Pull-off resistance of a screwless implant-abutment connection and surface evaluation after cyclic loading

  • Alevizakos, Vasilios (Department of CAD/CAM and Digital Technologies in Dentistry, Danube Private University) ;
  • Mosch, Richard (Department of CAD/CAM and Digital Technologies in Dentistry, Danube Private University) ;
  • Mitov, Gergo (Department of Prosthodontics and Biomaterials, Danube Private University) ;
  • Othman, Ahmed (Department of CAD/CAM and Digital Technologies in Dentistry, Danube Private University) ;
  • See, Constantin von (Department of CAD/CAM and Digital Technologies in Dentistry, Danube Private University)
  • Received : 2020.10.03
  • Accepted : 2021.05.28
  • Published : 2021.06.30

Abstract

Purpose. The aim of this study was to investigate to what extent cyclic load affects the screwless implant-abutment connection for Morse taper dental implants. Materials and Methods. 16 implants (SICvantage max) and 16 abutments (Swiss Cross) were used. The screwless implant-abutment connection was subjected to 10,000 cycles of axial loading with a maximum force of 120 N. For the pull-off testing, before and after the same cyclic loading, the required force for disconnecting the remaining 6 implant-abutment connections was measured. The surface of 10 abutments was examined using a scanning electron microscope 120× before and after loading. Results. The pull-off test showed a significant decrease in the vertical force required to pull the abutment from the implant with mean 229.39 N ± 18.23 before loading, and 204.30 N ± 13.51 after loading (P<.01). Apart from the appearance of polished surface areas and slight signs of wear, no visible damages were found on the abutments. Conclusion. The deformation on the polished abutment surface might represent the result of micro movements within the implant-abutment connection during loading. Although there was a decrease of the pull-off force values after cyclic loading, this might not have a notable effect on the clinical performance.

