Abstract
Purpose: This study aims to identify SUV, SNR, spatial resolution, and axial uniformity under the same reconstruction conditions and to find out the differences between equipment models. Materials and Methods: The equipment was GE's Discovery 600, 710, IQ, MI(GE Healthcare, USA), and the Phantom used ACR(American College of Radiology) Flangeless Esser Phantom and PET/SPECT Performance Phantom. The PET/SPECT Performance Phantom injected 18F-FDG at a concentration of 3.8 kBq/mL, and the ACR Flangeless Esser Phantom made the conditions for Hot Spot and Background activity for 4 : 1. Image evaluation was compared and evaluated for SUV, SNR, spatial resolution, and axial uniformity with the same reconstruction that added SharpIR of VPHD. Results: The SUVmax showed a difference up to 4.6% with an average of 2.71, 2.35, 1.89, and 1.43 from Hot Spot 1 to 4, and the SUVmean showed a difference up to 4.7% with 2.06, 1.75, 1.49, and 1.27. There was a difference up to 5% between equipment, and there was no significant difference between both SUVmax and SUVmean. SNR showed a difference up to 0.04 with an average of 0.37, 0.26, 0.18, and 0.11. FWHM showed a difference up to 0.27. Lastly, COV of axial uniformity was up to 0.018. Conclusion: SUV showed differences within 5% between equipment and showed no significant difference. This is considered to be used as basic data that can be used for the development and replacement of equipment because it has the advantage of being able to observe with a large number of equipment.