Feasibility and Safety of a Technique Intended to Place the Catheter Tip in the Right Atrium without Abutment Against the Cardiac Wall during Implantation of the Totally Implantable Venous Access Port

완전이식형 정맥 접근 포트 삽입 시 카테터 팁을 심벽에 닿지 않고 우심방 내 위치시키는 방법의 타당성과 안전성

  • Hyejin Park (Department of Radiology, Korea University Guro Hospital, Korea University College of Medicine) ;
  • Tae-Seok Seo (Department of Radiology, Korea University Guro Hospital, Korea University College of Medicine) ;
  • Myung Gyu Song (Department of Radiology, Korea University Guro Hospital, Korea University College of Medicine) ;
  • Woo Jin Yang (Department of Radiology, Korea University Guro Hospital, Korea University College of Medicine)
  • 박혜진 (고려대학교 의과대학 고려대학교 구로병원 영상의학과) ;
  • 서태석 (고려대학교 의과대학 고려대학교 구로병원 영상의학과) ;
  • 송명규 (고려대학교 의과대학 고려대학교 구로병원 영상의학과) ;
  • 양우진 (고려대학교 의과대학 고려대학교 구로병원 영상의학과)
  • Received : 2023.05.24
  • Accepted : 2023.07.26
  • Published : 2024.01.01

Abstract

Purpose To assess the safety and feasibility of intentionally positioning the catheter tip in the right atrium (RA) without an abutment during implantation of a totally implantable venous access port (TIVAP). Materials and Methods We enrolled 330 patients who had undergone TIVAP implantation between January and December 2016 and postoperative chest CT. The TIVAP was placed using the single-incision technique to access the axillary vein directly from the incision line. To position the catheter tip in the RA without abutment, blood return was checked before cutting. Catheter length and complications were evaluated by retrospectively reviewing medical images and records. Results All patients achieved successful catheter tip positioning without abutment or dysfunction. The median tip position was 15.3 mm distal to the cavoatrial junction (CAJ) on fluoroscopy and 6 mm distal to the CAJ on CT. Catheter tips migrated a median of 10.4 mm cephalically on CT compared to fluoroscopy. Thromboses were detected in the RA and superior vena cava in one patient each. Conclusion Intentional catheter tip positioning in the RA without abutment is a safe and feasible technique with a low incidence of thrombosis and no observed dysfunction.

목적 완전이식형 정맥 접근 포트 이식 시 카테터 팁을 심벽에 닿지 않으면서 우심방 내 위치시키는 기술의 타당성과 안전성에 대하여 평가하였다. 대상과 방법 2016년 1월부터 12월까지 완전이식형 정맥 접근 포트를 삽입하고 추적관찰 CT를 시행한 환자 총 330명이 연구에 포함되었다. 정맥 접근 포트는 절개선에서 직접 액와정맥으로 접근하는 단일절개법을 적용하여 이식하였다. 카테터 팁을 자르기 전 정맥 환류를 확인하여 우심방 내벽에 닿지 않도록 위치를 확인하였다. 카테터의 길이 및 연관된 합병증은 의료영상 그리고 의료기록을 통해 후향적으로 평가하였다. 결과 모든 환자에서 카테터 팁은 심벽과의 닿음과 기능장애 없이 성공적으로 위치 조정되었다. 카테터 팁의 위치는 cavoatrial junction 기준으로 fluoroscopy에서 15.3 mm, 첫 번째 CT에서 6 mm 원위부에 있었다(중앙값). 카테터 팁은 fluoroscopy 대비 첫 번째 CT에서 10.4 mm 머리 방향으로 이동하였다(중앙값). 추적 CT에서 우심방과 상대정맥에서 각각 1건씩의 혈전증이 발견되었다. 결론 카테터 팁을 심벽에 닿지 않으면서 우심방 내 의도적으로 위치시키는 것은 혈전증 발생률이 낮고 기능장애가 없는 기술적으로 가능하고 안전한 방법이다.

