DOI QR코드

DOI QR Code

PVDF 압전소자를 이용한 심장박동 및 호흡수 동시측정센서개발

Development of New Stacked Element Piezoelectric Polyvinylidene Fluoride Pressure Sensor for Simultaneous Heartbeat and Respiration Measurements

  • 박창용 (금오공과대학교 기계시스템공학과) ;
  • 권현규 (금오공과대학교 기계시스템공학과) ;
  • 이소진 (금오공과대학교 기계시스템공학과) ;
  • 롱원만 (금오공과대학교 기계시스템공학과)
  • Park, Chang-Yong (Department of Mechanical System Engineering, Kumoh National Institute of Technology) ;
  • Kweon, Hyun-Kyu (Department of Mechanical System Engineering, Kumoh National Institute of Technology) ;
  • Lee, So-Jin (Department of Mechanical System Engineering, Kumoh National Institute of Technology) ;
  • Manh, Long-Nguyen (Department of Mechanical System Engineering, Kumoh National Institute of Technology)
  • 투고 : 2019.03.05
  • 심사 : 2019.04.09
  • 발행 : 2019.04.30

초록

In this paper, a new stacked element pressure sensor has proposed for heartbeat and respiration measurement. This device can be directly attached to an individual's chest; heartbeat and respiration are detected by the pulsatile vibration and deformation of the chest. A key feature of the device is the simultaneous measurement of heart rate and respiration. The structure of the sensor consists of two stacked elements, in which one element includes one polyvinylidene fluoride (PVDF) thin film bonded on polydimethylsiloxane (PDMS) substrate. In addition, for the measurement and signal processing, the electric circuit and the filter are simply constructed with an OP-amp, resistance, and a capacitor. One element (element1, PDMS) maximizes the respiration signal; the other (element2, PVDF) is used to measure heartbeat. Element1 and element2 had sensitivity of 0.163V/N and 0.209V/N, respectively, and element2 showed improved characteristics compared with element1 in response to force. Thus, element1 and element2 were optimized for measuring respiration heart rate, respectively. Through mechanical and vivo human tests, this sensor shows the great potential to optimize the signals of heartbeat and respiration compared with commercial devices. Moreover, the proposed sensor is flexible, light weight, and low cost. All of these characteristics illustrate an effective piezoelectric pressure sensor for heartbeat and respiration measurements.

키워드

참고문헌

  1. Folke, M., Cernerud, L., Ekstrom, M., and Hok, B., Critical review of non-invasive respiratory monitoring in medical care, Medical and Biological Engineering and Computing 41, pp. 377-383, 2003. https://doi.org/10.1007/BF02348078
  2. Moise, R., Christian, C. and Joseph, V., "Einthoven's string galvanometer", Texas Heart Institute Journal, Vol. 35 No. 2, pp. 174-178, 2008.
  3. Jacobs, J. L., Embree, P., Glei, M., Christensen, S. and Sulivan, P. K., "Characterization of a novel heart and respiratory rate sensor", Conference Proceeding-IEEE Engineering in Medicine and Biology Society 3, pp.2223-2226, 2004.
  4. Choi, S. and Jiang, Z., "A novel wearable sensor device for conductive fabric and PVDF film for monitoring cardiorespiratory signals", Sensors and Actuators A: Physics, Vol. 128, Issue 2, pp.317-326, 2006. https://doi.org/10.1016/j.sna.2006.02.012
  5. Chiu, Y. Y., Lin, W. Y., Wang, H. Y., Huang, S. B. and Wu, M. H., "Development of a piezoelectric polyvinylidene fluoride (PVDF) polymer-based sensor patch for simultaneous heartbeat and respiration monitor", Sensors and Actuators A, Vol. 189, pp.328-334, 2013. https://doi.org/10.1016/j.sna.2012.10.021
  6. Lindh, W. Q., Pooler, M., Tamparo, C. D., Dahl, B. M. and Moris, J., Delmar's, Comprehensive Medical Assisting: Administrative and Clinical Competencies, 5th Edition Delmar Cengage Learning, Stamford, 2013.
  7. Ramachandram, G. and Singh, M., "Three dimensional reconstruction of cardiac displacement patterns on the chest wall during the P, QRS and T-segments of the ECG by laser speckle inteferometry", Medicine & Biological Engineering & Computing, Vol. 27, Issue. 5, pp.525-530, 1989. https://doi.org/10.1007/BF02441473
  8. Groote, A. D. E., Wantier, M., Cheron, G., Estenne, M. and Paiva, M., "Chest wall motion during tidal breathing", Journal applied physiology: respiratory, environmental and exercise physiology, Vol. 83, Issue. 5, pp.1531-1537, 1997. https://doi.org/10.1152/jappl.1997.83.5.1531
  9. Sajala, S. and Lekkala, J., "Film-type sensor Materials PVDF and EMFi in measurement of cardiorespiratory signals-a Review", IEEE Sensors Journal, Vol. 12, Issue. 3, pp.439-446, 2012. https://doi.org/10.1109/JSEN.2010.2089510
  10. Tresler, J. F., Alkoy, S. and Newham, R. E., "Piezoelectric Sensors and Sensor Materials", Journal of Electroceramics, Vol. 2, Issue. 4, pp.252-257, 1998.
  11. Safari, A. and Akdogan, E. K., Piezoelectric and Acoustic Materials for Transducer Applications, Springer, 2008.
  12. Lotters, J. C., Olthuis, W., Veltink, P. H. and Bergveld, P., "The mechanical properties of the rubber elastic polymer polydimethylsiloxane for sensor applications", Journal Micromechanics and Microengineering, Vol. 7, Issue. 3, pp.145-147, 1997. https://doi.org/10.1088/0960-1317/7/3/017
  13. McDonald, J. C., Duffy, D. C., Anderson, J. R., Chiu, D. T., Wu, H., Schueller, O. J. A. and Whitesides, G. M., "Fabrication of microfluidic systems in poly(dimethylsiloxane)", Electrophoresis, Vol. 21, Issue. 1, pp.27-40, 2000. https://doi.org/10.1002/(SICI)1522-2683(20000101)21:1<27::AID-ELPS27>3.0.CO;2-C
  14. "CH-456:HEALTH CARE PRODUCTS:CITIZEN SYSTEMS JAPAN CO.,LTD."(2019), http://www.citizen-systems.co.jp/english/electronic/health/ch_456.html(accesed 22, Feb., 2019)