Development of a Web-Based Program for Cross-Calibration and Record Management of Radiation Measuring Equipment

  • Park, So Hyun (Department of Radiation Oncology, Jeju National University Hospital, Jeju University College of Medicine) ;
  • Lee, Rena (Department of Radiation Oncology, Ewha Womans University School of Medicine) ;
  • Kim, Kyubo (Department of Radiation Oncology, Ewha Womans University School of Medicine) ;
  • Ahn, Sohyun (Department of Radiation Oncology, Kangwon National University Hospital, Kangwon University College of Medicine) ;
  • Lim, Sangwook (Department of Radiation Oncology, Kosin University College of Medicine) ;
  • Cho, Samju (Department of Radiation Oncology, Ewha Womans University Medical Center, Ewha Womans University School of Medicine)
  • Received : 2019.04.17
  • Accepted : 2019.06.21
  • Published : 2019.06.30

Abstract

Purpose: To manage radiation measurement equipment, a web-based management program has been developed in this study. Materials and Methods: This program is based on a web service and Java Server Pages (JSP) and employs compatibility and accessibility. Results: The first step in the workflow has been designed to create accounts for each user or organization and to log in. The program consists of two parts: fields for listed instruments, and measurement information. The instruments for measuring radiation listed in this program are as follows: ionization chambers, survey meters, thermometers, barometers, electrometers, and phantoms. Instrument properties can be put in the recording fields and browsing for associated instruments can be performed. The main part of the program is the cross-calibration for each ion chamber. For instance, the ionization chamber to be used as a relative dosimeter can be registered by cross-calibration data with a reference chamber calibrated by an accredited laboratory. This program supports methods using the central axis transfer theory for cross-calibration for the ionization chambers. The reference and field ionization chambers were placed in a solid water phantom along the beam central axis at two different depths, and then the positions were switched. Each measured value was used for calculating the cross-calibration factor. Conclusions: Because many instruments are used and managed in radiation oncology departments, systematic, traceable recording is very important. The web-based program developed in this study is expected to be used effectively in the maintenance of radiation measurement instruments.

Keywords

References

  1. Han Yy, Huh SJ, Ju SG, Ahn YC, Lim DH, Lee JE, Par W. Impact of an electronic chart on the staff workload in a radiation oncology department. Jpn J Clin Oncol. 2005;35(8):470-474. https://doi.org/10.1093/jjco/hyi129
  2. Lee DH. The development of medical information management system of radiation oncology department. Journal of the korea institute of maritime information & communication sciences. 2010;14(3):655-662. https://doi.org/10.6109/jkiice.2010.14.3.655
  3. Absorbed dose determination in external beam radiotherapy: An international code of practice for dosimetry based on standards of absorbed dose to water. IAEA TRS-398 2006;12:1-183.
  4. Klein EE, Hanley JS, Bayouth J, Yin FF, Simon W, Dresser S, Serago Ch, Aguirre F, Ma LJ, Arhomandy BJ, Liu Ch. Task group 142 report: quality assurance of medical accelerators. Am. Assoc. Phys. Med. 2009;36(9):4197-4212.
  5. Kutcher GJ, Coia L, Gillin M, Hanson WF, Leibel S, Morton RJ, Palta JR, Purdy JA, Reinstein LE, Svensson GK, Weller M, Wingfield L. Comprehensive QA for radiation oncology: report of AAPM radiation therapy committee task group 40. Med. Phys.1994;21(4):581-618. https://doi.org/10.1118/1.597316
  6. Serre L. Transfer of ionization chamber calibration coefficients in linac MV x-ray beams. 2008:1-123.
  7. Scott LS, Joel ET, Larry MM, Julian GR, Edward LC. RAPID: An electronic medical records system for radiation oncology. Seminars in Radiation Oncology. 1997;7(1):4-10 https://doi.org/10.1016/S1053-4296(97)80012-4
  8. John AK, Aleksandra Z, Jeffery S, Paul F, Vythialingam S, Cynthia B, Alfred WR, Nick R, Bharat BM. A comprehensive quality assurance program for personnel and procedures in radiation oncology: value of voluntary error reporting and checklists. Int J Radiat Oncol Biol Phys. 2013;86(2):241-248. https://doi.org/10.1016/j.ijrobp.2013.02.003
  9. Lawrence BM, Marianne J, Liyi X, Sha XC, Katharin DB, Lukasz M, Ellen LJ, Patricia S, Dana L, Dee CB, Robert DA. Practical Radiation Oncology. 2011;1:2-14. https://doi.org/10.1016/j.prro.2010.10.001