Keywords

References

  1. Barrachina-Diez JM, Tashkandi E, Stampf S, Att W. Long-term outcome of one-piece implants. Part II: Prosthetic outcomes. A systematic literature review with meta-analysis. Int J Oral Maxillofac Implants 2013;28:1470-82. https://doi.org/10.11607/jomi.2945
  2. El-Sheikh MAY, Mostafa TMN, El-Sheikh MM. Effect of different angulations and collar lengths of conical hybrid implant abutment on screw loosening after dynamic cyclic loading. Int J Implant Dent 2018;4:39. https://doi.org/10.1186/s40729-018-0149-z
  3. Lauritano D, Moreo G, Lucchese A, Viganoni C, Limongelli L, Carinci F. The impact of implant-abutment connection on clinical outcomes and microbial colonization: a narrative review. Materials (Basel) 2020; 13:1131. https://doi.org/10.3390/ma13051131
  4. Maeda Y, Satoh T, Sogo M. In vitro differences of stress concentrations for internal and external hex implant-abutment connections: a short communication. J Oral Rehabil 2006;33:75-8. https://doi.org/10.1111/j.1365-2842.2006.01545.x
  5. Gehrke SA, Pereira Fde A. Changes in the abutment-implant interface in Morse taper implant connections after mechanical cycling: a pilot study. Int J Oral Maxillofac Implants 2014;29:791-7. https://doi.org/10.11607/jomi.3113
  6. Lopes PA, Carreiro AFP, Nascimento RM, Vahey BR, Henriques B, Souza JCM. Physicochemical and microscopic characterization of implant-abutment joints. Eur J Dent 2018;12:100-4. https://doi.org/10.4103/ejd.ejd_3_17
  7. Prado AM, Pereira J, Silva FS, Henriques B, Nascimento RM, Benfatti CAM, Lopez-Lopez J, Souza JCM. Wear of Morse taper and external hexagon implant joints after abutment removal. J Mater Sci Mater Med 2017;28:65. https://doi.org/10.1007/s10856-017-5879-6
  8. Geckili E, Geckili O, Bilhan H, Kutay O, Bilgin T. Clinical comparison of screw-retained and screwless morse taper implant-abutment connections: one-year post-loading results. Int J Oral Maxillofac Implants 2017; 32:1123-31. https://doi.org/10.11607/jomi.5314
  9. Urdaneta RA, Marincola M, Weed M, Chuang SK. A screwless and cementless technique for the restoration of single-tooth implants: a retrospective cohort study. J Prosthodont 2008;17:562-71. https://doi.org/10.1111/j.1532-849X.2008.00343.x
  10. Ugurel CS, Steiner M, Isik-Ozkol G, Kutay O, Kern M. Mechanical resistance of screwless morse taper and screw-retained implant-abutment connections. Clin Oral Implants Res 2015;26:137-42. https://doi.org/10.1111/clr.12303
  11. Bressan E, Lops D, Tomasi C, Ricci S, Stocchero M, Carniel EL. Experimental and computational investigation of Morse taper conometric system reliability for the definition of fixed connections between dental implants and prostheses. Proc Inst Mech Eng H 2014; 228:674-81.
  12. Moris IC, Faria AC, Ribeiro RF, Rodrigues RC. Torque loss of different abutment sizes before and after cyclic loading. Int J Oral Maxillofac Implants 2015;30:1256-61. https://doi.org/10.11607/jomi.3868
  13. Wang TM, Lee MS, Wang JS, Lin LD. The effect of implant design and bone quality on insertion torque, resonance frequency analysis, and insertion energy during implant placement in low or low- to medium-density bone. Int J Prosthodont 2015;28:40-7. https://doi.org/10.11607/ijp.4063
  14. Ettinger S, Schwarze M, Yao D, Ettinger M, Claassen L, Stukenborg-Colsman C, Thermann H, Plaass C. Stability of supramalleolar osteotomies using different implants in a sawbone model. Arch Orthop Trauma Surg 2018;138:1359-63. https://doi.org/10.1007/s00402-018-2981-2
  15. Bischel OE, Nadorf J, Klein SB, Gantz S, Jakubowitz E, Kretzer JP, Arnholdt J, Seeger JB. Modular tumor prostheses: are current stem designs suitable for distal femoral reconstruction? A biomechanical implant stability analysis in Sawbones. Arch Orthop Trauma Surg 2019;139:843-9. https://doi.org/10.1007/s00402-019-03158-y
  16. Danza M, Carinci F. Flapless surgery and immediately loaded implants: a retrospective comparison between implantation with and without computer-assisted planned surgical stent. Stomatologija 2010;12:35-41.
  17. Hellwig E, Schafer E, Klimek J, Attin T. Introduction to tooth preservation. 6th ed. Collogne: Deutscher Zahnarzte Verlag; 2013.
  18. Mangano F, Bakaj R, Frezzato I, Frezzato A, Montini S, Mangano C. Morse taper connection implants placed in grafted sinuses in 65 patients: a retrospective clinical study with 10 years of follow-up. Int J Dent 2017; 2017:4573037.
  19. Mangano F, Lucchina AG, Brucoli M, Migliario M, Mortellaro C, Mangano C. Prosthetic complications affecting single-tooth morse-taper connection implants. J Craniofac Surg 2018;29:2255-62. https://doi.org/10.1097/scs.0000000000004495
  20. Urdaneta RA, Marincola M. The integrated abutment crown, a screwless and cementless restoration for single-tooth implants: a report on a new technique. J Prosthodont 2007;16:311-8. https://doi.org/10.1111/j.1532-849x.2007.00194.x
  21. Arshad M, Mahgoli HA, Rasouli K, Refoua S. A novel technique to retrieve cold-welded implant abutment: a case series. Clin Case Rep 2019;7:1854-7. https://doi.org/10.1002/ccr3.2375
  22. Norton MR. Assessment of cold welding properties of the internal conical interface of two commercially available implant systems. J Prosthet Dent 1999;81: 159-66. https://doi.org/10.1016/S0022-3913(99)70243-X
  23. Merz BR, Hunenbart S, Belser UC. Mechanics of the implant-abutment connection: an 8-degree taper compared to a butt joint connection. Int J Oral Maxillofac Implants 2000;15:519-26.
  24. Bozkaya D, Muftu S. Mechanics of the tapered interference fit in dental implants. J Biomech 2003;36:1649-58. https://doi.org/10.1016/S0021-9290(03)00177-5
  25. Pintinha M, Camarini ET, Sabio S, Pereira JR. Effect of mechanical loading on the removal torque of different types of tapered connection abutments for dental implants. J Prosthet Dent 2013;110:383-8. https://doi.org/10.1016/j.prosdent.2013.06.007
  26. Hsu PF, Yao KT, Kao HC, Hsu ML. Effects of axial loading on the pull-out force of conical connection abutments in ankylos implant. Int J Oral Maxillofac Implants 2018;33:788-94. https://doi.org/10.11607/jomi.6016