Load response of the natural tooth and dental implant: A comparative biomechanics study

  • Robinson, Dale (Department of Biomedical Engineering, University of Melbourne) ;
  • Aguilar, Luis (Department of Biomedical Engineering, University of Melbourne) ;
  • Gatti, Andrea (Department of Biomedical Engineering, University of Melbourne) ;
  • Abduo, Jaafar (Melbourne Dental School, University of Melbourne) ;
  • Lee, Peter Vee Sin (Department of Biomedical Engineering, University of Melbourne) ;
  • Ackland, David (Department of Biomedical Engineering, University of Melbourne)
  • 투고 : 2018.10.03
  • 심사 : 2019.06.11
  • 발행 : 2019.06.28

초록

PURPOSE. While dental implants have displayed high success rates, poor mechanical fixation is a common complication, and their biomechanical response to occlusal loading remains poorly understood. This study aimed to develop and validate a computational model of a natural first premolar and a dental implant with matching crown morphology, and quantify their mechanical response to loading at the occlusal surface. MATERIALS AND METHODS. A finite-element model of the stomatognathic system comprising the mandible, first premolar and periodontal ligament (PDL) was developed based on a natural human tooth, and a model of a dental implant of identical occlusal geometry was also created. Occlusal loading was simulated using point forces applied at seven landmarks on each crown. Model predictions were validated using strain gauge measurements acquired during loading of matched physical models of the tooth and implant assemblies. RESULTS. For the natural tooth, the maximum vonMises stress (6.4 MPa) and maximal principal strains at the mandible ($1.8m{\varepsilon}$, $-1.7m{\varepsilon}$) were lower than those observed at the prosthetic tooth (12.5 MPa, $3.2m{\varepsilon}$, and $-4.4m{\varepsilon}$, respectively). As occlusal load was applied more bucally relative to the tooth central axis, stress and strain magnitudes increased. CONCLUSION. Occlusal loading of the natural tooth results in lower stress-strain magnitudes in the underlying alveolar bone than those associated with a dental implant of matched occlusal anatomy. The PDL may function to mitigate axial and bending stress intensities resulting from off-centered occlusal loads. The findings may be useful in dental implant design, restoration material selection, and surgical planning.