Keywords

References

  1. Caers J, Fontaine C, Vinh-Hung V, De Mey J, Ponnet G, Oost C, et al. Catheter tip position as a risk factor for thrombosis associated with the use of subcutaneous infusion ports. Support Care Cancer 2005;13:325-331
  2. Petersen J, Delaney JH, Brakstad MT, Rowbotham RK, Bagley CM Jr. Silicone venous access devices positioned with their tips high in the superior vena cava are more likely to malfunction. Am J Surg 1999;178:38-41
  3. Xiang DZ, Verbeken EK, Van Lommel AT, Stas M, De Wever I. Composition and formation of the sleeve enveloping a central venous catheter. J Vasc Surg 1998;28:260-271
  4. Fletcher SJ, Bodenham AR. Safe placement of central venous catheters: where should the tip of the catheter lie? Br J Anaesth 2000;85:188-191
  5. Vesely TM. Central venous catheter tip position: a continuing controversy. J Vasc Interv Radiol 2003;14:527-534
  6. Tesselaar ME, Ouwerkerk J, Nooy MA, Rosendaal FR, Osanto S. Risk factors for catheter-related thrombosis in cancer patients. Eur J Cancer 2004;40:2253-2259
  7. Schutz JC, Patel AA, Clark TW, Solomon JA, Freiman DB, Tuite CM, et al. Relationship between chest port catheter tip position and port malfunction after interventional radiologic placement. J Vasc Interv Radiol 2004;15:581-587
  8. Pittiruti M, Hamilton H, Biffi R, MacFie J, Pertkiewicz M. ESPEN guidelines on parenteral nutrition: central venous catheters (access, care, diagnosis and therapy of complications). Clin Nutr 2009;28:365-377
  9. Bodenham Chair A, Babu S, Bennett J, Binks R, Fee P, Fox B, et al. Association of anaesthetists of Great Britain and Ireland: safe vascular access 2016. Anaesthesia 2016;71:573-585
  10. Hade AD, Beckmann LA, Basappa BK. A checklist to improve the quality of central venous catheter tip positioning. Anaesthesia 2019;74:896-903
  11. Dariushnia SR, Wallace MJ, Siddiqi NH, Towbin RB, Wojak JC, Kundu S, et al. Quality improvement guidelines for central venous access. J Vasc Interv Radiol 2010;21:976-981
  12. Kowalski CM, Kaufman JA, Rivitz SM, Geller SC, Waltman AC. Migration of central venous catheters: implications for initial catheter tip positioning. J Vasc Interv Radiol 1997;8:443-447
  13. Stonelake PA, Bodenham AR. The carina as a radiological landmark for central venous catheter tip position. Br J Anaesth 2006;96:335-340
  14. Wallace JA, Afonso E, Yu H, Birchard K, Isaacson A. Factors that predict increased catheter tip movement in left internal jugular vein implantable venous access ports upon standing. J Vasc Access 2014;16:223-226
  15. Samman M, Mujo T, Harris JJ, Coldwell DM, Hite-Potts M, Vyleta M. Subcutaneous port malfunction: a retrospective comparison between internal jugular and subclavian vein access. J Assoc Vasc Access 2015;20:229-234
  16. Schummer W, Herrmann S, Schummer C, Funke F, Steenbeck J, Fuchs J, et al. Intra-atrial ECG is not a reliable method for positioning left internal jugular vein catheters. Br J Anaesth 2003;91:481-486
  17. Nazarian GK, Bjarnason H, Dietz CA Jr, Bernadas CA, Hunter DW. Changes in tunneled catheter tip position when a patient is upright. J Vasc Interv Radiol 1997;8:437-441
  18. Aslamy Z, Dewald CL, Heffner JE. MRI of central venous anatomy: implications for central venous catheter insertion. Chest 1998;114:820-826
  19. Schuster M, Nave H, Piepenbrock S, Pabst R, Panning B. The carina as a landmark in central venous catheter placement. Br J Anaesth 2000;85:192-194
  20. Baskin KM, Jimenez RM, Cahill AM, Jawad AF, Towbin RB. Cavoatrial junction and central venous anatomy: implications for central venous access tip position. J Vasc Interv Radiol 2008;19:359-365
  21. Seo TS, Song MG, Kang EY, Lee CH, Yong HS, Doo K. A single-incision technique for placement of implantable venous access ports via the axillary vein. J Vasc Interv Radiol 2014;25:1439-1446
  22. Pittiruti M, Lamperti M. Late cardiac tamponade in adults secondary to tip position in the right atrium: an urban legend? A systematic review of the literature. J Cardiothorac Vasc Anesth 2015;29:491-495
  23. Song MG, Seo TS, Kim YH, Cho SB, Jung E, Chung HH, et al. Effectiveness of breast fixation to reduce migration of the tip of a totally implantable venous access port in women. J Vasc Access 2016;17:348-352
  24. Chabanier A, Dany F, Brutus P, Vergnoux H. Iatrogenic cardiac tamponade after central venous catheter. Clin Cardiol 1988;11:91-99
  25. Aldridge HE, Jay AW. Central venous catheters and heart perforation. CMAJ 1986;135:1082-1084
  26. Korones DN, Buzzard CJ, Asselin BL, Harris JP. Right atrial thrombi in children with cancer and indwelling catheters. J Pediatr 1996;128:841-846
  27. Bivins MH, Callahan MJ. Position-dependent ventricular tachycardia related to a peripherally inserted central catheter. Mayo Clin Proc 2000;75:414-416
  28. Lewis CA, Allen TE, Burke DR, Cardella JF, Citron SJ, Cole PE, et al. Quality improvement guidelines for central venous access. J Vasc Interv Radiol 2003;14(9 Pt 2):S231-S235
  29. Kalso E, Rosenberg PH, Vuorialho M, Pietila K. How much do arm movements displace cubital central venous catheters? Acta Anaesthesiol Scand 1982;26:354-356
  30. Wyschkon S, Loschmann JP, Scheurig-Munkler C, Nagel S, Hamm B, Elgeti T. Apparent migration of implantable port devices: normal variations in consideration of BMI. J Vasc Access 2016;17:155-161
  31. Behrendt FF, Wingen M, Katoh M, Guenther RW, Buecker A. Evaluation of catheter loops in central venous port systems. Invest Radiol 2006;41:777-780
  32. Caruso LJ, Gravenstein N, Layon AJ, Peters K, Gabrielli A. A better landmark for positioning a central venous catheter. J Clin Monit Comput 2002;17:331-334
  33. Puel V, Caudry M, Le Metayer P, Baste JC, Midy D, Marsault C, et al. Superior vena cava thrombosis related to catheter malposition in cancer chemotherapy given through implanted ports. Cancer 1993;72:2248-2252