A Convergence Study on the Measurement of Bacterial Pollution in Medical Ultrasonic Practice

의료용 초음파 실습 시 장치의 세균오염도 측정에 관한 융합적 연구

  • Kim, Dong-Heun (Department of Radiological Science, Kangwon National University) ;
  • Park, Sang-Hee (Department of Radiological Science, Kangwon National University) ;
  • Park, Gyu-Tae (Department of Radiological Science, Kangwon National University) ;
  • Jung, Won-Hee (Department of Radiological Science, Kangwon National University) ;
  • Kim, So-Yeon (Department of Radiological Science, Kangwon National University) ;
  • Hong, Hee-Jin (Department of Dental Hygiene, College of Health Sciences, Kangwon National University) ;
  • Son, Na-Ra (Department of Dental Hygiene, College of Health Sciences, Kangwon National University) ;
  • Nam, Seoul-Hee (Department of Dental Hygiene, College of Health Sciences, Kangwon National University) ;
  • Han, Man-Seok (Department of Radiological Science, Kangwon National University)
  • 김동흔 (강원대학교 방사선학과) ;
  • 박상희 (강원대학교 방사선학과) ;
  • 박규태 (강원대학교 방사선학과) ;
  • 정원희 (강원대학교 방사선학과) ;
  • 김소연 (강원대학교 방사선학과) ;
  • 홍희진 (강원대학교 치위생학과) ;
  • 손나라 (강원대학교 치위생학과) ;
  • 남설희 (강원대학교 치위생학과) ;
  • 한만석 (강원대학교 방사선학과)
  • Received : 2019.08.22
  • Accepted : 2019.10.20
  • Published : 2019.10.28

Abstract

During the medical ultrasound exercise at school, we randomly select parts of the ultrasound device and areas with the most contact in the abdominal phantom to detect bacteria that are above the probe and determine the number of pathogens. I want to find out. The experimental method was rubbed 20 times with the sterilized cotton swab for sterilization and then smeared on Lysogeny broth (LB) agar, put into the incubator and incubated for 48 hours, and the colony forming unit (CFU) count was measured. The bacterial distribution of probe handle and abdominal phantom was evaluated by evaluation. As a result, the CFU value is the lens was $3.0{\pm}0.87$, print button was $5.5{\pm}1.06$, freeze button was $8.0{\pm}4.95$, phantom was $20.0{\pm}2.78$, line was $23.5{\pm}2.50$, and probe handle was measured as $35.3{\pm}10.75$. In this study, it is expected that attention to infection control of equipment during practice during medical ultrasound practice can be highlighted and further contributed to the reduction of bacterial infection rate of ultrasound devices.

학교에서 의료용 초음파 실습 시 초음파 장치의 Freeze 및 Print Button, 프로브의 Lens, 손잡이, Line 부분 그리고 복부 Phantom에서 가장 접촉이 많은 부위를 임의로 선정하여 프로브에 상제하고 있는 세균을 검출하여 병원균의 수를 알아보고자 한다. 실험 방법으로는 실험 대상 위치에 균채집용 멸균된 면봉으로 검사 대상 부위를 20번 문지른 후 Lysogeny broth(LB) agar에 도말을 실시한 후 배양기에 넣어 48시간 동안 배양하고 colony forming unit (CFU) 수를 평가하여 프로브 손잡이, 복부 Phantom의 세균 분포정도를 알아보았다. 그 결과 CFU 값은 Lens는 $3.0{\pm}0.87$, Print button는 $5.5{\pm}1.06$, Freeze button은 $8.0{\pm}4.95$, Phantom은 $20.0{\pm}2.78$, Line은 $23.5{\pm}2.50$, 그리고 Probe handle의 값은 $35.3{\pm}10.75$로 측정되었다. 본 연구에서는 의료용 초음파 실습 시 실습 시 장비의 감염관리에 대한 주의를 부각시키고, 나아가서 초음파 장치의 세균 감염률 감소에 확실하게 기여 할 수 있을 것으로 기대된다.

