DOI QR코드

DOI QR Code

Assessment of Validity and Reliability of Plantar Pressure in Smart Insole

스마트 인솔의 족저압 측정 결과에 대한 타당도 및 신뢰도 평가

  • Kang, Ho Won (Department of Orthopedic Surgery, Ewha Womans University Mokdong Hospital) ;
  • An, Yae Lynn (Department of Orthopedic Surgery, Seoul National University College of Medicine) ;
  • Kim, Dae-Yoo (Department of Orthopedic Surgery, Busan Paik Hospital) ;
  • Lee, Dong-Oh (Department of Orthopedic Surgery, SNU Seoul Hospital) ;
  • Park, Gil Young (Department of Orthopedic Surgery, Seoul National University Hospital) ;
  • Lee, Dong Yeon (Department of Orthopedic Surgery, Seoul National University College of Medicine)
  • Received : 2022.04.11
  • Accepted : 2022.08.17
  • Published : 2022.09.15

Abstract

Purpose: Smart insoles are wearable devices that are inserted into shoes. Smart insoles with built-in pressure and acceleration sensors can measure the plantar pressure, stride length, and walking speed. This study evaluated the validity and reliability of the plantar pressure measurements of smart insoles during walking on flat ground. Materials and Methods: Twenty one subjects were included in this study. After wearing smart insoles, I-SOL® (Gilon, Seongnam, Korea), the subjects walked a 10 m corridor six times at a rate of 100 steps/min, and the middle three steps, free from direction changes, were chosen for data analysis. The same protocol was repeated after wearing Pedar-X (Novel Corporation, Munich, Germany), an insoletype plantar pressure measurement equipment with proven validity. The average maximum pressure (Ppeak, kPa) and the time at which Ppeak appeared (Ptime, %stride) were calculated for each device. The validity of smart insoles was evaluated by using the interclass correlation coefficient (ICC) of Ppeak and Ptime between the two instruments, and Cronbach's alpha was obtained from the Ppeak values to evaluate the reliability. Results: The ICC of Ppeak was 0.651 (good) in the hallux, 0.744 (good) in the medial forefoot, 0.839 (excellent) in the lateral forefoot, and 0.854 (excellent) in the hindfoot. The ICC of Ptime showed 0.868 (excellent) in the hallux, 0.892 (excellent) in the medial forefoot, 0.721 (good) in the lateral forefoot, and 0.832 (excellent) in the hindfoot. All ICC values showed good or excellent results. The Cronbach's alpha of Ppeak measured in the smart insoles was 0.990 in the hallux, 0.961 in the medial forefoot, 0.973 in the lateral forefoot, and 0.995 in the hindfoot; all indicated excellent reliability in all areas. Conclusion: The plantar pressure measurements of smart insoles during walking on a flat ground showed validity compared to Pedar-X, and high reliability after repeated measurements.

Keywords

Acknowledgement

This work was supported by the Korea Medical Device Development Fund grant funded by the Korea government (the Ministry of Science and ICT, the Ministry of Trade, Industry, and Energy, the Ministry of Health & Welfare, the Ministry of Food and Drug Safety) (Project Number: 1711138420, KMDF_PR_20200901_0193).

