DOI QR코드

DOI QR Code

Effect of fluid contamination on reverse torque values in implant-abutment connections under oral conditions

  • Mostafavi, Azam Sadat (Department of Prosthodontics, School of Dentistry, Tehran University of Medical Sciences) ;
  • Memarian, Maryam (Department of Prosthodontics, School of Dentistry, Tehran University of Medical Sciences) ;
  • Seddigh, Mohammad Ali (Department of Prosthodontics, School of Dentistry, Tehran University of Medical Sciences)
  • 투고 : 2020.09.15
  • 심사 : 2020.12.30
  • 발행 : 2021.02.26

초록

Purpose. Implant mechanical complications, including screw loosening, can influence dental implant success. It has been shown that torque values are affected by contamination occurred in implant-abutment (I/A) interface. This study aimed to examine the effects of blood, saliva, fluoride and chlorhexidine contamination on reverse torque values (RTVs) of abutment screws in oral conditions. Materials and Methods. 50 fixtures were mounted into the stainless-steel holders and divided into five groups (n = 10). Except control group (NC), fixture screw holes in other groups were contaminated with chlorhexidine (CG), saliva (SG), blood (BG), or fluoride (FG). Abutment screws were tightened with a digital torque meter. I/A assemblies were subjected to thermocycling and cyclic loading. The mean RTVs were recorded and data were analyzed with one-way ANOVA and Tukey test. Results. Except for specimens in SG (20.56 ± 1.33), other specimens in BG (21.11 ± 1.54), CG (22.89 ± 1.1) and FG (24.00 ± 1.12) displayed significantly higher RTVs compared to NC (19.00 ± 1.87). The highest RTVs were detected in CG and FG. Conclusion. The obtained data robustly suggest that RTVs were significantly affected by fluid contaminations. Specimens in FG and CG displayed the highest RTVs. Therefore, clinicians should have enough knowledge about probable contaminations in I/A interface in order to manage them during clinical procedure and to inform patients about using oral care products.

