Accuracy of different electronic torque drivers: A comparative evaluation

  • Ko, Byeong-Dae (Department of Dentistry, Graduate School, Kyungpook National University) ;
  • Son, KeunBaDa (Department of Dental Science, Graduate School, Kyungpook National University) ;
  • Kang, Seok Hyon (Advanced Dental Device Development Institute, Kyungpook National University) ;
  • Lee, Kyu-Bok (Department of Dentistry, Graduate School, Kyungpook National University)
  • Received : 2019.08.04
  • Accepted : 2019.12.10
  • Published : 2019.12.31

Abstract

PURPOSE. This study aims to evaluate the loosening torque on the implant fixture, and to assess the accuracy of difference electronic torque drivers. MATERIALS AND METHODS. Three electronic torque drivers were used to measure the loosening torque on the implant system (AnyOne; MegaGen). The implant fixtures were divided among the 3 electronic torque driver types (W&H, SAESHIN, and NSK group) and 9 for each group. The screws were fastened at the implant fixture by three electronic torque drivers using the tightening torques recommended by the manufacturers of the drivers. After 10 minutes, the screws were again fastened at the implant fixture with equal torque. Then, the loosening torques were measured with an MGT12 torque gauge (MARK-10, Inc.). This measurement procedure was repeated 10 times under loosening torques of 15 Ncm, 25 Ncm, and 35 Ncm. In the statistical analysis, all values of loosening torque were analyzed with the one-way ANOVA and Kruskal-Wallis test (α=.05) for comparative evaluation. RESULTS. There were significant inter-group differences at loosening torques of 15 Ncm and 25 Ncm (P<.05). The accuracy of the NSK driver was the highest, followed by SAESHIN and W&H. There was no significant difference between NSK and W&H at 35 Ncm (P>.05). The SAESHIN driver showed the closest loosening torque at 35 Ncm. CONCLUSION. The most accurate loosening torques were SAESHIN at 35 Ncm, and NSK at 15 Ncm and 25 Ncm. Since the loosening torque may vary depending on the tightening torques and electronic torque drivers, periodic calibration of the electronic torque driver is recommended.

