Comparison of Ablation Performance between Octopus Multipurpose Electrode and Conventional Octopus Electrode

  • Sae-Jin Park (Department of Radiology, Seoul National University Hospital) ;
  • Jae Hyun Kim (Department of Radiology, Seoul National University Hospital) ;
  • Jeong Hee Yoon (Department of Radiology, Seoul National University Hospital) ;
  • Jeong Min Lee (Department of Radiology, Seoul National University Hospital)
  • Received : 2022.07.10
  • Accepted : 2022.11.01
  • Published : 2023.02.01

Abstract

Objective: To compare Octopus multipurpose (MP) electrodes, which are capable of saline instillation and direct tissue temperature measurement, and conventional electrodes for radiofrequency ablation (RFA) in porcine livers in vivo. Materials and Methods: Sixteen pigs were used in this study. In the first experiment, RFA was performed in the liver for 6 minutes using Octopus MP electrodes (n = 15 ablation zones) and conventional electrodes (n = 12 ablation zones) to investigate the effect of saline instillation. The ablation energy, electrical impedance, and ablation volume of the two electrodes were compared. In the second experiment, RFA was performed near the gallbladder (GB) and colon using Octopus MP electrodes (n = 12 ablation zones for each) with direct tissue temperature monitoring and conventional electrodes (n = 11 ablation zones for each). RFA was discontinued when the temperature increased to > 60℃ in the Octopus MP electrode group, whereas RFA was performed for a total of 6 minutes in the conventional electrode group. Thermal injury was assessed and compared between the two groups by pathological examination. Results: In the first experiment, the ablation volume and total energy delivered in the Octopus MP electrode group were significantly larger than those in the conventional electrode group (15.7 ± 4.26 cm3 vs. 12.5 ± 2.14 cm3, p = 0.027; 5.48 ± 0.49 Kcal vs. 5.04 ± 0.49 Kcal, p = 0.029). In the second experiment, thermal injury to the GB and colon was less frequently noted in the Octopus MP electrode group than that in the conventional electrode group (16.7% [2/12] vs. 90.9% [10/11] for GB and 8.3% [1/12] vs. 90.9% [10/11] for colon, p < 0.001 for all). The total energy delivered around the GB (2.65 ± 1.07 Kcal vs. 5.04 ± 0.66 Kcal) and colon (2.58 ± 0.57 Kcal vs. 5.17 ± 0.90 Kcal) were significantly lower in the Octopus MP electrode group than that in the conventional electrode group (p < 0.001 for all). Conclusion: RFA using the Octopus MP electrodes induced a larger ablation volume and resulted in less thermal injury to the adjacent organs compared with conventional electrodes.

Keywords

Acknowledgement

This work was supported by project for Cooperative R&D between Industry, Academy, and Research Institute funded Korea Ministry of SMEs and Startups in 2020 (Grants No. S2889928).