키워드

참고문헌

  1. Ali Z, Baker SR, Shahrbaf S, Martin N, Vettore MV. Oral health-related quality of life after prosthodontic treatment for patients with partial edentulism: a systematic review and meta-analysis. J Prosthet Dent 2019; 121:59-68. https://doi.org/10.1016/j.prosdent.2018.03.003
  2. Sargozaie N, Moeintaghavi A, Shojaie H. Comparing the quality of life of patients requesting dental implants before and after implant. Open Dent J 2017;11: 485-91. https://doi.org/10.2174/1874210601711010485
  3. Goodacre CJ, Kan JY, Rungcharassaeng K. Clinical complications of osseointegrated implants. J Prosthet Dent 1999;81:537-52. https://doi.org/10.1016/S0022-3913(99)70208-8
  4. Lee HW, Alkumru H, Ganss B, Lai JY, Ramp LC, Liu PR. The effect of contamination of implant screws on reverse torque. Int J Oral Maxillofac Implants. 2015;30: 1054-60. https://doi.org/10.11607/jomi.3976
  5. Tzenakis GK, Nagy WW, Fournelle RA, Dhuru VB. The effect of repeated torque and salivary contamination on the preload of slotted gold implant prosthetic screws. J Prosthet Dent 2002;88:183-91. https://doi.org/10.1067/mpr.2002.127604
  6. Tan KB, Nicholls JI. The effect of 3 torque delivery systems on gold screw preload at the gold cylinder-abutment screw joint. Int J Oral Maxillofac Implants 2002; 17:175-83.
  7. Guda T, Ross TA, Lang LA, Millwater HR. Probabilistic analysis of preload in the abutment screw of a dental implant complex. J Prosthet Dent 2008;100:183-93. https://doi.org/10.1016/S0022-3913(08)60177-8
  8. Jorn D, Kohorst P, Besdo S, Rucker M, Stiesch M, Borchers L. Influence of lubricant on screw preload and stresses in a finite element model for a dental implant. J Prosthet Dent 2014;112:340-8. https://doi.org/10.1016/j.prosdent.2013.10.016
  9. Guzaitis KL, Knoernschild KL, Viana MA. Effect of repeated screw joint closing and opening cycles on implant prosthetic screw reverse torque and implant and screw thread morphology. J Prosthet Dent 2011;106:159-69. https://doi.org/10.1016/S0022-3913(11)60115-7
  10. Sahin C, Ayyildiz S. Correlation between microleakage and screw loosening at implant-abutment connection. J Adv Prosthodont 2014;6:35-8. https://doi.org/10.4047/jap.2014.6.1.35
  11. Cardoso M, Torres MF, Lourenço EJ, de Moraes Telles D, Rodrigues RC, Ribeiro RF. Torque removal evaluation of prosthetic screws after tightening and loosening cycles: an in vitro study. Clin Oral Implants Res 2012; 23:475-80. https://doi.org/10.1111/j.1600-0501.2011.02165.x
  12. Al Rafee MA, Nagy WW, Fournelle RA, Dhuru VB, Tzenakis GK, Pechous CE. The effect of repeated torque on the ultimate tensile strength of slotted gold prosthetic screws. J Prosthet Dent 2002;88:176-82. https://doi.org/10.1067/mpr.2002.127652
  13. Duarte AR, Neto JP, Souza JC, Bonachela WC. Detorque evaluation of dental abutment screws after immersion in a fluoridated artificial saliva solution. J Prosthodont 2013;22:275-81. https://doi.org/10.1111/j.1532-849X.2012.00941.x
  14. Norton MR. Assessment of cold welding properties of the internal conical interface of two commercially available implant systems. J Prosthet Dent 1999;81: 159-66. https://doi.org/10.1016/S0022-3913(99)70243-X
  15. Ghanbarzadeh J, Dashti H, Abasi M, Nakhaei M. Torque removal evaluation of one-piece and twopiece abutment screws under dry and wet conditions. J Mashad Dent Sch 2015;39:89-98.
  16. Gumus HO, Zortuk M, Albayrak H, Dincel M, Kocaagaoglu HH, Kilinc HI. Effect of fluid contamination on reverse torque values in bone-level implants. Implant Dent 2014;23:582-7.
  17. Koosha S, Toraji S, Mostafavi AS. Effect of fluid contamination on the reverse torque values of abutment screws at implant-abutment connections. J Prosthet Dent 2020;123:618-21. https://doi.org/10.1016/j.prosdent.2019.05.001
  18. Micarelli C, Canullo L, Baldissara P, Clementini M. Implant abutment screw reverse torque values before and after plasma cleaning. Int J Prosthodont 2013;26: 331-3. https://doi.org/10.11607/ijp.3396
  19. Nigro F, Sendyk CL, Francischone CE Jr, Francischone CE. Removal torque of zirconia abutment screws under dry and wet conditions. Braz Dent J 2010;21:225-8. https://doi.org/10.1590/S0103-64402010000300009
  20. Jalali H, Moslemion N, Nokar S. Effect of blood and saliva contamination of the implant-abutment interface on the amount of preload. Dentistry Adv Res 2018;3:1-7.
  21. Kozlovsky A, Artzi Z, Moses O, Kamin-Belsky N, Greenstein RB. Interaction of chlorhexidine with smooth and rough types of titanium surfaces. J Periodontol 2006;77:1194-200. https://doi.org/10.1902/jop.2006.050401
  22. Zavanelli RA, Henriques GEP, Ferreira I, de Almeida Rollo JM. Corrosion-fatigue life of commercially pure titanium and Ti-6Al-4V alloys in different storage environments. J Prosthet Dent 2000;84:274-9. https://doi.org/10.1067/mpr.2000.108758
  23. Souza J, Barbosa S, Ariza E, Celis J-P, Rocha L. Simultaneous degradation by corrosion and wear of titanium in artificial saliva containing fluorides. Wear 2012; 292-3:82-8.