Association between False Positive, False Negative, High-Background Cases and Humidity in One Lambda LABTypeTM HLA-DR Typing

One Lambda LABTypeTM을 이용한 HLA- DR Typing시 나타나는 위양성과 위음성, High-Background 사례와 검사실 습도의 관계분석

  • Hyang Son AN (Department of Laboratory Medicine, Seoul National University Hospital) ;
  • Minsung SOHN (Division of Health Care Science, The Cyber University of Korea)
  • 안향선 (서울대학교병원 진단검사의학과) ;
  • 손민성 (고려사이버대학교 보건의료학부)
  • Received : 2023.07.23
  • Accepted : 2023.08.25
  • Published : 2023.09.30

Abstract

Human leukocyte antigen (HLA) typing was performed in the diagnostic immunology laboratory of the Seoul National University Hospital. Among 611 HLA-DR tests, specific bead reactions suspected of being false positive and false negative in Lot 20 reagents were found. Therefore, we aimed to identify the factors causing cut-off corrections by examining cases where cut-off corrections were not made for 533 test results and cases where cut-off corrections were made for 78 cases after the cut-off corrections of specific beads. Frequency analysis was conducted to verify the demographic characteristics, and descriptive statistics were used to assess the humidity in the laboratory as a variable. Cross-tabulation was done to examine the association between cut-off corrections and demographic characteristics. Independent samples t-tests were conducted to verify the difference in humidity based on cut-off corrections. Finally, logistic regression analysis was conducted to examine the relationship between humidity levels and the rate of cut-off corrections, and results showed as the humidity level in the laboratory highs, the number of cut-off corrections decreased by a factor of 0.986. This suggests cut-off corrections rate increases when the humidity lowers. Therefore, it indicates that humidity in the laboratory is also a factor that affects HLA typing results.

서울대학교병원 진단면역검사실에서 HLA 형별 검사를 위한 결과 산출하는 과정에서 HLA-DR 검사 시행 건수 총 611건 중 Lot No.20의 시약에서 빈번히 나타나는 위양성과 위음성으로 의심되는 특이 bead 들의 반응을 발견하게 되었다. 따라서 정확한 결과 산출과정을 모색하기 위하여 특이 bead들의 cut-off 수정을 하지 않은 검사결과 533건과 특이 bead의 cut-off 수정을 한 뒤 결과 산출을 한 78건의 사례들을 가지고 cut-off 수정을 야기시키는 요인을 여러 변수로서 규명하고자 하였다. 검체대상의 인구통계특성과 cut-off 수정 여부를 확인하기 위해 빈도분석, 검사실의 습도를 변수로 넣어 기술통계를 진행하였고, cut-off 수정 여부와 인구통계특성의 연관성을 확인하기 위하여 교차분석을 시행, cut-off 수정 여부에 따른 습도의 차이를 검증하기 위해 독립표본 t검증을 실시하였다. 마지막으로 습도수준에 따른 cut-off 수정비율의 관계를 검증하기 위해 로지스틱 회귀분석을 실시하였다. 결과적으로 검사실의 습도 수준이 증가함에 따라 cut-off 수정사례는 0.986배 감소하는 것으로 나타났다. 이는 습도가 낮아질수록 cut-off 수정률이 증가한다는 것을 의미한다. 따라서 검사실의 습도 또한 HLA typing 결과에 영향을 미치는 요인임을 시사한다.

Keywords

Acknowledgement

This article is based on a part of the first author's master's thesis.