Keywords

References

  1. Branemark PI. Osseointegration and its experimental background. J Prosthet Dent 1983;50:399-410. https://doi.org/10.1016/S0022-3913(83)80101-2
  2. Sumi T, Braian M, Shimada A, Shibata N, Takeshita K, Vandeweghe S, Coelho PG, Wennerberg A, Jimbo R. Characteristics of implant-CAD/CAM abutment connections of two different internal connection systems. J Oral Rehabil 2012;39:391-8. https://doi.org/10.1111/j.1365-2842.2011.02273.x
  3. Lang LA, Sierraalta M, Hoffensperger M, Wang RF. Evaluation of the precision of fit between the Procera custom abutment and various implant systems. Int J Oral Maxillofac Implants 2003;18:652-8.
  4. Adell R, Eriksson B, Lekholm U, Branemark PI, Jemt T. Long-term follow-up study of osseointegrated implants in the treatment of totally edentulous jaws. Int J Oral Maxillofac Implants 1990;5:347-59.
  5. Binon PP. The effect of implant/abutment hexagonal misfit on screw joint stability. Int J Prosthodont 1996;9:149-60.
  6. Lee WJ, Lim JH, Cho IH. Stress of dental implant abutment screw by the tightening torque. J Korean Acad Prosthodont 1998;36:721-37.
  7. Lim JB, Yim SH, Cho IH. The effect of screw hole sealing method on abutment screw loosening in dental implant. J Korean Acad Prosthodont 1997;35:767-80.
  8. Cho SM, Lim JH, Cho IH. A study on geometric comparison of four interchangeable implant prosthetic retaining screws and measurement of loosening torgue. J Korean Acad Prosthodont 1998;36:468-82.
  9. Balshi TJ, Hernandez RE, Pryszlak MC, Rangert B. A comparative study of one implant versus two replacing a single molar. Int J Oral Maxillofac Implants 1996;11:372-8.
  10. Kemppainen P, Eskola S, Ylipaavalniemi P. A comparative prospective clinical study of two single-tooth implants: a preliminary report of 102 implants. J Prosthet Dent 1997;77:382-7. https://doi.org/10.1016/S0022-3913(97)70163-X
  11. Cho IH. Biomechanical analysis of implant prosthodontic restorations. J Korean Dent Assoc 1993;31:438-43.
  12. Binon PP. The effect of implant/abutment hexagonal misfit on screw joint stability. Int J Prosthodont 1996;9:149-60.
  13. Binon PP. Implants and components: entering the new millennium. Int J Oral Maxillofac Implants 2000;15:76-94.
  14. Misch CE. Principles for screw-retained prostheses. In: Misch CE. Contemporary implant dentistry. 2nd ed., Missouri: CV Mosby, 1999. p. 669-85.
  15. Ivanoff CJ, Sennerby L, Lekholm U. Reintegration of mobilized titanium implants. An experimental study in rabbit tibia. Int J Oral Maxillofac Surg 1997;26:310-5. https://doi.org/10.1016/S0901-5027(97)80878-8
  16. Ko SM, Park SJ, Kim IS, Song SI, Lee IK. Observation of the change of the dental implant stability and bone density evaluation methods. J Korean Acad Periodontol 2009;39:185-92. https://doi.org/10.5051/jkape.2009.39.2.185
  17. Park YS, Yi KY, Lee IS, Jung YC. Correlation between microtomography and histomorphometry for assessment of implant osseointegration. Clin Oral Implants Res 2005;16:156-60. https://doi.org/10.1111/j.1600-0501.2004.01083.x
  18. Sennerby L, Dasmah A, Larsson B, Iverhed M. Bone tissue responses to surface-modified zirconia implants: A histomorphometric and removal torque study in the rabbit. Clin Implant Dent Relat Res 2005;7:S13-20. https://doi.org/10.1111/j.1708-8208.2005.tb00070.x
  19. Klokkevold PR, Johnson P, Dadgostari S, Caputo A, Davies JE, Nishimura RD. Early endosseous integration enhanced by dual acid etching of titanium: a torque removal study in the rabbit. Clin Oral Implants Res 2001;12:350-7. https://doi.org/10.1034/j.1600-0501.2001.012004350.x
  20. Westover L, Faulkner G, Hodgetts W, Raboud D. Comparison of implant stability measurement devices for bone-anchored hearing aid systems. J Prosthet Dent 2018;119:178-84. https://doi.org/10.1016/j.prosdent.2017.02.021
  21. Friberg B, Sennerby L, Linden B, Grondahl K, Lekholm U. Stability measurements of one-stage Branemark implants during healing in mandibles. A clinical resonance frequency analysis study. Int J Oral Maxillofac Surg 1999;28:266-72. https://doi.org/10.1016/S0901-5027(99)80156-8
