Validity of Three-dimensional Facial Scan Taken with Facial Scanner and Digital Photo Wrapping on the Cone-beam Computed Tomography: Comparison of Soft Tissue Parameters

  • Aljawad, Hussein (Department of Orthodontics, School of Dentistry, Chonnam National University) ;
  • Lee, Kyungmin Clara (Department of Orthodontics, School of Dentistry, Chonnam National University)
  • 투고 : 2022.01.09
  • 심사 : 2022.05.19
  • 발행 : 2022.06.30

초록

Purpose: The purpose of the study was to assess the validity of three-dimensional (3D) facial scan taken with facial scanner and digital photo wrapping on the cone-beam computed tomography (CBCT). Materials and Methods: Twenty-five patients had their CBCT scan, two-dimensional (2D) standardized frontal photographs and 3D facial scan obtained on the same day. The facial scans were taken with a facial scanner in an upright position. The 2D standardized frontal photographs were taken at a fixed distance from patients using a camera fixed to a cephalometric apparatus. The 2D integrated facial models were created using digital photo wrapping of frontal photographs on the corresponding CBCT images. The 3D integrated facial models were created using the integration process of 3D facial scans on the CBCT images. On the integrated facial models, sixteen soft tissue landmarks were identified, and the vertical, horizontal, oblique and angular distances between soft tissue landmarks were compared among the 2D facial models and 3D facial models, and CBCT images. Result: The results showed no significant differences of linear and angular measurements among CBCT images, 2D and 3D facial models except for Se-Sn vertical linear measurement which showed significant difference for the 3D facial models. The Bland-Altman plots showed that all measurements were within the limit of agreement. For 3D facial model, all Bland-Altman plots showed that systematic bias was less than 2.0 mm and 2.0° except for Se-Sn linear vertical measurement. For 2D facial model, the Bland-Altman plots of 6 out of 11 of the angular measurements showed systematic bias of more than 2.0°. Conclusion: The facial scan taken with facial scanner showed a clinically acceptable performance. The digital 2D photo wrapping has limitations in clinical use compared to 3D facial scans.

키워드

과제정보

This work was supported by the National Research Foundation of Korea (NRF) grant funded by the Korea government (Ministry of Science and ICT) (No. 2020R1F1A1070617).

참고문헌

  1. Kau CH, Richmond S, Incrapera A, English J, Xia JJ. Three-dimensional surface acquisition systems for the study of facial morphology and their application to maxillofacial surgery. Int J Med Robot. 2007; 3: 97-110. https://doi.org/10.1002/rcs.141
  2. Maal TJ, van Loon B, Plooij JM, Rangel F, Ettema AM, Borstlap WA, Berge SJ. Registration of 3-dimensional facial photographs for clinical use. J Oral Maxillofac Surg. 2010; 68: 2391-401. https://doi.org/10.1016/j.joms.2009.10.017
  3. Altorkat Y, Khambay BS, McDonald JP, Cross DL, Brocklebank LM, Ju X. Immediate effects of rapid maxillary expansion on the naso-maxillary facial soft tissue using 3D stereophotogrammetry. Surgeon. 2016; 14: 63-8. https://doi.org/10.1016/j.surge.2014.04.005
  4. van Loon B, Maal TJ, Plooij JM, Ingels KJ, Borstlap WA, Kuijpers-Jagtman AM, Spauwen PH, Berge SJ. 3D Stereophotogrammetric assessment of pre- and postoperative volumetric changes in the cleft lip and palate nose. Int J Oral Maxillofac Surg. 2010; 39: 534-40. https://doi.org/10.1016/j.ijom.2010.03.022
  5. Petrides G, Clark JR, Low H, Lovell N, Eviston TJ. Three-dimensional scanners for soft-tissue facial assessment in clinical practice. J Plast Reconstr Aesthet Surg. 2021; 74: 605-14. https://doi.org/10.1016/j.bjps.2020.08.050
  6. Gibelli D, Dolci C, Cappella A, Sforza C. Reliability of optical devices for three-dimensional facial anatomy description: a systematic review and metaanalysis. Int J Oral Maxillofac Surg. 2020; 49: 1092-106. https://doi.org/10.1016/j.ijom.2019.10.019
  7. Zhao YJ, Xiong YX, Wang Y. Three-dimensional accuracy of facial scan for facial deformities in clinics: a new evaluation method for facial scanner accuracy. PLoS One. 2017; 12: e0169402. https://doi.org/10.1371/journal.pone.0169402
  8. Premjani P, Al-Mulla AH, Ferguson DJ. Accuracy of 3D facial models obtained from CBCT volume wrapping. J Clin Orthod. 2015; 49: 641-6.
  9. Almulla S, Premjani P, Vaid NR, Fadia DF, Ferguson DJ. Evaluating the accuracy of facial models obtained from volume wrapping: 2D images on CBCT versus 3D on CBCT. Semin Orthod. 2018; 24: 443-50. https://doi.org/10.1053/j.sodo.2018.10.008
  10. Modabber A, Peters F, Kniha K, Goloborodko E, Ghassemi A, Lethaus B, Holzle F, Mohlhenrich SC. Evaluation of the accuracy of a mobile and a stationary system for three-dimensional facial scanning. J Craniomaxillofac Surg. 2016; 44: 1719-24. https://doi.org/10.1016/j.jcms.2016.08.008
  11. Ozsoy U, Demirel BM, Yildirim FB, Tosun O, Sarikcioglu L. Method selection in craniofacial measurements: advantages and disadvantages of 3D digitization method. J Craniomaxillofac Surg. 2009; 37: 285-90. https://doi.org/10.1016/j.jcms.2008.12.005
  12. Dindaroglu F, Kutlu P, Duran GS, Gorgulu S, Aslan E. Accuracy and reliability of 3D stereophotogrammetry: a comparison to direct anthropometry and 2D photogrammetry. Angle Orthod. 2016; 86: 487-94. https://doi.org/10.2319/041415-244.1
  13. Ma L, Xu T, Lin J. Validation of a three-dimensional facial scanning system based on structured light techniques. Comput Methods Programs Biomed. 2009; 94: 290-8. https://doi.org/10.1016/j.cmpb.2009.01.010
  14. Plooij JM, Swennen GR, Rangel FA, Maal TJ, Schutyser FA, Bronkhorst EM, Kuijpers-Jagtman AM, Berge SJ. Evaluation of reproducibility and reliability of 3D soft tissue analysis using 3D stereophotogrammetry. Int J Oral Maxillofac Surg. 2009; 38: 267-73. https://doi.org/10.1016/j.ijom.2008.12.009
  15. Heike CL, Upson K, Stuhaug E, Weinberg SM. 3D digital stereophotogrammetry: a practical guide to facial image acquisition. Head Face Med. 2010; 6: 18. https://doi.org/10.1186/1746-160X-6-18