References

  1. Goodman JW. Antigen presentation & the major histocompatibility complex. In: Stites DP, Terr AI, Parslow TG, editors. Basic & clinical immunology. 8th ed. Appleton & Lange: 1994. p. 58-65.
  2. Johnson AH, Hurley CK, Hartzman RJ, Wade JA. The major histocompatibility of the man. In: Henry JB, editor. Clinical diagnosis and management by laboratory methods. 20th ed. WB Saunders: 2001. p. 927-948.
  3. Mytilineos J, Scherer S, Dunckley H, Chapman J, Middleton D, Opelz G. Comparison of serological and DNA HLA-DR typing results for transplantation in Western Europe, Eastern Europe, North America and South America. Transpl Int. 1994;7 Suppl 1: S519-S521. https://doi.org/10.1111/j.1432-2277.1994.tb01433.x
  4. Tiercy JM, Goumaz C, Mach B, Jeannet M. Application of HLA-DR oligotyping to 110 kidney transplant patients with doubtful serological typing. Transplantation. 1991;51:1110-1114. https://doi.org/10.1097/00007890-199105000-00034
  5. Buyse I, Decorte R, Baens M, Cuppens H, Semana G, Emonds MP, et al. Rapid DNA typing of class II HLA antigens using the polymerase chain reaction and reverse dot blot hybridization. Tissue Antigens. 1993;41:1-14. https://doi.org/10.1111/j.1399-0039.1993.tb01970.x
  6. Terasaki PI, Bernoco D, Park MS, Ozturk G, Iwaki Y. Microdroplet testing for HLA-A, -B, -C, and -D antigens. The Phillip Levine Award Lecture. Am J Clin Pathol. 1978;69:103-120. https://doi.org/10.1093/ajcp/69.2.103
  7. Bannai M, Tokunaga K, Lin L, Kuwata S, Mazda T, Amaki I, et al. Discrimination of human HLA-DRB1 alleles by PCR-SSCP (single-strand conformation polymorphism) method. Eur J Immunogenet. 1994;21:1-9. https://doi.org/10.1111/j.1744-313x.1994.tb00170.x
  8. Olerup O, Zetterquist H. HLA-DR typing by PCR amplification with sequence-specific primers (PCR-SSP) in 2 hours: an alternative to serological DR typing in clinical practice including donor-recipient matching in cadaveric transplantation. Tissue Antigens. 1992;39:225-235. https://doi.org/10.1111/j.1399-0039.1992.tb01940.x
  9. Ota M, Seki T, Fukushima H, Tsuji K, Inoko H. HLA-DRB1 genotyping by modified PCR-RFLP method combined with group-specific primers. Tissue Antigens. 1992;39:187-202. https://doi.org/10.1111/j.1399-0039.1992.tb01935.x
  10. Wu J, Griffith BB, Bassinger S, Moehlenkamp C, Brodie SG, Wu Y, et al. Strategies for unambiguous detection of allelic heterozygosity via direct DNA sequencing of PCR products: application to the HLA DRB1 locus. Mol Diagn. 1996;1:89-98. https://doi.org/10.1016/S1084-8592(96)70024-3
  11. Park JS, Han TJ, Oh HB. HLA-DR15 and HLA-DR16 typing with sequence-specific primer-polymerase chain reaction. J Clin Pathol Qual Control. 1998;20:407-410.
  12. Song EY, Park SK, Kim SM, Kim BC, Han BY, Lim YM, et al. Analysis of Ambiguities of HLA-DR typing using the Dynal RELI(TM) SSO HLA-DRB kit and development of an 'Interpretation Program for Koreans'. Korean J Lab Med. 2002;22:267-277.
  13. Yang YS, Kim DW. Evaluation of a commercial kit "INNO-LiPA HLA DRB" for HLA-DR genotyping. Korean J Clin Pathol. 1996;16:228-237.
  14. Choi SJ, Won CY, Oh HB. HLA-DR typing with PEL-FREEZ(R) DR4 TEST-SSP UNITRAY(TM). J Clin Pathol Qual Control. 1999;21:237-241.
  15. Thonnard J, Deldime F, Heusterspreute M, Delepaut B, Hanon F, De Bruyere M, et al. HLA class II genotyping: two assay systems compared. Clin Chem. 1995;41:553-556. https://doi.org/10.1093/clinchem/41.4.553
  16. Sada M, Hashimoto M, Kinoshita T, Ichikawa Y, Takahara S, Tada M, et al. Importance of HLA-DRB1 amino acid residue matching between recipient and donor in cadaveric renal transplantation. Transplant Proc. 1995;27:698-700.
  17. Clinical and Laboratory Standards Institute (CLSI). Laboratory design; approved guideline. 2nd ed. CLSI: 2007. p73.
  18. Burtis CA, Begovich JM, Watson JS. Factors influencing evaporation from sample cups and assessment of their effect on analytical error. Clin Chem. 1975;21:1907-1917. https://doi.org/10.1093/clinchem/21.13.1907
  19. Um JH, Park HR, Jung KE, Cho MJ. The effect of sample evaporation on the che mical analysis results. J Pathol Transl Med. 1983;17:21-25.
  20. Glenn GC, Hathaway TK. Effects of specimen evaporation on quality control. Am J Clin Pathol. 1976;66:645-652. https://doi.org/10.1093/ajcp/66.4.645
  21. Takahara S, Sada M, Hatori M, Wang JD, Tsuji T, Kokado Y, et al. Importance of HLA-DRB1 molecular matching between recipient and donor in cadaveric renal transplantation. Transplant Proc. 1996;28:1255-1256. https://doi.org/10.1016/0198-8859(96)85301-8
  22. One Lambda, Inc. LABTypeTM SSO typing tests [Internet]. One Lambda, Inc [cited 2019 December 1]. Available from: https://www.thermofisher.com/onelambda/us/en/products.html