A Study on Image Change according to Grid Ratio in Digital Radiography

디지털 방사선에서 격자 비에 따른 영상변화에 관한 연구

  • Sung-Hun Jeong (Department of Radiological Science, Gimcheon University)
  • 정성훈 (김천대학교 방사선학과)
  • Received : 2023.01.24
  • Accepted : 2023.03.05
  • Published : 2023.04.30

Abstract

With the introduction of the D.R system, the grid with high removal rate of scattered radiation is selected and used without considering the grid performance evaluation. Despite the high removal rate of scattered radiation belonging to high grid ratio, it is also possible to see the cut-off phenomenon in which the primary ray involved in the formation of image could be removed as well. Thus, the purpose of this study is to provide the basic data for researches on the usefulness of using the grid by presenting the appropriate grid ratio in the D.R system through the measurement of image in grid such as non-grid, 6:1, 8:1, 10:1, 12:1. The equipments used for this study were radiation generator with grid in 12:1, 10:1, 8:1, 6:1, indirect-type detector, and acryl phantom. As the study for image evaluation, this study measured the SNR, PSNR, MSE, and Entropy. In the results of this study, the PSNR was the highest in 6:1 and the lowest in 8:1. The SNR was high in 6:1 and 8:1, and the lowest in 12:1. In case of Entropy, it was high in 8:1 and 10:1, and the lowest in 12:1. Therefore, when the grid is used, it would be more proper to choose the grid in 8:1 or 10:1 with less loss in information content of primary ray rather than the high grid ratio showing the increased patient exposure dose.

Keywords

Acknowledgement

This work was supported by the Gimcheon University Research Grant of 2021

References

  1. Seo WJ, Seo JB, Lee JW. Using Image J program, compared of focusing distance and grid rate. Korean Journal Digital Imaging Med. 2012;14(1):10-5.
  2. Kim DS. Artifact Reduction in Digital Radiography Images with the Stationary Grid Based on 1-Dimensional Filters. Korean Society of Electronic Engineering. 2010;47(5):117-8.
  3. Lee SJ, Cho HS, Choi SG, Cho HM, Oh JN, Lee SY, Park YN, et al. Study on a moirm Artifact in the Use of Carbon Interspaced Antiscatter Grids for Digital Radiography. Journal of the Korean Society of Radiology. 2008;2(4):5-7. https://doi.org/10.3348/jkrs.2008.59.1.5
  4. Seo JH, Jeong JH, Shim JH, Won JG, Jung SW, Ha SY, Baek CH. Investigation of Scattered Radiation on Images Effect in Digital Radiograph. Journal of the Korean Society of Radiology. Spring Conference Discussion. 2014;8(1):77-9.
  5. Ko IH, Kim SH, Kang CS, Lee KS. An experimental study on performance evaluation and use of diagnostic X-ray Grid being used in che-ju province. Korea Radiological Association. 1999;25(1):192-5.
  6. Kim DS. Grid Angle Optimization for Grid Artifact Reduction in Digital Radiography. Korean Society of Broadcasting Engineers. 2010;2010(7):439-40.
  7. Lee SH, Kim GH, Yang ON. Image Analysis of Micro Lesions According to Grid Frequency After Removal of Moire Artifact. Journal of Radiological Science and Technology. 2018;41(5):463-5. https://doi.org/10.17946/JRST.2018.41.5.463
  8. Choi SH. A Study on Dose and Image Quality in D.R System Evaluation According to Use of the Grid, Department of Radiological Science. The Graduate School of Health Promotion. Hanseo University. 2012:9-13.
  9. Kwon DM, Kim GJ, Kim YG, Im YI, Kim HJ, Park BR. Analog & Digital.PACS medical imaging informatics. Daihak Seorim. 2021:60-9.
  10. Nam TH. Medical imaging practice using matlab. Wolsong. 2013:108-12.
  11. Choi SH. A Study on Dose and Image Quality in DR System Evaluation According to Use of the Grid, Department of Radiological Science, The Graduate School of Health Promotion. Hanseo University. 2012:27-32.
  12. Go SJ, Goo HG, Kim GG, Kim BS, Kim SG, Kim SC, et al. Introduction to Radiologic Science, cheong-gu. 2016:110-1.
  13. Baek CH. A Study of Scattered Radiation Effect on Digital Radiography Imaging System. Journal of the Korean Society of Radiology. 2017;40(1):71-5. https://doi.org/10.17946/JRST.2017.40.1.11
  14. Lee KH, Lee CH, Jin GH. Stationary and Moving Computed Radiography Grids: Comparative Observer's Perception. Journal of the Korean Society of Radiology. 2015;9(7):515-6. https://doi.org/10.7742/jksr.2015.9.7.515
  15. Seo WJ, Seo JB, Lee JW. Using Image J program, compared of focusing distance and grid rate. Journal of the Korean Society for Digital Imaging in Medicine. 2012;14(1):37-9.
  16. Lee IJ, Yeo YB, Lee TS. Entrance Skin Dose and Image Quality Evaluation According to Use Grid Radiography for the Extremity in FPD System. Journal of Radiological Science and Technology. 2010;33(4):343-6.
  17. Jeung H, Lee W, Kim MC. Digital Medical Imaging. Jeongmungak. 2003:174-5.