DOI QR코드

DOI QR Code

주파수 출력 제어 최적화를 통한 고강도 집속 초음파 피부치료 시스템 개발 연구

A Study on the Development of High-Intensity Focused Ultrasound Skin Treatment System Through Frequency Output Control Optimization

  • Park, Jong-Cheol (Department of Electrical, Electronics & Information Communications Engineering, Graduate School, Tong Myong University) ;
  • Kim, Min-Sung (School of Artificial Intelligence, College of Artificial Intelligence Convergence, TongMyong University)
  • 투고 : 2022.08.15
  • 심사 : 2022.08.22
  • 발행 : 2022.08.31

초록

It is important to develop a transducer that generates uniform output power through frequency control of the HIFU at 4 MHz frequency for the high intensity focused ultrasound (HIFU) skin diseases treatment. In this paper, a 4 MHz frequency band HIFU system for skin disease treatment was designed, manufactured and developed. In HIFU, even for the ultrasonic vibrator in the 4 MHz frequency band, the characteristics of the output power of the HIFU are different depending on the difference in the thickness of the PZT material. Through the development of a system amplifier, the sound output of the HIFU transducer was improved to more than 48 W and uniform output power control was possible. And, it is possible to control the output power even in a frequency band of 4.0 to 4.7 MHz, which is wider than 4.0 MHz, and shows the resonance frequency of the transducer. The maximum output power for each frequency was 49.969 W and the minimum value was 48.018 W. The maximum output power compared to the minimum output power is 49.969 W, which is uniform within 4.1%. It was confirmed that the output power of the HIFU through the amplifier can be uniformly controlled in the 4 MHz frequency band.

키워드

과제정보

This research was supported by the TongMyong University Research Grants 2020(2020A030)

참고문헌

  1. M.L. Jewell, P.B. Fodor, E.B. de Souza Pinto, and M.A. Al Shammari, "Clinical Application of VASER-Assisted Lipoplasty: A Pilot Clinical Study," Aesthetic Surgery Journal, Vol. 22, No. 2, pp. 131-146, 2002. https://doi.org/10.1067/maj.2002.123377
  2. G.T. Haar and C. Coussios, "High Intensity Focused Ultrasound Physical Principles and Devices," International Journal of Hyperthermia, Vol. 23, No. 2, pp. 89-104, 2007. https://doi.org/10.1080/02656730601186138
  3. O. Lee, T. Kim, and H. Song, "Development of Electric Surgical Machine by LC Series Resonance," The Korean Institute of Electrical Engineers (KIEE), pp. 153-157, 2002.
  4. J. Lim and H. Jung, "Comparison of Use of Ultrasonic Surgical and Bipolar Coagulator System on a Laparoscopic Assisted Vaginal Hysterectomy," Korea Society of Obstetrics and Gynecology, pp. 1363-1366, 2003.
  5. K. Song, Y. Ko, and W. Lee, "Development of the Skin Treatment Unit Using Low Temperature Plasma," Journal of the Korea Institute of Electronic Communication Science, Vol. 9, No. 12, pp. 1427-1433, 2014. https://doi.org/10.13067/JKIECS.2014.9.12.1427
  6. J.E. Kennedy, G.R. Haar, and D. Cranston, "High Intensity Focused Ultrasound: Surgery of the Future?," The British Journal of Radiology, Vol. 76, No. 909, pp. 590-599, 2003. https://doi.org/10.1259/bjr/17150274
  7. G. Oni, R. Hoxworth, S. Teotia, S. Brown, and J.M. Kenkel, "Evaluation of a Microfocused Ultrasound System for Improving Skin Laxity and Tightening in the Lower Face," The American Society for Aesthetic Plastic Surgery, Vol. 34, No. 7, pp. 1099-1110, 2014. https://doi.org/10.1177/1090820X14541956
  8. M. Alam, L.E. White, N. Martin, J. Witherspoon, S. Yoo, and D.P. West, "Ultrasound Tightening of Facial and Neck Skin: A Rater-Blinded Prospective Cohort Study," Journal of the American Academy of Dermatology, Vol. 62, No. 2, pp. 262-269, 2010. https://doi.org/10.1016/j.jaad.2009.06.039
  9. K.H. Ahn and J.H. Kim, "Face Tracking System Using Updated Skin Color," Journal of Korea Multimedia Society, Vol. 18, No. 5, pp. 610-619, 2015. https://doi.org/10.9717/KMMS.2015.18.5.610