DOI QR코드

DOI QR Code

Investigation of the Characteristics of New, Uniform, Extremely Small Iron-Based Nanoparticles as T1 Contrast Agents for MRI

  • Young Ho So (Department of Radiology, SMG-SNU Boramae Medical Center) ;
  • Whal Lee (Department of Radiology, Seoul National University College of Medicine) ;
  • Eun-Ah Park (Department of Radiology, Seoul National University College of Medicine) ;
  • Pan Ki Kim (Department of Radiology and Research Institute of Radiological Science, Severance Hospital, Yonsei University College of Medicine)
  • Received : 2020.12.13
  • Accepted : 2021.05.10
  • Published : 2021.10.01

Abstract

Objective: The purpose of this study was to evaluate the magnetic resonance (MR) characteristics and applicability of new, uniform, extremely small iron-based nanoparticles (ESIONs) with 3-4-nm iron cores using contrast-enhanced magnetic resonance angiography (MRA). Materials and Methods: Seven types of ESIONs were used in phantom and animal experiments with 1.5T, 3T, and 4.7T scanners. The MR characteristics of the ESIONs were evaluated via phantom experiments. With the ESIONs selected by the phantom experiments, animal experiments were performed on eight rabbits. In the animal experiments, the in vivo kinetics and enhancement effect of the ESIONs were evaluated using half-diluted and non-diluted ESIONs. The between-group differences were assessed using a linear mixed model. A commercially available gadolinium-based contrast agent (GBCA) was used as a control. Results: All ESIONs showed a good T1 shortening effect and were applicable for MRA at 1.5T and 3T. The relaxivity ratio of the ESIONs increased with increasing magnetic field strength. In the animal experiments, the ESIONs showed peak signal intensity on the first-pass images and persistent vascular enhancement until 90 minutes. On the 1-week follow-up images, the ESIONs were nearly washed out from the vascular structures and organs. The peak signal intensity on the first-pass images showed no significant difference between the non-diluted ESIONs with 3-mm iron cores and GBCA (p = 1.000). On the 10-minutes post-contrast images, the non-diluted ESIONs showed a significantly higher signal intensity than did the GBCA (p < 0.001). Conclusion: In the phantom experiments, the ESIONs with 3-4-nm iron oxide cores showed a good T1 shortening effect at 1.5T and 3T. In the animal experiments, the ESIONs with 3-nm iron cores showed comparable enhancement on the first-pass images and superior enhancement effect on the delayed images compared to the commercially available GBCA at 3T.

Keywords

Acknowledgement

This study was supported by a grant from the Hanwha Chemical Corp. The extremely small-sized iron-based nanoparticles were supplied from Hanwha Chemical Corp. The assistance of Bong-sik Jeon, PhD, Eung-gyu Kim, PhD, and Wan-Jae Myeong, PhD in some of the experiments is gratefully acknowledged.