  22. Meredith N, Shagaldi F, Alleyne D, Sennerby L, Cawley P. The application of resonance frequency measurements to study the stability of titanium implants during healing in the rabbit tibia. Clin Oral Implants Res 1997;8:234-43. https://doi.org/10.1034/j.1600-0501.1997.080310.x
  23. Sennerby L, Wennerberg A, Pasop F. A new microtomographic technique for non-invasive evaluation of the bone structure around implants. Clin Oral Implants Res 2001;12:91-4. https://doi.org/10.1034/j.1600-0501.2001.012001091.x
  24. Fanuscu MI, Chang TL. Three-dimensional morphometric analysis of human cadaver bone: microstructural data from maxilla and mandible. Clin Oral Implants Res 2004;15:213-8. https://doi.org/10.1111/j.1600-0501.2004.00969.x
  25. Brandt J, Bierögel C, Holweg K, Hein W, Grellmann W. Extended push-out test to characterize the failure of boneimplant interface. Biomed Tech (Berl) 2005;50:201-6. https://doi.org/10.1515/BMT.2005.028
  26. Dhert WJ, Verheyen CC, Braak LH, de Wijn JR, Klein CP, de Groot K, Rozing PM. A finite element analysis of the pushout test: influence of test conditions. J Biomed Mater Res 1992;26:119-30. https://doi.org/10.1002/jbm.820260111
  27. Cibirka RM, Nelson SK, Lang BR, Rueggeberg FA. Examination of the implant-abutment interface after fatigue testing. J Prosthet Dent 2001;85:268-75. https://doi.org/10.1067/mpr.2001.114266
  28. Elshahawy W, Ajlouni R, Ajlouni K, Sadakah A. Failure torque of ceramic neck titanium implant. J Prosthet Dent 2018;119:432-6. https://doi.org/10.1016/j.prosdent.2017.04.019
  29. Faulkner MG, Wolfaardt JF, Chan A. Measuring abutment/implant joint integrity with the Periotest instrument. Int J Oral Maxillofac Implants 1999;14:681-8.
  30. Haack JE, Sakaguchi RL, Sun T, Coffey JP. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants 1995;10:529-36.
  31. Martin WC, Woody RD, Miller BH, Miller AW. Implant abutment screw rotations and preloads for four different screw materials and surfaces. J Prosthet Dent 2001;86:24-32. https://doi.org/10.1067/mpr.2001.116230
  32. Gross M, Abramovich I, Weiss EI. Microleakage at the abutment-implant interface of osseointegrated implants: a comparative study. Int J Oral Maxillofac Implants 1999;14:94-100.
  33. Kim DG, Park CJ, Cho LR. Comparative analysis of the implant torque controller. J Dent Rehabil Appl Sci 2012;28:27-36.
  34. Jaarda MJ, Razzoog ME, Gratton DG. Providing optimum torque to implant prostheses: a pilot study. Implant Dent 1993;2:50-2. https://doi.org/10.1097/00008505-199304000-00010
  35. Goheen KL, Vermilyea SG, Vossoughi J, Agar JR. Torque generated by handheld screwdrivers and mechanical torquing devices for osseointegrated implants. Int J Oral Maxillofac Implants 1994;9:149-55.
  36. Tan KB, Nicholls JI. The effect of 3 torque delivery systems on gold screw preload at the gold cylinder-abutment screw joint. Int J Oral Maxillofac Implants 2002;17:175-83.
  37. Vallee MC, Conrad HJ, Basu S, Seong WJ. Accuracy of friction-style and spring-style mechanical torque limiting devices for dental implants. J Prosthet Dent 2008;100:86-92. https://doi.org/10.1016/S0022-3913(08)60154-7
  38. Dellinges MA, Tebrock OC. A measurement of torque values obtained with hand-held drivers in a simulated clinical setting. J Prosthodont 1993;2:212-4. https://doi.org/10.1111/j.1532-849X.1993.tb00411.x
  39. Gross M, Kozak D, Laufer BZ, Weiss EI. Manual closing torque in five implant abutment systems: an in vitro comparative study. J Prosthet Dent 1999;81:574-8. https://doi.org/10.1016/S0022-3913(99)70212-X
  40. Jaarda MJ, Razzoog ME, Gratton DG. Providing optimum torque to implant prostheses: a pilot study. Implant Dent 1993;2:50-2. https://doi.org/10.1097/00008505-199304000-00010
  41. Tan KB, Nicholls JI. The effect of 3 torque delivery systems on gold screw preload at the gold cylinder-abutment screw joint. Int J Oral Maxillofac Implants 2002;17:175-83.
  42. Johansson CB. Sennerby L, Albrektsson T. A removal torque and histomorphometric study of bone tissue reaction to commercially pure titanium and vitallium implant. Int J Oral Maxillofac Implants 1991;6:437-41.
  43. Johansson CB, Albrektsson T. A removal torque and histomorphometric study of commercially pure niobium and titanium implants in rabbit bone. Clin Oral Implants Res 1991;2:24-9. https://doi.org/10.1034/j.1600-0501.1991.020103.x
