DOI QR코드

DOI QR Code

Source and LVis based coincidence summing correction in HPGe gamma-ray spectrometry

  • Lee, Jieun (Research Center, Dongnam Institute of Radiological and Medical Sciences) ;
  • Kim, HyoJin (Research Center, Dongnam Institute of Radiological and Medical Sciences) ;
  • Kye, Yong Uk (Research Center, Dongnam Institute of Radiological and Medical Sciences) ;
  • Lee, Dong Yeon (Research Center, Dongnam Institute of Radiological and Medical Sciences) ;
  • Kim, Jeung Kee (Research Center, Dongnam Institute of Radiological and Medical Sciences) ;
  • Jo, Wol Soon (Research Center, Dongnam Institute of Radiological and Medical Sciences) ;
  • Kang, Yeong-Rok (Research Center, Dongnam Institute of Radiological and Medical Sciences)
  • Received : 2021.07.07
  • Accepted : 2021.11.07
  • Published : 2022.05.25

Abstract

The activity of gamma-ray emitting nuclides is calculated assuming that each gamma-ray is detected individually; thus, the magnitude of the coincidence summing signal must be considered during activity calculations. Here, the correction factor for the coincidence summing effect was calculated, and the detection efficiencies of two HPGe detectors were compared. The CANBERRA Inc. GC4018 high-purity Ge detector provided an estimate for the peak-to-total ratio using a point source to determine the coincidence summing correction factor. The ORTEC Inc. GEM60 high-purity Ge detector uses EFFTRAN in LVis to obtain the parameters of the detector and source model and the gamma-gamma and gamma-X match estimates, in order to determine the coincidence summing correction factor. Nuclide analyses, radioactivity comparisons, and analyses of reference material samples were performed utilizing certified reference materials to accurately determine the detection efficiencies. For both Co-60 and Y-88, the detection efficiency for a point source increased by an average of at least 12-13%, whereas the detection efficiency determined using LVis increased by an average of at least 13-15%. The calculated radioactivity values of the certified reference material and reference material samples were accurate to within 3% and 6% of the measured values, respectively.

Keywords

Acknowledgement

This work was supported by the Dongnam Institute of Radiological & Medical Sciences (DIRAMS) grant funded by the Korean Government (MSIT) (No. 50494-2021).

References

  1. G. Giubrone, J. Ortiz, S. Gallardo, S. Martorell, M.C. Bas, Calculation of coincidence summing correction factors for an HPGe detector using GEANT4, J. Environ. Radioact. 158-159 (2016) 114-118, https://doi.org/10.1016/j.jenvrad.2016.04.008.
  2. V.P. Kolotov, V.V. Atrashkevich, S.J. Gelsema, Estimation of true coincidence corrections for voluminous sources, J. Radioanal. Nucl. Chem. 210 (No. 1) (1996) 183-196, https://doi.org/10.1007/BF02055417.
  3. M. Blaauw, The use of sources emitting coincident g-rays for determination of absolute efficiency curves of highly efficient Ge detectors, Nucl. Instrum. Methods Phys. Res. A 332 (1993) 493-500, https://doi.org/10.1016/0168-9002(93)90305-2.
  4. S. Rizzo, E. Tomarchio, Numerical expressions for the computation of coincidence-summing correction factors in gamma-ray spectrometry with HPGe detectors, Appl. Radiat. Isot. 68 (2010) 555-560, https://doi.org/10.1016/j.apradiso.2009.10.024.
  5. S. Hurtado, M. Garcia-Leon, R. Garcia-Tenorio, GEAMT4 Code for simulation of a germanium gamma-ray detector and its application to efficiency calibration, Nucl. Instrum. Methods Phys. Res. A 518 (2004) 764-774, https://doi.org/10.1016/j.nima.2003.09.057.
  6. T. Vidmar, EFFTRAN-A Monte Carlo efficiency transfer code for gamma-ray spectrometry, Nucl. Instrum. Methods Phys. Res. A 550 (2005) 603-608, https://doi.org/10.1016/j.nima.2005.05.055.
  7. L.-C. He, L.-J. Diao, B.-H. Sun, L.-H. Zhu, J.-W. Zhao, M. Wang, K. Wang, Summing coincidence correction for γ-ray measurements using the HPGe detector with a low background shielding system, Nucl. Instrum. Methods Phys. Res. A 880 (2018) 22-27, https://doi.org/10.1016/j.nima.0017.09.043.
  8. M.S. Badawi, S.I. Jovanovic, A.A. Thabet, A.M. Ei-Khatib, A.D. Dlabac, B.A. Salem, M.M. Gouda, N.N. Mihaljevic, K.S. Almugren, M.I. Abbas, Calibration of 4π NaI(Tl) detectors with coincidence summing correction using new numerical procedure and ANGLE4 software, AIP Adv. 7 (2017), 035005, https://doi.org/10.1063/1.4978214.
  9. J. Kastlander, L.-E. De Geer, S. Jonsson, H. Rameback, Uncertainties in calculated correction factors for true coincidence-summing (TCS), Appl. Radiat. Isot. 122 (2017) 174-179, https://doi.org/10.1016/j.apradiso.2017.01.039.
  10. Genie Canberra, Customization Tools Manual, Canberra Industries, 2000.
  11. Ortec, LVis 2.7 Counting Laboratory Application Manager Software for Gamma Vision, Software User's Manual, ORTEC Part No. 931082.