Keywords

References

  1. J. S. Choi. (1995). Principle and Application of Ultrasonic Image Diagnosis. The Processing of the Institute of Electrical Engineers, 34(9), 541-547
  2. M. J. Kim, M. S. Han & J. U. Jang. (2017). Evaluation of convergence Elasticity of Liver Fibroscan used measurement with Ultrasonography. Journal of the Korea Convergence Society, 8(5), 79-85 https://doi.org/10.15207/JKCS.2017.8.5.079
  3. C. H. Kim et al. (2004). Investigation into the Actual State or Sanitary Management and Recognition Degree and Infection Level of Ultrasonographic Probes. Journal of radiological science and technology. 27(3), 51 -58
  4. J. E Moon & M. O. Song. (2017). A Convergence Study about the Performance of Healthcare-Associated Infection Control Guidelines of Hospital Nurses-based on the Theory of Planned Behavior. Journal of the Korea Convergence Society, 8(1), 51-59, https://doi.org/10.15207/JKCS.2017.8.1.051
  5. B. J. Misset et al. (2004). A continuous quality improvement program reduces nosocomial infection rates in the ICU. Intensive Care Med, 30(3), 395-400, https://doi.org/10.1007/s00134-003-2096-1
  6. S. Y. Jeong, H. S. Oh & H. K. Chun. (2015). Analysis of the Status of Infection Controls after Application of the Healthcare Accreditation System. The Korean journal of health service management, 9(4), 33 https://doi.org/10.12811/kshsm.2015.9.4.033
  7. C. H. Woo, J. Y. Park, S. Y. Lee & J. E. Oh. (2018). Factors Influencing the Infection Control Practice of Clinical Nurses based on Health Belief Model. Journal of the Korea Convergence Society 9(3), 121-129 https://doi.org/10.15207/JKCS.2018.9.3.121
  8. J. Yoon & H. J. Kim. (2018). A study on effective disinfection methods of medical ultrasound probe resident floras. Journal of the Korea Academia-Industrial cooperation Society, 19(1), 346-354. https://doi.org/10.5762/KAIS.2018.19.1.346
  9. H. K. Lee, S. S. Kim, Y. C. Heo & D.K Han. (2018). A Study on Microbial Contamination and Disinfection of Ultrasonic Probe in Metropolitan Area. Journal of Radiological Science and Technology, 41(5), 427-435 https://doi.org/10.17946/JRST.2018.41.3.427
  10. D. H. Hong & M.S. Park. (2017). Analysis of Healthcare Convergence on Bacterial Contamination of Radiological Equipment in Emergency Rooms of General Hospitals. Journal of the Korea Convergence Society 8(1), 51-59 https://doi.org/10.15207/JKCS.2017.8.1.051
  11. K. Y. Choi, S, J. Yoo, J. H. Lee & S. Y. Hon. (2019). A Study on the Necessity Verification of Convex Probe Disinfection. Journal of the Korean Society of Radiology, 13(2), 193-200 https://doi.org/10.7742/JKSR.2019.13.2.193
  12. S. H. Bae, M. S. Lee, C. S. Lim & G. J. Kim. (2008). A Study on the Measurement of the Pollution Level of Bacteria and Disinfection of Table and IP Cassette, Journal of radiological science and technology, 31(3), 229-237
  13. S. H. Shin. (2019). Relationships between Health Promoting Lifestyle, Health Belief about Emerging Infectious Disease and Hygiene Behavior of College Students, Journal of the Korea Convergence Society. 10(4), 285-293 https://doi.org/10.15207/JKCS.2019.10.4.285
  14. S. Hayashi, H. Koibuchi, N. Taniguchi & Y. Hirai. (2012). Evaluation of procedures for decontaminating ultrasoundprobes, Journal of Medical Ultrasonics. 39(1), 11-4. https://doi.org/10.1007/s10396-011-0332-9
  15. H. Koibuchi, K. Kotani & N. Taniguchi. (2013). Ultrasoundprobes as a possible vector of bacterial transmission. Medical ultrasonography, 15(1), 41-4. https://doi.org/10.11152/mu.2013.2066.151.hk1upp2
  16. J. S. Jeong, J. H. Choi, S. H. Lee & Y. S. Kim. (2003). Hand Hygiene Effects Measured by Hand Culture in Intensive Care Unit. The Journal of Korean Biological Nursing Science, 5(2), 21-30,
  17. S. Y. Jeong, H. S. Oh & H. K. Chun. (2015). Analysis of the Status of Infection Controls after Application of the Healthcare Accreditation System. The Korean journal of health service management, 9(4), 33. https://doi.org/10.12811/kshsm.2015.9.4.033
  18. H. Y. Lee & S. G. Shin. (2017). Bacteriological Research for the Contamination of Digital Portable Radiography. Journal of Radiological Science and Technology, 40(1), 7-12. https://doi.org/10.17946/JRST.2017.40.1.02
  19. D. H. Hong & H. G. Kim (2016). Analysis of Bacterial Contamination on Surface of General Radiography Equipment and CT Equipment in Emergency Room of Radiology. Journal of Radiological Science and Technology, 39(3), 421-27. https://doi.org/10.17946/JRST.2016.39.3.17
  20. J. Y. Lee. (2017). Influence of Diagnostic X-ray Laboratory Automatic System on Pathogen Infection. Radiology, Mater's thesis. Cheongju University, Cheongju.