DOI QR코드

DOI QR Code

휴대용 인큐베이터의 개발에 관한 연구

Study on Development of Portable Incubator

  • Eizad, Amre (School of Mechanical & Aerospace Engineering, Gyeongsang National University) ;
  • Zahra, Falak (Department of Mechatronics Engineering, Air University) ;
  • Alam, Hamza (Department of Mechatronics Engineering, Air University) ;
  • Tahir, Hassan (Department of Mechatronics Engineering, Air University) ;
  • Bangash, Afrasiab Khan (Department of Mechatronics Engineering, Air University) ;
  • Lyu, Sung-Ki (School of Mechanical & Aerospace Engineering, Gyeongsang National University)
  • 투고 : 2019.05.12
  • 심사 : 2019.05.29
  • 발행 : 2019.09.30

초록

Preterm children require a controlled environment that is as close as possible to that inside the womb. Incubators are well equipped to fulfill this requirement; however, they are cumbersome and expensive, thereby restricting their portability and availability in less developed and rural areas. This research comprises the development and system validation of a portable incubator. The system consists of a collapsible baby enclosure that can be stowed inside the system base when not in use. The enclosure is made from acrylic such that it is easy to clean and allows unhindered visual observation of the occupant while being robust enough to withstand transit conditions. The system can be powered either by a mains supply or a 12-VDC automobile power supply. Additionally, it has an onboard battery to ensure a continuous supply during transit. A Peltier plate controlled using a microcontroller ensures the desired enclosure temperature irrespective of the ambient temperature. Built-in sensor probes can measure the skin temperature, pulse rate, blood oxygenation level, and ECG of the infant and display them on the system screen. The system function is validated by testing its peak power consumption and the heating and cooling performances of the environment control system.

키워드

참고문헌

  1. Goldenberg, R. L., Rouse, D.J., "Prevention of premature birth," New England Journal of Medicine, Vol. 339, No. 5, pp. 313-320, 1998. https://doi.org/10.1056/NEJM199807303390506
  2. Blencowe, H., Cousens, S., Oestergaard, M. Z., Chou, D., Moller, A. B., Narwal, R., Adler, A., Garcia, C. V., Rohde, S., Say, L., Lawn, J. E., "National, regional, and worldwide estimates of preterm birth rates in the year 2010 with time trends since 1990 for selected countries: a systematic analysis and implications," The lancet, Vol. 379, No. 9832, pp. 2162-2172, 2012. https://doi.org/10.1016/S0140-6736(12)60820-4
  3. Liu, L., Oza, S., Hogan, D., Chu, Y., Perin, J., Zhu, J., Lawn, J. E., Cousens, S., Mathers, C., Black, R. E., "Global, regional, and national causes of under-5 mortality in 2000-15: an updated systematic analysis with implications for the Sustainable Development Goals," The Lancet, Vol. 388, No. 10063, pp. 3027-3035, 2016. https://doi.org/10.1016/S0140-6736(16)31593-8
  4. "WHO recommendations on interventions to improve preterm birth outcomes," World Health Organization, 2015.
  5. Aita, M., Snider, L., "The art of developmental care in the NICU: a concept analysis," Journal of Advanced Nursing, Vol. 41, No. 3, pp. 223-232, 2003. https://doi.org/10.1046/j.1365-2648.2003.02526.x
  6. Delaporte, S. E., Gazes, M. J., Keefe, T. J., "Modular neonatal intensive care system," United States patent application, US20100168502A1, 2010.
  7. Poll, S. G., "Infant incubator," United States patent, US1914288A, 1933.
  8. Gordon, A., "Portable baby incubator," United States patent, US2417962A, 1947.
  9. Louis, P. J., "Portable infant incubator," United States patent, US2470721A, 1949.
  10. Grosholz, J. R., "Portable incubator for infants," United States patent, US3529590A, 1970.
  11. Belval, J. C., Brox, P. T., Dean, I. V. H., Raver, B. P., "Method and Apparatus for Providing a Portable Neonatal Transport Incubator," United States patent application, US20130204074A1, 2013.
  12. Breiner, S., Sipter, F., "Incubator," UK Patent, GB2137504A, 1984.
  13. Roberts, J. M., "A collapsible incubator," United States patent application, US20170119609A1, 2017.
  14. Ballesteros, T., Arana, I., Latorre-Biel, J. I., Perez-Ezcurdia, A., ALFARO, J. R., "Development and Tests of a Neonatal Portable Foldable Emergency Incubator," Journal of Mechanics in Medicine and Biology, Vol. 18, No. 06, pp. 1850058, 2018. https://doi.org/10.1142/S0219519418500586

피인용 문헌

  1. A Short Review of Appropriate Technology and Engineering Design Education for Underdeveloped Countries vol.19, pp.7, 2020, https://doi.org/10.14775/ksmpe.2020.19.07.112
  2. 가정용 영유아 스마트 인큐베이터 개발 vol.24, pp.10, 2020, https://doi.org/10.6109/jkiice.2020.24.10.1325