DOI QR코드

DOI QR Code

유헬스케어 패치형 무선 체온 모니터링 시스템 구현

U-Healthcare Patch Type Wireless Body Temperature Monitoring System

  • 박영상 ((재)대구경북첨단의료산업진흥재단 첨단의료기기개발지원센터) ;
  • 권오언 ((재)대구경북첨단의료산업진흥재단 첨단의료기기개발지원센터) ;
  • 조현성 ((재)대구경북첨단의료산업진흥재단 첨단의료기기개발지원센터) ;
  • 손재범 ((재)대구경북첨단의료산업진흥재단 첨단의료기기개발지원센터)
  • Park, Young-Sang (Medical Device Development Center, Daegu-Gyeongbuk Medical Innovation Foundation) ;
  • Kwon, Oeon (Medical Device Development Center, Daegu-Gyeongbuk Medical Innovation Foundation) ;
  • Cho, Hyun-Sung (Medical Device Development Center, Daegu-Gyeongbuk Medical Innovation Foundation) ;
  • Son, Jaebum (Medical Device Development Center, Daegu-Gyeongbuk Medical Innovation Foundation)
  • 투고 : 2019.10.15
  • 심사 : 2020.02.20
  • 발행 : 2020.02.29

초록

Recently, there have been many research on fever management using u-healthcare technology. Especially, fever of infants requires continuous monitoring of body temperature by parents. For infants between 4 weeks and under 5 years old, it is recommended to use an electronic thermometer or chemical thermometer in the axilla, or to use an infrared thermometer. However, in order to overcome the reality of not being able to waste significant time on continuous monitoring, there have been demands of patch type thermometers with the internet of things (IoT) and wireless communication technologies. Existing IoT thermometers are difficult to attach to infants' body because they do not take into account its size, and their interoperability is not guaranteed because they do not comply with standards in communication. Therefore, in this study, a patch-type thermometer with a diameter of 20 mm and a weight of 2.9 g was developed to manage the fever of infants, while it communicates wirelessly with Bluetooth Low Energy (BLE) communication protocol and complies with IEEE 11073 PHD(Personal Health Device) at the same time. We verified its performance under the requirements of thermometers regulated by the Korean Ministry of Food and Drug Safety.

키워드

참고문헌

  1. Jeong YS, Kim JS. Fever and Fever Management in Children: A Literature Review. Child Health Nursing Research. 2010;16(1):30-40. https://doi.org/10.4094/jkachn.2010.16.1.30
  2. Fever in under 5s: assessment and initial management. National Institute for Health and Care Excellence; 2019. pp. 1-35.
  3. Yun KW, Lim IS. A study for accuracy and usefulness of tympanic membrane and forehead thermometers. Korean J Pediatr. 2005;48(8):820-25.
  4. Lee YD, Sohn J, Ko JW, Kim JY, Ahn DH. Lee EK. A Study of Reliability of Tympanic Temperature Measurement Using Infrared Thermometer (Home Model) in Neonates. J Korean Soc Neonatol. 1999;6(1):116-20.
  5. Park SH. Reliability and Accuracy of Infrared Temperature: A Systematic Review. Korean Journal of Adult Nursing. 2014;26(6):668-80. https://doi.org/10.7475/kjan.2014.26.6.668
  6. Temperature measurement in paediatrics. Paediatr Child Health. 2000;5(5). pp273-276.
  7. Shin EJ, Kim HK. Analytical Study on Developing IoT Healthcare Device for Infants and Toddlers. Asia-pacific Journal of Multimedia Services Convergent with Art, Humanities, and Sociology. 2018;8(2):41-50.
  8. The Magazine of the IEIE. The Institute of Electronics and Information Engineers; 2018;45(11) pp. 1-98.
  9. Kim HJ, Yang HH. Patch Type Body Temperature Measurement System for Ubiquitous Healthcare. Journal of the Korea Institute of Information and Communication Engineering. 2011;15(7):1628-34. https://doi.org/10.6109/jkiice.2011.15.7.1628
  10. Woo SH, Park SY, Din Z, Won CH, Lee JH, Park HJ, Lee JW, Hong YG, Suh JH, Cho JH. Design and implement of patch type wireless skin temperature measuring system. Journal of Sensor Science and Technology. 2008;17(5):350-60. https://doi.org/10.5369/JSST.2008.17.5.350
  11. Sohng KY, Kang SS, Hwang JS, Kim MJ. Accuracy of Temperature Measurements, Nursing Time for Measuring Temperature and the Validity of Fever Detection. Journal of Korean Academy of Fundamentals of Nursing. 1998;5(1):33-45.
  12. Myung KD, Park SW, Jeong G, Lee HC, Yoon SY, Shin SM. Appropriate Body Temperature Measurement Method for Parents of Infants. Journal of The Korean Society of Emergency Medicine. 2016;27(5):458-63.
  13. Electronic thermometer. Medical device standard. Korea: Ministry of Food and Drug Safety; 2019. pp. 2012-2055.
  14. Park HN, Kim SH, Yoo DS. Present status and analysis for IEEE 11073 personal health device specializations. The Journal of Korean Institute of Communications and Information Sciences. 2012;37(6C):469-75. https://doi.org/10.7840/KICS.2012.37.6C.469
  15. ISO/IEEE 11073-10408:2010(E), Health informatics - Personal health device communication - Part 10408: Device specialization - Thermometer. 2010. pp. 1-48.
  16. Chun SM, Nah JW, Park JT. Design and Implementation of IEEE 11073/HL7 Translation Gateway Based on U-Healthcare Application Service for M2M. The Journal of Korean Institute of Communications and Information Sciences. 2011;36(3B):275-86. https://doi.org/10.7840/KICS.2011.36B.3.275
  17. U-Healthcare Medical Device System License Review Guidelines. Korea: Ministry of Food and Drug Safety; 2016. pp. 1-17.