키워드

참고문헌

  1. Moraschini V, Poubel LA, Ferreira VF, Barboza Edos S. Evaluation of survival and success rates of dental implants reported in longitudinal studies with a follow-up period of at least 10 years: a systematic review. Int J Oral Maxillofac Surg 2015;44:377-88. https://doi.org/10.1016/j.ijom.2014.10.023
  2. Pjetursson BE, Asgeirsson AG, Zwahlen M, Sailer I. Improvements in implant dentistry over the last decade: comparison of survival and complication rates in older and newer publications. Int J Oral Maxillofac Implants 2014;29:308-24. https://doi.org/10.11607/jomi.2014suppl.g5.2
  3. Jung RE, Pjetursson BE, Glauser R, Zembic A, Zwahlen M, Lang NP. A systematic review of the 5-year survival and complication rates of implant-supported single crowns. Clin Oral Implants Res 2008;19:119-30. https://doi.org/10.1111/j.1600-0501.2007.01453.x
  4. Esposito M, Hirsch JM, Lekholm U, Thomsen P. Biological factors contributing to failures of osseointegrated oral implants. (I). Success criteria and epidemiology. Eur J Oral Sci 1998;106:527-51. https://doi.org/10.1046/j.0909-8836..t01-2-.x
  5. Esposito M, Hirsch JM, Lekholm U, Thomsen P. Biological factors contributing to failures of osseointegrated oral implants. (II). Etiopathogenesis. Eur J Oral Sci 1998;106:721-64. https://doi.org/10.1046/j.0909-8836..t01-6-.x
  6. Pye AD, Lockhart DE, Dawson MP, Murray CA, Smith AJ. A review of dental implants and infection. J Hosp Infect 2009;72:104-10. https://doi.org/10.1016/j.jhin.2009.02.010
  7. Misch CE, Perel ML, Wang HL, Sammartino G, Galindo-Moreno P, Trisi P, Steigmann M, Rebaudi A, Palti A, Pikos MA, Schwartz-Arad D, Choukroun J, Gutierrez-Perez JL, Marenzi G, Valavanis DK. Implant success, survival, and failure: the International Congress of Oral Implantologists (ICOI) Pisa Consensus Conference. Implant Dent 2008;17:5-15. https://doi.org/10.1097/ID.0b013e3181676059
  8. Eskitascioglu G, Usumez A, Sevimay M, Soykan E, Unsal E. The influence of occlusal loading location on stresses transferred to implant-supported prostheses and supporting bone: A three-dimensional finite element study. J Prosthet Dent 2004;91:144-50. https://doi.org/10.1016/j.prosdent.2003.10.018
  9. Sevimay M, Turhan F, Kilicarslan MA, Eskitascioglu G. Three-dimensional finite element analysis of the effect of different bone quality on stress distribution in an implant-supported crown. J Prosthet Dent 2005;93:227-34. https://doi.org/10.1016/j.prosdent.2004.12.019
  10. Soncini M, Rodriguez y Baena R, Pietrabissa R, Quaglini V, Rizzo S, Zaffe D. Experimental procedure for the evaluation of the mechanical properties of the bone surrounding dental implants. Biomaterials 2002;23:9-17. https://doi.org/10.1016/S0142-9612(01)00073-4
  11. Geng JP, Tan KB, Liu GR. Application of finite element analysis in implant dentistry: a review of the literature. J Prosthet Dent 2001;85:585-98. https://doi.org/10.1067/mpr.2001.115251
  12. Lin CL, Lin YH, Chang SH. Multi-factorial analysis of variables influencing the bone loss of an implant placed in the maxilla: prediction using FEA and SED bone remodeling algorithm. J Biomech 2010;43:644-51. https://doi.org/10.1016/j.jbiomech.2009.10.030
  13. de Las Casas EB, de Almeida AF, Cimini Junior CA, Gomes Pde T, Cornacchia TP, Saffar JM. Determination of tangential and normal components of oral forces. J Appl Oral Sci 2007;15:70-6. https://doi.org/10.1590/S1678-77572007000100015
  14. Du J, Lee JH, Jang AT, Gu A, Hossaini-Zadeh M, Prevost R, Curtis DA, Ho SP. Biomechanics and strain mapping in bone as related to immediately-loaded dental implants. J Biomech 2015;48:3486-94. https://doi.org/10.1016/j.jbiomech.2015.05.014
  15. Chee WW, Mordohai N. Tooth-to-implant connection: a systematic review of the literature and a case report utilizing a new connection design. Clin Implant Dent Relat Res 2010;12:122-33. https://doi.org/10.1111/j.1708-8208.2008.00144.x
  16. Fill TS, Carey JP, Toogood RW, Major PW. Experimentally determined mechanical properties of, and models for, the periodontal ligament: critical review of current literature. J Dent Biomech 2011;2011:312980.
  17. Melsen B. Biological reaction of alveolar bone to orthodontic tooth movement. Angle Orthod 1999;69:151-8.
  18. Cattaneo PM, Dalstra M, Melsen B. Strains in periodontal ligament and alveolar bone associated with orthodontic tooth movement analyzed by finite element. Orthod Craniofac Res 2009;12:120-8. https://doi.org/10.1111/j.1601-6343.2009.01445.x
  19. Brunski JB. In vivo bone response to biomechanical loading at the bone/dental-implant interface. Adv Dent Res 1999;13:99-119. https://doi.org/10.1177/08959374990130012301
  20. Taddei F, Cristofolini L, Martelli S, Gill HS, Viceconti M. Subject-specific finite element models of long bones: An in vitro evaluation of the overall accuracy. J Biomech 2006;39:2457-67. https://doi.org/10.1016/j.jbiomech.2005.07.018
  21. Morneburg TR, Proschel PA. In vivo forces on implants influenced by occlusal scheme and food consistency. Int J Prosthodont 2003;16:481-6.
