Kang F, Zhao H, Wang W, Du Y, He F, et al. Analysis and evaluation of the external quality assessment results of quality indicators in laboratory medicine all over China from 2015 to 2018. Clin Chem Lab Med 2019;57:812-21. https://doi.org/10.1515/cclm-2018-0983
  6. Westgard JO and Burnett RW. Precision requirements for cost-effective operation of analytical processes. Clin Chem 1990;36:1629-32. https://doi.org/10.1093/clinchem/36.9.1629
  7. CNAS-CL38:2012. Guidance on the application of accreditation criteria for the medical laboratory quality and competence in the field of clinical chemistry. Beijing, China: China National Accreditation Service for Conformity Assessment, 2012.
  8. Sandberg S, Fraser CG, Horvath AR, Jansen R, Jones G, Oosterhuis W, et al. Defining analytical performance specifications: Consensus Statement from the 1st Strategic Conference of the European Federation of Clinical Chemistry and Laboratory Medicine. Clin Chem Lab Med 2015;53:833-5. https://doi.org/10.1515/cclm-2015-0067
  9. European Federation of Clinical Chemistry and Laboratory Medicine. BV data for measurands. (Updated in Feb 2023).
  10. Kang F, Li W, Lou Y, Shan Z. Application of biological variation and sigmametrics to evaluate the performance of HbA1c in external quality assessment. Scand J Clin Lab Invest 2022;82:398-403. https://doi.org/10.1080/00365513.2022.2100822
  11. Ricós C, Alvarez V, Cava F, García-Lario JV, Hernandez A, Jiménez CV, et al. Current databases on biologic variation: pros, cons and progress. Scand J Clin Lab Invest 1999;59:491-500. https://doi.org/10.1080/00365519950185229
  12. Panteghini M and Sandberg S. Defining analytical performance specifications 15 years after the Stockholm conference. Clin Chem Lab Med 2015;53:829-32. https://doi.org/10.1515/cclm-2015-0303
  13. Braga F and Panteghini M. Performance specifications for measurement uncertainty of common biochemical measurands according to Milan models. Clin Chem Lab Med 2021;59:1362-8. https://doi.org/10.1515/cclm-2021-0170
  14. Panteghini M, Braga F, Camara JE, Delatour V, Van Uytfanghe K, Vesper HW, et al. Optimizing available tools for achieving result standardization: value added by joint committee on traceability in laboratory medicine (JCTLM). Clin Chem 2021;67:1590-605. https://doi.org/10.1093/clinchem/hvab178
  15. Oosterhuis WP, Coskun A, Sandberg S, Theodorsson E. Performance specifications for sodium should not be based on biological variation. Clin Chim Acta 2023;540:117221. https://doi.org/10.1016/j.cca.2023.117221
  16. Keleş M. Evaluation of the clinical chemistry tests analytical performance with Sigma Metric by using different quality specifications - Comparison of analyser actual performance with manufacturer data. Biochem Med (Zagreb) 2022;32:010703. https://doi.org/10.11613/BM.2022.010703
  17. Molina A, Guiñon L, Perez A, Segurana A, Bedini JL, Reverter JC, et al. State of the art vs biological variability: comparison on hematology parameters using Spanish EQAS data. Int J Lab Hematol 2018;40:284-91. https://doi.org/10.1111/ijlh.12783