DOI QR코드

DOI QR Code

A Study of Orthognathic Surgical Guides with Two-stage Split Path

2단 절개 형태를 가지는 악교정 수술 장치 연구

  • Min Uk, Kim (Medical Device Development Center, Daegu Gyeongbuk Medical Innovation Foundation) ;
  • Chung Hwan, Park (Medical Device Development Center, Daegu Gyeongbuk Medical Innovation Foundation) ;
  • Ji Hyoung, Rho (Medical Device Development Center, Daegu Gyeongbuk Medical Innovation Foundation) ;
  • Eui Sung, Jung (Medical Device Development Center, Daegu Gyeongbuk Medical Innovation Foundation) ;
  • Young Sang, Park (Medical Device Development Center, Daegu Gyeongbuk Medical Innovation Foundation) ;
  • Dong Guk, Kim (Medical Device Development Center, Daegu Gyeongbuk Medical Innovation Foundation) ;
  • Yohan, Seo (New Drug Development Center, Daegu Gyeongbuk Medical Innovation Foundation) ;
  • Young Jea, Woo (Medical Device Development Center, Daegu Gyeongbuk Medical Innovation Foundation) ;
  • Jong Min, Lee (Department of Radiology, School of Medicine, Kyungpook National University)
  • 김민욱 (대구경북첨단의료산업진흥재단 첨단의료기기개발지원센터) ;
  • 박충환 (대구경북첨단의료산업진흥재단 첨단의료기기개발지원센터) ;
  • 노지형 (대구경북첨단의료산업진흥재단 첨단의료기기개발지원센터) ;
  • 정의성 (대구경북첨단의료산업진흥재단 첨단의료기기개발지원센터) ;
  • 박영상 (대구경북첨단의료산업진흥재단 첨단의료기기개발지원센터) ;
  • 김동국 (대구경북첨단의료산업진흥재단 첨단의료기기개발지원센터) ;
  • 서요한 (대구경북첨단의료산업진흥재단 신약개발지원센터) ;
  • 우영재 (대구경북첨단의료산업진흥재단 첨단의료기기개발지원센터) ;
  • 이종민 (경북대학교 의과대학 영상의학교실)
  • Received : 2022.11.17
  • Accepted : 2022.11.23
  • Published : 2022.12.31

Abstract

In this study, the accuracy of the orthognathic surgical guides with single-stage split path was upgraded to realize orthognathic surgical guides with two-stage split path and simulated surgery was performed to verify its accuracy. As a result, the average error distance between the simulation model and the scan model was + 0.289 / - 0.468 mm (standard deviation 0.128), which was confirmed to be within ± 0.5 mm, which is a clinically acceptable level. Also, there was no significant difference compared with the average value of + 0.313 / - 0.456 mm (average standard deviation 0.106) of the conventional single-stage split path type device. It is judged that the use of this device can contribute to the reduction of surgical time and increase in accuracy since a separate finishing operation for bone preparation is unnecessary.

Keywords

Acknowledgement

본 연구는 대구경북첨단의료산업진흥재단 첨단의료기기개발지원센터의 지원을 받아 수행하였음.

References

  1. Hashemi S, Parvizi S, Baghbanijavid H, Tan A, Nematollahi M, Ramazani A, Fang N, Elahinia M. Computational modelling of process-structure-property-performance relationships in metal additive manufacturing: a review. International materials reviews. 2022. pp. 1-46. 
  2. Ministry of Food and Drug Safety. Understanding Medical Device 3D Printing Technology: Part I Medical Device 3D Printing Technology. 2018. 
  3. Schouman T, Rouch P, Imholz B, Fasel J, Courvoisier, D. Accuracy Evaluation of CAD/CAM Generated Splints in Orthognathic Surgery : A Cadaveric Study. Head & Face Medicine. 2015;11(1):11-24. 
  4. Rinaldo Z, Giorgio O, Carlo T, Gaetano I, Domenico D, Umberto Z, Ilenia B, Stefano B. Clinical Accuracy of Splintless Maxillary Positioning with Aid of CAD/CAM Fabricated Surgical Cutting Guides and Titanium Plates. J Coatings. 2022;12(10):147-157.  https://doi.org/10.3390/coatings12020147
  5. Kim MU, Kim YH, Kim YS. A Study of Orthognathic Surgical Splints Development Using 3D Convergence Technology. J Korean Society for Precision Engineering. 2019;36(3):301-309.  https://doi.org/10.7736/KSPE.2019.36.3.301
  6. http://snuh.org/health/nMedInfo/nView.do?category=DIS&medid=AA000517 Accessed on 7 Sep 2022. 
  7. http://www.yon.co.kr/bbs/content.php?co_id=bjkp03 Accessed on 12 Feb 2022. 
  8. http://www.lineupdent.co.kr/58/?pCode=58 Accessed on 17 Feb 2022. 
  9. Kang SH, Kim HJ, Park HW, and Lee SH. Maxillary Cutting Guide for Executing a Simulated Osteotomy and Removing the Bony Interference During Orthognathic Surgery. Journal of Medical Devices, 2015, pp. 44-50. 
  10. Kim YS, Hong SM, Jung YH, No GW. Computer Aided Design and Applications. Sigma Press. 2018. 
  11. Kim MU, Kim YH, Kim YS. A Study of Orthognathic Surgical Splints Development Using 3D Convergence Technology. J Korean Society for Precision Engineering. 2019;36(3):301-309.  https://doi.org/10.7736/kspe.2019.36.3.301
  12. https://www.hani.co.kr/arti/economy/biznews Accessed on 27 Dec 2021. 
  13. https://www.amc.seoul.kr/asan/healthinfo/management/managementDetail.do?managementId=360 Accessed on 17 Nov 2022. 
  14. Vale F, Scherzberg J, Cavaleiro J, Sanz D, Caramelo Fl. 3D Virtual Planning in Orthognathic Surgery and CAD/CAM Surgical Splints Generation in One Patient with Craniofacial Microsomia : A Case Report. Dental Press J Orthodontics. 2016;21(1):89-100.  https://doi.org/10.1590/2177-6709.21.1.089-100.oar
  15. Li B, Zhang L, Sun H, Yuan J, Shen S. G. A Novel Method of Computer Aided Orthognathic Surgery Using Individual CAD/CAM Templates: A Combination of Osteotomy and Repositioning Guides. British J Oral and Maxillofacial Surgery. 2013;51(8):239-244. 
  16. Kim MU, Kim YH, Kim YS. A Study of Orthognathic Surgical Splints Development Using 3D Convergence Technology. J Korean Society for Precision Engineering. 2019;36(3):301-309. https://doi.org/10.7736/kspe.2019.36.3.301