References

  1. Gozzi M, Amorico MG, Colopi S, Favali M, Gallo E, Torricelli P, et al. Peripheral arterial occlusive disease: role of MR angiography. Radiol Med 2006;111:225-237 
  2. Lim RP, Shapiro M, Wang EY, Law M, Babb JS, Rueff LE, et al. 3D time-resolved MR angiography (MRA) of the carotid arteries with time-resolved imaging with stochastic trajectories: comparison with 3D contrast-enhanced Bolus-Chase MRA and 3D time-of-flight MRA. AJNR Am J Neuroradiol 2008;29:1847-1854 
  3. Nael K, Fenchel M, Krishnam M, Finn JP, Laub G, Ruehm SG. 3.0 tesla high spatial resolution contrast-enhanced magnetic resonance angiography (CE-MRA) of the pulmonary circulation: initial experience with a 32-channel phased array coil using a high relaxivity contrast agent. Invest Radiol 2007;42:392-398 
  4. Prince MR. Gadolinium-enhanced MR aortography. Radiology 1994;191:155-164 
  5. Yucel EK. MR angiography for evaluation of abdominal aortic aneurysm: has the time come? Radiology 1994;192:321-323 
  6. Government of Canada. Gadolinium-containing contrast agents - Update on nephrogenic systemic fibrosis/nephrogenic fibrosing eermopathy (NSF/NFD) - Notice to hospitals. Healthycanadians.gc.ca Web site. http://healthycanadians.gc.ca/recall-alert-rappel-avis/hc-sc/2013/36711a-eng.php. Accessed August 15, 2018 
  7. U.S. Food and Drug Administration. FDA Drug Safety Communication: new warnings for using gadolinium-based contrast agents in patients with kidney dysfunction. Fda.gov Web site. https://www.fda.gov/Drugs/DrugSafety/ucm223966.htm. Accessed August 15, 2018 
  8. European Medicines Agency. Gadolinium-containing contrast agents. Ema.europa.eu Web site. http://www.ema.europa.eu/ema/index.jsp?curl=pages/medicines/human/referrals/Gadolinium-containing_contrast_agents/human_referral_000182.jsp. Accessed August 20, 2018 
  9. Khawaja AZ, Cassidy DB, Al Shakarchi J, McGrogan DG, Inston NG, Jones RG. Revisiting the risks of MRI with gadolinium based contrast agents-review of literature and guidelines. Insights Imaging 2015;6:553-558 
  10. Wang Y, Alkasab TK, Narin O, Nazarian RM, Kaewlai R, Kay J, et al. Incidence of nephrogenic systemic fibrosis after adoption of restrictive gadolinium-based contrast agent guidelines. Radiology 2011;260:105-111 
  11. Altun E, Martin DR, Wertman R, Lugo-Somolinos A, Fuller ER 3rd, Semelka RC. Nephrogenic systemic fibrosis: change in incidence following a switch in gadolinium agents and adoption of a gadolinium policy--report from two U.S. universities. Radiology 2009;253:689-696 
  12. Roch A, Muller RN. Theory of proton relaxation induced by superparamagnetic particles. J Chem Phys 1999;110:5403-5411 
  13. Antell H, Numminen J, Abo-Ramadan U, Niemela MR, Hernesniemi JA, Kangasniemi M. Optimization of high-resolution USPIO magnetic resonance imaging at 4.7 T using novel phantom with minimal structural interference. J Magn Reson Imaging 2010;32:1184-1196 
  14. Ling D, Park W, Park YI, Lee N, Li F, Song C, et al. Multiple-interaction ligands inspired by mussel adhesive protein: synthesis of highly stable and biocompatible nanoparticles. Angew Chem Int Ed Engl 2011;50:11360-11365 
  15. Hifumi H, Yamaoka S, Tanimoto A, Citterio D, Suzuki K. Gadolinium-based hybrid nanoparticles as a positive MR contrast agent. J Am Chem Soc 2006;128:15090-15091 
  16. Jun YW, Lee JH, Cheon J. Chemical design of nanoparticle probes for high-performance magnetic resonance imaging. Angew Chem Int Ed Engl 2008;47:5122-5135 
  17. Kim BH, Lee N, Kim H, An K, Park YI, Choi Y, et al. Large-scale synthesis of uniform and extremely small-sized iron oxide nanoparticles for high-resolution T1 magnetic resonance imaging contrast agents. J Am Chem Soc 2011;133:12624-12631 
  18. Hagberg GE, Scheffler K. Effect of r1 and r2 relaxivity of gadolinium-based contrast agents on the T1-weighted MR signal at increasing magnetic field strengths. Contrast Media Mol Imaging 2013;8:456-465 
  19. Na HB, Lee IS, Seo H, Park YI, Lee JH, Kim SW, et al. Versatile PEG-derivatized phosphine oxide ligands for water-dispersible metal oxide nanocrystals. Chem Commun (Camb) 2007;48:5167-5169 
  20. Pouliquen D, Le Jeune JJ, Perdrisot R, Ermias A, Jallet P. Iron oxide nanoparticles for use as an MRI contrast agent: pharmacokinetics and metabolism. Magn Reson Imaging 1991;9:275-283 
  21. Park EA, Lee W, So YH, Lee YS, Jeon BS, Choi KS, et al. Extremely small pseudoparamagnetic iron oxide nanoparticle as a novel blood pool T1 magnetic resonance contrast agent for 3 T whole-heart coronary angiography in canines: comparison with gadoterate meglumine. Invest Radiol 2017;52:128-133 
  22. Malikova H, Holesta M. Gadolinium contrast agents - are they really safe? J Vasc Access 2017;18:1-7 
  23. Schieda N, Blaichman JI, Costa AF, Glikstein R, Hurrell C, James M, et al. Gadolinium-based contrast agents in kidney disease: comprehensive review and clinical practice guideline issued by the Canadian association of radiologists. Can Assoc Radiol J 2018;69:136-150 
  24. Nandwana SB, Moreno CC, Osipow MT, Sekhar A, Cox KL. Gadobenate dimeglumine administration and nephrogenic systemic fibrosis: is there a real risk in patients with impaired renal function? Radiology 2015;276:741-747 
  25. Janus N, Launay-Vacher V, Karie S, Clement O, Ledneva E, Frances C, et al. Prevalence of nephrogenic systemic fibrosis in renal insufficiency patients: results of the FINEST study. Eur J Radiol 2010;73:357-359 
  26. Deray G, Rouviere O, Bacigalupo L, Maes B, Hannedouche T, Vrtovsnik F, et al. Safety of meglumine gadoterate (Gd-DOTA)-enhanced MRI compared to unenhanced MRI in patients with chronic kidney disease (RESCUE study). Eur Radiol 2013;23:1250-1259 
  27. Amet S, Launay-Vacher V, Clement O, Frances C, Tricotel A, Stengel B, et al. Incidence of nephrogenic systemic fibrosis in patients undergoing dialysis after contrast-enhanced magnetic resonance imaging with gadolinium-based contrast agents: the Prospective Fibrose Nephrogenique Systemique study. Invest Radiol 2014;49:109-115 
  28. Soyer P, Dohan A, Patkar D, Gottschalk A. Observational study on the safety profile of gadoterate meglumine in 35,499 patients: The SECURE study. J Magn Reson Imaging 2017;45:988-997 
  29. Schieda N, Maralani PJ, Hurrell C, Tsampalieros AK, Hiremath S. Updated clinical practice guideline on use of gadolinium-based contrast agents in kidney disease issued by the Canadian Association of Radiologists. Can Assoc Radiol J 2019;70:226-232 
  30. Attari H, Cao Y, Elmholdt TR, Zhao Y, Prince MR. A systematic review of 639 patients with biopsy-confirmed nephrogenic systemic fibrosis. Radiology 2019;292:376-386 
  31. Liu G, Gao J, Ai H, Chen X. Applications and potential toxicity of magnetic iron oxide nanoparticles. Small 2013;9:1533-1545 
  32. Storey P, Lim RP, Chandarana H, Rosenkrantz AB, Kim D, Stoffel DR, et al. MRI assessment of hepatic iron clearance rates after USPIO administration in healthy adults. Invest Radiol 2012;47:717-724 
  33. Sharifi S, Behzadi S, Laurent S, Forrest ML, Stroeve P, Mahmoudi M. Toxicity of nanomaterials. Chem Soc Rev 2012;41:2323-2343