Effects of Exposure Dose Reduction by Optimization of Automatic Exposure Control Factors in Digital Radiographic Examination of Paranasal Sinus

부비동 디지털 엑스선검사에서 자동노출제어 조절인자의 최적화를 통한 조사선량 감소 효과

  • Jeong, Min-Gyu (Department of Radiology, ChungBuk National University Hospital) ;
  • Seoung, Youl-Hun (Department of Radiological Science, College of Health Medical Science, Cheongju University)
  • 정민규 (충북대학교병원 영상의학과) ;
  • 성열훈 (청주대학교 보건의료과학대학 방사선학과)
  • Received : 2021.06.15
  • Accepted : 2021.06.28
  • Published : 2021.06.30

Abstract

The purpose of this study was to reduce dose while maintaining image quality during digital radiographic examination of paranasal sinus by using the automatic exposure control (AEC) system. The tube voltage was set as six stages that increased by about 10 kVp to 70 kVp, 81 kVp, 90 kVp, 102 kVp, 109 kVp and 117 kVp. And then the AEC system conditions were consisted of 9 setting environments, that change mode of the sensitivity (S200, S400, S800) and the density (+2.5, 0, -2.5). We measured automatically exposed tube current (mAs) under 54 conditions with combined these, and assessed SNR and PSNR through the acquired images. In addition, four radiologists performed a qualitative assessment of the acquired images for each combination on a five-point scale of the Likert. As a result, the lowest dose and the highest values of SNR and PSNR in images with a qualitative assessment more than 4 point were the AEC control factors of 90 kVp, S800, D2.5. We applied this condition to the clinical trial, it showed an effect of 83.1% reduction in exposure radiation dose (mR). Therefore, AEC system could be used as dose reduction technology if it understood and used related regulatory factors and physical characteristics.

Keywords

References

  1. Chiu P, Chen J, Chang C, Wei J. The diagnostic value of sinus radiography in the evaluation of sinusitis. Journal of Chin J Radiology. 2010;35(2):143-8.
  2. Lee JS, Ko SJ, Kang SS, Kim JH, Kim DH, Kim CS. Evaluation of image quality using automatic exposure control & sensitivity in the digital chest image. Journal of The Korea Contents Society. 2013;13(8):275-83.
  3. Hamer OW, Sirlin CB, Strotzer M, Borisch I, Zorger N, Feuerbach S, et al. Chest radiography with a flat-panel detector: Image quality with dose reduction after copper filtration. radiology. 2005;237(2):691-700. https://doi.org/10.1148/radiol.2372041738
  4. ICRP. Statement on Tissue Reactions. Approved by the Commission on April. 2007.
  5. Hong SS, Kim HC. A study on dose and image quality according to X-ray photon detection method in digital radiography system. Journal of the Institute of Electronics and Information Engineers. 2013; 50(12):247-53. https://doi.org/10.5573/IEEK.2013.50.12.247
  6. Hwang JH, Lee KB. A study on the quantitative analysis method through the absorbed dose and the histogram in the performance evaluation of the detector according to the sensitivity change of Auto Exposure Control(AEC) in digital radiography. Journal of The Korea Contents Society. 2018;18(1):232-40.
  7. Seibert JA. Digital radiography: Image quality and radiation dose. Health Physics. 2008;95(5):586-98. https://doi.org/10.1097/01.HP.0000326338.14198.a2
  8. Schaefer-Prokop C, Neitzel U, Venema HW, Uffmann M, Prokop M. Digital chest radiography: An update on modern technology, dose containment and control of image quality. European Radiology. 2008;18(9):1818-30. https://doi.org/10.1007/s00330-008-0948-3
  9. Lee JS, Park HH. Evaluation on organ dose and image quality by changing kVp and ion chamber combination while taking digital chest lateral decubitus PA projection. Journal of The Korea Contents Society. 2015;15(1):316-23.
  10. Hawking N, Elmore A. Effects of AEC chamber selection on patient dose and image quality. Radiologic Technoloy. 2009;80(5):411-20.
  11. Kim SC, Lee HK, Lee YS, Cho JH. Evaluation of automatic exposure control system chamber for The dose optimization when examining pelvic in digital radiography. Journal of X-ray Science and Technology. 2015;23(3):321-30. https://doi.org/10.3233/XST-150492
  12. Lee JS, Kim CS. The additional filter and ion chamber sensor combination for reducing patient dose in digital chest X-ray projection. Journal of the Korean Society of Radiology. 2015;9(3):175-81. https://doi.org/10.7742/jksr.2015.9.3.175
  13. Kim JS, Kim SH, Jeon MC, Ju WH, Jeong MG, Kim MJ, et al. Development of indirect dosimetry by calculation method in the diagnostic X-ray equipment. Journal of Radiological Science and Technology. 2018;41(6):587-94. https://doi.org/10.17946/JRST.2018.41.6.587
  14. Jeong MG, Seoung YH. Effects of field configuration shielding area and changing of density and sensitivity on tube current and image quality in automatic exposure control system. Journal of Radiological Science and Technology. 2020;14(5):635-42.
  15. Lee JS, Choi BW, Kim SH, Kim YM, Shim JN, Ahn HS, et al. Effect of automatic exposure control marker with chest radiography in radiation reduction. Korean Society of Radiological Science. 2014;37(33):177-85.
  16. Hwang JH, Ahn CJ, Lee KB. A study on the dynamic range performance evaluation method of detector with variation of tube voltage and automatic exposure control in digital radiography: Focused on the dynamic step wedge and histogram evaluation. Journal of The Korea Contents Society. 2019;19(4):368-80.
  17. Lim SH, Seoung YH. Correlation analysis of control factors in automatic exposure control of digital radiography system based on fine contrast images. Journal of Radiological Science and Technology. 2021;44(1):1-8. https://doi.org/10.17946/JRST.2021.44.1.1
  18. Lee KJ, Kim MG, Lee JW, Kim HC. Research for the environmental optimization of dose and image quality in digital radiography. Journal of the Institute of Electronics and Information Engineers. 2013;50(2):203-9. https://doi.org/10.5573/IEEK.2013.50.2.203
  19. Huang W, Van Metter R, Yang CYJ, Yorkston J. Configuration of AEC kVp dependence for digital radiography systems. Progress in Biomedical Optics and Imaging Proceedings Series. 2007;6510(1):651017.
  20. Bowden L, Faulkner R, Clancy C, Gallagher A, Devine M, Gorman D, et al. Doses under Automatic Exposure Control (AEC) for direct digital radiographic (DDR) X-ray systems. Radiation Protection Dosimetry. 2011;147(1-2):210-4. https://doi.org/10.1093/rpd/ncr301