References

  1. Fadahunsi KP, O'Connor S, Akinlua JT, Wark PA, Gallagher J, Carroll C, et al. Information quality frameworks for digital health technologies: systematic review. J Med Internet Res. 2021;23:e23479. doi: 10.2196/23479.
  2. Mickle KJ, Munro BJ, Lord SR, Menz HB, Steele JR. Foot pain, plantar pressures, and falls in older people: a prospective study. J Am Geriatr Soc. 2010;58:1936-40. doi: 10.1111/j.1532-5415.2010.03061.x.
  3. Kim EJ, Shin HS, Lee JH, Kyung MG, Yoo HJ, Yoo WJ, et al. Repeatability of a multi-segment foot model with a 15-marker set in normal children. Clin Orthop Surg. 2018;10:484-90. doi: 10.4055/cios.2018.10.4.484.
  4. Murphy DF, Beynnon BD, Michelson JD, Vacek PM. Efficacy of plantar loading parameters during gait in terms of reliability, variability, effect of gender and relationship between contact area and plantar pressure. Foot Ankle Int. 2005;26:171-9. doi: 10.1177/107110070502600210.
  5. Ramanathan AK, Kiran P, Arnold GP, Wang W, Abboud RJ. Repeatability of the Pedar-X in-shoe pressure measuring system. Foot Ankle Surg. 2010;16:70-3. doi: 10.1016/j.fas.2009.05.006.
  6. Hsiao H, Guan J, Weatherly M. Accuracy and precision of two in-shoe pressure measurement systems. Ergonomics. 2002;45:537-55. doi: 10.1080/00140130210136963.
  7. Putti AB, Arnold GP, Cochrane L, Abboud RJ. The Pedar in-shoe system: repeatability and normal pressure values. Gait Posture. 2007;25:401-5. doi: 10.1016/j.gaitpost.2006.05.010.
  8. Kernozek TW, LaMott EE, Dancisak MJ. Reliability of an in-shoe pressure measurement system during treadmill walking. Foot Ankle Int. 1996;17:204-9. doi: 10.1177/107110079601700404.
  9. Marshall SJ, Levy SS, Tudor-Locke CE, Kolkhorst FW, Wooten KM, Ji M, et al. Translating physical activity recommendations into a pedometer-based step goal: 3000 steps in 30 minutes. Am J Prev Med. 2009;36:410-5. doi: 10.1016/j.amepre.2009.01.021.
  10. Cicchetti DV, Sparrow SA. Developing criteria for establishing interrater reliability of specific items: applications to assessment of adaptive behavior. Am J Ment Defic. 1981;86:127-37.
  11. George D, Mallery P. SPSS for Windows step by step : a simple guide and reference 11.0 update. 4th ed. Boston: Allyn and Bacon; 2003.
  12. Marey EJ. De la locomotion terrestre chez les bipedes et les quadrupedes. [Paris]: Editeur inconnu; 1873. French.
  13. Salpavaara T, Verho J, Lekkala J, Halttunen J. Wireless insole sensor system for plantar force measurements during sport events. Paper presented at: IMEKO XIX World Congress on Fundamental and Applied Metrology; 2009 Sep 6-11; Lisbon, Portugal. p.2118-23.
  14. MacWilliams BA, Armstrong PF. Clinical applications of plantar pressure measurement in pediatric orthopedics. Paper presented at: Pediatric Gait, A new Millennium in Clinical Care and Motion Analysis Technology; 2000 Jul 22; Chicago, USA. p.143-50.
  15. Razak AH, Zayegh A, Begg RK, Wahab Y. Foot plantar pressure measurement system: a review. Sensors (Basel). 2012;12:9884-912. doi: 10.3390/s120709884.
  16. Gefen A. Pressure-sensing devices for assessment of soft tissue loading under bony prominences: technological concepts and clinical utilization. Wounds. 2007;19:350-62.
  17. Barnett S, Cunningham JL, West S. A comparison of vertical force and temporal parameters produced by an in-shoe pressure measuring system and a force platform. Clin Biomech (Bristol, Avon). 2001;16:353-7. doi: 10.1016/s0268-0033(01)00026-2.
  18. Frykberg RG, Gordon IL, Reyzelman AM, Cazzell SM, Fitzgerald RH, Rothenberg GM, et al. Feasibility and efficacy of a smart mat technology to predict development of diabetic plantar ulcers. Diabetes Care. 2017;40:973-80. doi: 10.2337/dc16-2294.
  19. Najafi B, Ron E, Enriquez A, Marin I, Razjouyan J, Armstrong DG. Smarter sole survival: will neuropathic patients at high risk for ulceration use a smart insole-based foot protection system? J Diabetes Sci Technol. 2017;11:702-13. doi: 10.1177/1932296816689105.
  20. Macdonald EM, Perrin BM, Kingsley MIC. Factors influencing Australian podiatrists' behavioural intentions to adopt a smart insole into clinical practice: a mixed methods study. J Foot Ankle Res. 2020;13:28. doi: 10.1186/s13047-020-00396-x.
  21. Chatzaki C, Skaramagkas V, Tachos N, Christodoulakis G, Maniadi E, Kefalopoulou Z, et al. The Smart-Insole dataset: gait analysis using wearable sensors with a focus on elderly and Parkinson's patients. Sensors (Basel). 2021;21:2821. doi: 10.3390/s21082821.
  22. Putti AB, Arnold GP, Abboud RJ. Foot pressure differences in men and women. Foot Ankle Surg. 2010;16:21-4. doi: 10.1016/j.fas.2009.03.005.
  23. Yamamoto T, Hoshino Y, Kanzaki N, Nukuto K, Yamashita T, Ibaraki K, et al. Plantar pressure sensors indicate women to have a significantly higher peak pressure on the hallux, toes, forefoot, and medial of the foot compared to men. J Foot Ankle Res. 2020;13:40. doi: 10.1186/s13047-020-00410-2.