  44. Klokkevold PR, Nishimura RD, Adachi M, Caputo A. Osseointegration enhanced by chemical etching of the titanium surface. A torque removal study in the rabbit. Clin Oral Implants Res 1997;8:442-7. https://doi.org/10.1034/j.1600-0501.1997.080601.x
  45. Buser D, Nydegger T, Hirt HP, Cochran DL, Nolte LP. Removal torque values of titanium implants in the maxilla of miniature pigs. Int J Oral Maxillofac Implants 1998;13:611-9.
  46. Buser D, Nydegger T, Oxland T, Cochran DL, Schenk RK, Hirt HP, Snétivy D, Nolte LP. Interface shear strength of titanium implants with a sandblasted and acid-etched surface: a biomechanical study in the maxilla of miniature pigs. J Biomed Mater Res 1999;45:75-83. https://doi.org/10.1002/(SICI)1097-4636(199905)45:2<75::AID-JBM1>3.0.CO;2-P
  47. Inadome T, Hayashi K, Nakashima Y, Tsumura H, Sugioka Y. Comparison of bone-implant interface shear strength of hydroxyapatite-coated and alumina-coated metal implants. J Biomed Mater Res 1995;29:19-24. https://doi.org/10.1002/jbm.820290104
  48. Ellingsen JE, Johansson CB, Wennerberg A, Holmen A. Improved retention and bone-tolmplant contact with fluoride-modified titanium implants. Int J Oral Maxillofac Implants 2004;19:659-66.
  49. Paepoemsin T, Reichart PA, Chaijareenont P, Strietzel FP, Khongkhunthian P. Removal torque evaluation of three different abutment screws for single implant restorations after mechanical cyclic loading. Oral Implantol (Rome) 2016;9:213-21.
  50. Stüker RA, Teixeira ER, Beck JC, da Costa NP. Preload and torque removal evaluation of three different abutment screws for single standing implant restorations. J Appl Oral Sci 2008;16:55-8. https://doi.org/10.1590/S1678-77572008000100011
  51. Ghanbarzadeh J, Rajabzadeh M, Nakhaei M, Rajatihaghi H, Tayarani Najjaran N. Torque removal evaluation of screw in one-piece and two-piece abutments tightened with a handheld screwdriver. J Dent Mater Tech 2013;3:11-5.
  52. Saliba FM, Cardoso M, Torres MF, Teixeira AC, Lourenço EJ, Telles Dde M. A rationale method for evaluating unscrewing torque values of prosthetic screws in dental implants. J Appl Oral Sci 2011;19:63-7. https://doi.org/10.1590/S1678-77572011000100013
  53. Hwang YJ, Cho JH, Lee CH. Investigation of osseointegration according to the healing time after having iatrogenic mobility of implant fixtures. J Korean Acad Prosthodont 2010;48:308-14. https://doi.org/10.4047/jkap.2010.48.4.308
  54. Koh YH, Kim YJ, Chung HJ. Bone healing around screwshaped titanium implants with three different surface topographies. J Korean Acad Periodontol 2001;31:41-55. https://doi.org/10.5051/jkape.2001.31.1.41
  55. Yi YA, Park YB, Choi H, Kim SJ, Kim KM, Oh S, Shim JS. The evaluation of osseointegration of dental implant surface with different size of TiO 2 nanotube in rats. J Nanomater 2015;2:1-10.
  56. Buser D, Nydegger T, Hirt HP, Cochran DL, Nolte LP. Removal torque values of titanium implants in the maxilla of miniature pigs. Int J Oral Maxillofac Implants 1998;13:611-9.
  57. Albayrak H, Gumus HO, Tursun F, Kocaagaoglu HH, Kilinc HI. Accuracy of torque-limiting devices: A comparative evaluation. J Prosthet Dent 2017;117:81-6. https://doi.org/10.1016/j.prosdent.2016.07.005
  58. Wadhwani CPK, O'Brien R, Rosen PS, Chung KH. Testing and calibrating the mechanical-type toggle torque wrenches used in implant dentistry: A dental technique. J Prosthet Dent 2019 Aug 2. pii: S0022-3913(19)30342-7.
  59. Kreissl ME, Gerds T, Muche R, Heydecke G, Strub JR. Technical complications of implant-supported fixed partial dentures in partially edentulous cases after an average observation period of 5 years. Clin Oral Implants Res 2007;18:720-6. https://doi.org/10.1111/j.1600-0501.2007.01414.x
  60. Sananez A, Lefebvre C, Looney S, Baker P, Mettenburg D, Rueggeberg FA. In vitro mechanical analysis of complete-arch mandibular implant-supported fixed prostheses abutment screws after cyclic loading. J Prosthet Dent 2015;113:432-9. https://doi.org/10.1016/j.prosdent.2014.09.026

Cited by

  1. Influence of Orthodontic Anchor Screw Anchorage Method on the Stability of Artificial Bone: An In Vitro Study vol.13, pp.14, 2019, https://doi.org/10.3390/ma13143205