  22. Stanford CM, Brand RA. Toward an understanding of implant occlusion and strain adaptive bone modeling and remodeling. J Prosthet Dent 1999;81:553-61. https://doi.org/10.1016/S0022-3913(99)70209-X
  23. Ogawa T, Koyano K, Suetsugu T. The relationship between inclination of the occlusal plane and jaw closing path. J Prosthet Dent 1996;76:576-80. https://doi.org/10.1016/S0022-3913(96)90432-1
  24. Ogawa T, Koyano K, Suetsugu T. Characteristics of masticatory movement in relation to inclination of occlusal plane. J Oral Rehabil 1997;24:652-7. https://doi.org/10.1046/j.1365-2842.1997.00543.x
  25. Martos J, Lubian C, Silveira LF, Suita de Castro LA, Ferrer Luque CM. Morphologic analysis of the root apex in human teeth. J Endod 2010;36:664-7. https://doi.org/10.1016/j.joen.2010.01.014
  26. Alvarez-Arenal A, Segura-Mori L, Gonzalez-Gonzalez I, DeLlanos-Lanchares H, Sanchez-Lasheras F, Ellacuria-Echevarria J. Stress distribution in the transitional peri-implant bone in a single implant-supported prosthesis with platform-switching under different angulated loads. Odontology 2017;105:68-75. https://doi.org/10.1007/s10266-016-0237-6
  27. Cheng HC, Peng BY, Chen MS, Huang CF, Lin Y, Shen YK. Influence of deformation and stress between bone and implant from various bite forces by numerical simulation analysis. Biomed Res Int 2017;2017:2827953.
  28. Weinberg LA. Reduction of implant loading using a modified centric occlusal anatomy. Int J Prosthodont 1998;11:55-69.
  29. Weinberg LA. Therapeutic biomechanics concepts and clinical procedures to reduce implant loading. Part I. J Oral Implantol 2001;27:293-301. https://doi.org/10.1563/1548-1336(2001)027<0293:TBCACP>2.3.CO;2
  30. Katona TR, Eckert GJ. The mechanics of dental occlusion and disclusion. Clin Biomech (Bristol, Avon) 2017;50:84-91. https://doi.org/10.1016/j.clinbiomech.2017.10.009
  31. Sheridan RA, Decker AM, Plonka AB, Wang HL. The role of occlusion in implant therapy: A comprehensive updated review. Implant Dent 2016;25:829-38. https://doi.org/10.1097/ID.0000000000000488
  32. Abichandani SJ, Bhojaraju N, Guttal S, Srilakshmi J. Implant protected occlusion: A comprehensive review. Eur J Prosthodont 2013;1:29-36. https://doi.org/10.4103/2347-4610.116588
  33. Kim Y, Oh TJ, Misch CE, Wang HL. Occlusal considerations in implant therapy: clinical guidelines with biomechanical rationale. Clin Oral Implants Res 2005;16:26-35. https://doi.org/10.1111/j.1600-0501.2004.01067.x
  34. Lo J, Abduo J, Palamara J. Effect of different lateral occlusion schemes on peri-implant strain: A laboratory study. J Adv Prosthodont 2017;9:45-51. https://doi.org/10.4047/jap.2017.9.1.45
  35. Taylor TD, Belser U, Mericske-Stern R. Prosthodontic considerations. Clin Oral Implants Res 2000;11:101-7. https://doi.org/10.1034/j.1600-0501.2000.011S1101.x
  36. Chen YY, Kuan CL, Wang YB. Implant occlusion: biomechanical considerations for implant-supported prostheses. J Dent Sci 2008;3:65-74.
  37. Rungsiyakull P, Rungsiyakull C, Appleyard R, Li Q, Swain M, Klineberg I. Loading of a single implant in simulated bone. Int J Prosthodont 2011;24:140-3.
  38. Koyano K, Esaki D. Occlusion on oral implants: current clinical guidelines. J Oral Rehabil 2015;42:153-61. https://doi.org/10.1111/joor.12239
  39. Rohrle O, Saini H, Lee PVS, Ackland DC. A novel computational method to determine subject-specific bite force and occlusal loading during mastication. Comput Methods Biomech Biomed Engin 2018;21:453-60. https://doi.org/10.1080/10255842.2018.1479744
  40. Giesen EBW, Ding M, Dalstra M, Van Eijden TM. Mechanical properties of cancellous bone in the human mandibular condyle are anisotropic. J Biomech 2001;34:799-803. https://doi.org/10.1016/S0021-9290(01)00030-6

피인용 문헌

  1. Effects of cementless fixation of implant prosthesis: A finite element study vol.11, pp.6, 2019, https://doi.org/10.4047/jap.2019.11.6.341
  2. Stress Distribution Patterns Associated with Dental Implants with Varying Thread Designs, Dimensions and Splinting Conditions: A Photoelastic Analysis vol.13, pp.1, 2021, https://doi.org/10.5005/jp-journals-10015-1885
  3. Mechanical stability and technical outcomes of monolithic CAD/CAM fabricated abutment‐crowns supported by titanium bases: An in vitro study vol.32, pp.2, 2019, https://doi.org/10.1111/clr.13693
  4. Biomechanical evaluation of a novel mandibular distraction osteogenesis protocol: an in-vitro validation and the practical use of the method vol.24, pp.7, 2021, https://doi.org/10.1080/10255842.2020.1850704
  5. An application of finite element method in material selection for dental implant crowns vol.66, pp.4, 2019, https://doi.org/10.1515/bmt-2020-0232
  6. Mechanical Performance of Chairside Ceramic CAD/CAM Restorations and Zirconia Abutments with Different Internal Implant Connections: In Vitro Study and Finite Element Analysis vol.14, pp.17, 2019, https://doi.org/10.3390/ma14175009