References

  1. Wu YZ, Li B, Wang T, Wang SJ, Zhou YM. Radiofrequency ablation vs hepatic resection for solitary colorectal liver metastasis: a meta-analysis. World J Gastroenterol 2011;17:4143-4148
  2. Salhab M, Canelo R. An overview of evidence-based management of hepatocellular carcinoma: a meta-analysis. J Cancer Res Ther 2011;7:463-475
  3. Bruix J, Sherman M. Management of hepatocellular carcinoma: an update. Hepatology 2011;53:1020-1022
  4. Byrne TJ, Rakela J. Loco-regional therapies for patients with hepatocellular carcinoma awaiting liver transplantation: selecting an optimal therapy. World J Transplant 2016;6:306-313
  5. European Association for the Study of the Liver, European Organisation for Research and Treatment of Cancer. EASL-EORTC clinical practice guidelines: management of hepatocellular carcinoma. J Hepatol 2012;56:908-943
  6. Chen MS, Li JQ, Zheng Y, Guo RP, Liang HH, Zhang YQ, et al. A prospective randomized trial comparing percutaneous local ablative therapy and partial hepatectomy for small hepatocellular carcinoma. Ann Surg 2006;243:321-328
  7. Wang JH, Wang CC, Hung CH, Chen CL, Lu SN. Survival comparison between surgical resection and radiofrequency ablation for patients in BCLC very early/early stage hepatocellular carcinoma. J Hepatol 2012;56:412-418
  8. Solazzo SA, Ahmed M, Liu Z, Hines-Peralta AU, Goldberg SN. High-power generator for radiofrequency ablation: larger electrodes and pulsing algorithms in bovine ex vivo and porcine in vivo settings. Radiology 2007;242:743-750
  9. Goldberg SN. Radiofrequency tumor ablation: principles and techniques. Eur J Ultrasound 2001;13:129-147
  10. Wang C, Wang Q, Zhang D, He Z, Chen W. Perfused hypertonic-saline-augmented needle enlarges ablation zones in ex vivo porcine livers. Oncol Lett 2018;15:972-978
  11. Cha J, Kim YS, Rhim H, Lim HK, Choi D, Lee MW. Radiofrequency ablation using a new type of internally cooled electrode with an adjustable active tip: an experimental study in ex vivo bovine and in vivo porcine livers. Eur J Radiol 2011;77:516-521
  12. Rhim H, Dodd GD 3rd, Chintapalli KN, Wood BJ, Dupuy DE, Hvizda JL, et al. Radiofrequency thermal ablation of abdominal tumors: lessons learned from complications. Radiographics 2004;24:41-52
  13. Rhim H, Yoon KH, Lee JM, Cho Y, Cho JS, Kim SH, et al. Major complications after radio-frequency thermal ablation of hepatic tumors: spectrum of imaging findings. Radiographics 2003;23:123-134; discussion 134-136
  14. Lee JM, Kim SH, Han JK, Sohn KL, Choi BI. Ex vivo experiment of saline-enhanced hepatic bipolar radiofrequency ablation with a perfused needle electrode: comparison with conventional monopolar and simultaneous monopolar modes. Cardiovasc Intervent Radiol 2005;28:338-345
  15. Schmidt D, Trubenbach J, Brieger J, Koenig C, Putzhammer H, Duda SH, et al. Automated saline-enhanced radiofrequency thermal ablation: initial results in ex vivo bovine livers. AJR Am J Roentgenol 2003;180:163-165
  16. Lee JM, Han JK, Kim SH, Choi SH, An SK, Han CJ, et al. Bipolar radiofrequency ablation using wet-cooled electrodes: an in vitro experimental study in bovine liver. AJR Am J Roentgenol 2005;184:391-397
  17. Bruners P, Lipka J, Gunther RW, Schmitz-Rode T, Mahnken AH. Bipolar radiofrequency ablation: is the shape of the coagulation volume different in comparison to monopolar RF-ablation using variable active tip lengths? Minim Invasive Ther Allied Technol 2008;17:267-274
  18. Widmann G, Bodner G, Bale R. Tumour ablation: technical aspects. Cancer Imaging 2009;9 Spec No A:S63-S67
  19. Ahmed M, Solbiati L, Brace CL, Breen DJ, Callstrom MR, Charboneau JW, et al. Image-guided tumor ablation: standardization of terminology and reporting criteria--a 10-year update. Radiology 2014;273:241-260
  20. Dhiman M, Kumawat AK, Ramjee Repaka. Directional ablation in radiofrequency ablation using a multi-tine electrode functioning in multipolar mode: an in-silico study using a finite set of states. Comput Biol Med 2020;126:104007
  21. Wang X, Sofocleous CT, Erinjeri JP, Petre EN, Gonen M, Do KG, et al. Margin size is an independent predictor of local tumor progression after ablation of colon cancer liver metastases. Cardiovasc Intervent Radiol 2013;36:166-175
  22. Kim YS, Lee WJ, Rhim H, Lim HK, Choi D, Lee JY. The minimal ablative margin of radiofrequency ablation of hepatocellular carcinoma (> 2 and < 5 cm) needed to prevent local tumor progression: 3D quantitative assessment using CT image fusion. AJR Am J Roentgenol 2010;195:758-765
  23. Lee DH, Lee JM, Kim PN, Jang YJ, Kang TW, Rhim H, et al. Whole tumor ablation of locally recurred hepatocellular carcinoma including retained iodized oil after transarterial chemoembolization improves progression-free survival. Eur Radiol 2019;29:5052-5062
  24. Lee ES, Lee JM, Kim WS, Choi SH, Joo I, Kim M, et al. Multiple-electrode radiofrequency ablations using Octopus® electrodes in an in vivo porcine liver model. Br J Radiol 2012;85:e609-e615
  25. Schutt DJ, Haemmerich D. Effects of variation in perfusion rates and of perfusion models in computational models of radio frequency tumor ablation. Med Phys 2008;35:3462-3470
  26. Chang W, Lee JM, Lee DH, Yoon JH, Kim YJ, Yoon JH, et al. Comparison of switching bipolar ablation with multiple cooled wet electrodes and switching monopolar ablation with separable clustered electrode in treatment of small hepatocellular carcinoma: a randomized controlled trial. PLoS One 2018;13:e0192173
  27. Suh YS, Choi JW, Yoon JH, Lee DH, Kim YJ, Lee JH, et al. No-touch vs. conventional radiofrequency ablation using twin internally cooled wet electrodes for small hepatocellular carcinomas: a randomized prospective comparative study. Korean J Radiol 2021;22:1974-1984
  28. Hong HP, Lee JY, Lee MY, Jung KU, Kim BI, Son BH, et al. Early experience with laparoscopic treatment of liver tumors using a separable cluster electrode with a no-touch technique. Wideochir Inne Tech Maloinwazyjne 2021;16:76-82
  29. Lee DH, Lee MW, Kim PN, Lee YJ, Park HS, Lee JM. Outcome of no-touch radiofrequency ablation for small hepatocellular carcinoma: a multicenter clinical trial. Radiology 2021;301:229-236
  30. Choi IY, Kim PN, Lee SG, Won HJ, Shin YM. Efficacy and safety of radiofrequency ablation for focal hepatic lesions adjacent to gallbladder: reconfiguration of the ablation zone through probe relocation and ablation time reduction. J Vasc Interv Radiol 2017;28:1395-1399