망막질환 치료를 위한 어플리케이터 허브와 캐뉼러 일체화 금형기술 및 접합강도 분석

Applicator parts hub and cannula integrated mold technology and bonding strength analysis for retinal disease treatment

  • 유정현 ((주)승우) ;
  • 김용대 (한국생산기술연구원 금형성형연구부문) ;
  • 이정원 (한국생산기술연구원 금형성형연구부문)
  • Jeong-Hyeon Yu (Seungwoo Co., Ltd) ;
  • Yong-Dae Kim (Molding & Metal Forming R&D Department, Korea Institute of Industrial Technology) ;
  • Jeong-Won Lee (Molding & Metal Forming R&D Department, Korea Institute of Industrial Technology)
  • 투고 : 2023.03.14
  • 심사 : 2023.03.31
  • 발행 : 2023.03.31

초록

Macular degeneration and glaucoma are representative age-related retinal diseases that rank second and third in the prevalence of retinal diseases, and are a kind of degenerative neurological disease. Irreversible visual acuity and visual field damage may occur, and the number of patients is rapidly increasing as the population ages. Since this retinal disease is a chronic disease, continuous drug treatment is required. There are various drug delivery methods for treatment, but direct injection of the drug into the intravitreal is the most effective for continuous delivery of the drug over a long period of time. In order to solidify Dexamethasone, a retinal disease treatment, and insert it into the primary intravitreal, it is important to develop a technology to miniaturize the treatment and an applicator to deliver the treatment. In this study, a mold technology was developed to integrate the cannula and hub, which are one part of applicator. In addition, surface treatment was performed on the outside of the cannula to improve the bonding strength between the cannula and the hub, and the bonding strength according to each condition was analyzed through a tensile test.

키워드

과제정보

본 연구는 중소벤처기업부 구매조건부신제품개발 사업(구매연계형) 지원으로 수행한 연구 결과임

참고문헌

  1. Oh, M.J., Lee, S.Y., 'Ocular Diseases and Hypertension", The Korean Journal of Medicine, Vol. 83, No. 5, 2012.
  2. Ryu, S.Y., Kim, D.W., Rim, T.H., Chung, E.J., Kim, J.W., "Prevalence of Exudative Age-related Macular Degeneration and Projections of the Cost of Ranibizumab in Korea" J Korean Ophthalmol Soc, Vol. 60, No. 3, pp. 253-260, 2019 https://doi.org/10.3341/jkos.2019.60.3.253
  3. Kim, J.H., Cheong, S.K., Kim, J., "Influence of Bonding Strength on Surface Pattern in Bonding of Carbon Fiber Reinforced Plastic and Metal", Journal of the Korean Society of Manufacturing Technology Engineers 26:4 pp. 430~435, 2017. https://doi.org/10.7735/ksmte.2017.26.4.430
  4. Lee, M.S., Kim, H.H., Kang, C.G., "Effect of Surface Roughness of Al5052/CFRP Composites on the Adhesion and Mechanical Properties", Composites Research, 26:5 pp. 295-302, 2013.sdsdsd 1151-1159, 2003. https://doi.org/10.7234/COMPOSRES.2013.26.5.295
  5. Rhee, K.Y., Yang, J.H., Choi, N.S., "A Study on the Curing Method to Improve Bonding Strength of Aluminum/CFRP Composites", International Journal of Automotive Technology, 10:3 pp. 130-135, 2002.
  6. Jung, Y.J., Kim, Y.S., Jeon, E.S., "Analysis of Variations in the Bonding Strength Characteristics of the AL6061-PBT-Polymer Composite with Injection Parameters", Journal of the Korean Society of Manufacturing Process Engineers, Vol. 20, No. 5, pp. 135~141, 2021. https://doi.org/10.14775/ksmpe.2021.20.05.135
  7. Park, H.P., Cha, B.S., Kim, S.G., Ko, Y.B., Lee, J.W., Jeon, Y.J., "A Study on the Effects in Adhesion of Liquid Silicone Rubbers in Mold Surface and PBT Insert at LSR/PBT Multi component Injection Molding", Korean Society for Precision Engineering, pp. 508-509, 2014.
  8. Lee, J.W., Son, M.K., "A Study on Gas Vent Control of Injection Mold for the Production of Precision Medical Device Parts", Korea Society of Die & Mold Engineering, Vol. 14, No. 2, 2020.
  9. .Lee, J.H., Lim, J.H., Han, Y.A., Shin, J.P., Cho, K.D., Lim, J.O., Kim, S.Y., "The Development of Biodegradable Intrascleral Implant for Slow Releasing of Triamcinolone Acetonide", J Korean Ophthalmol Soc, Vol. 46, No. 11, pp. 1894-1902, 2005.