DOI QR코드

DOI QR Code

치과용 이동형 방사선장치의 선량평가

Evaluation of Radiation Doses of Dental Portable Equipment

  • 투고 : 2018.10.09
  • 심사 : 2018.10.22
  • 발행 : 2018.10.31

초록

We aim to evaluate safety of radiation by measuring leakage dose and patient(phantom) incident dose of ZEN-PX II dental portable equipment developed by G company. Measurement for leakage dose of equipment is conducted on the top, at the bottom, on the left, on the right and at the back. Dose measurement incident on the subject with the area dosimeter when using the phantom and measurement the leakage dose of equipment when using the phantom are evaluated. Comparing the right with the highest leakage dose as a 0 cm, 25 cm, 50 cm, 75 cm and 100 cm dose measurement at the measurement height of 100 cm, 64.2 uR was reduced to 47.3 uR in the senser mode 0.32sec. Even in film mode it was measured at 414.4 uR and about 27% lower at 162.6 uR. As the result of this study, when the irradiation time is 2 sec the right side dose is 290.5 uR and sensor mode is 0.32 sec the right side dose is 64.2 uR.

키워드

참고문헌

  1. Shin KS, Kim YH, et al. The Actual State and the Utilization for Dental Radiography in Korea. Journal of Radiological Science and Technology. 2010;33(2):109-119.
  2. Kenneth PH, Stanley SJ, Mackernzie AJ. Radiation safty for the $NOMAD^{TM}$ portable X-ray system in a temporary morgue setting. J Forensic Sci. 2008;53:917-921. https://doi.org/10.1111/j.1556-4029.2008.00766.x
  3. Kim EK. Leakage and scattered radiation from hand-held dental x-ray unit. Korean J Oral Maxillofac Radiol. 2007;37:65-68.
  4. Kim EK. Radiation safety of portable dental X-ray systems, KFDA Research Report. 2011.
  5. KFDA. Radiation Safety of Portable Dental X-ray System, Radiation safety management series No.37, 2014.
  6. Schueler BA. Operator shielding: how and why. Tech Vasc Interv Radiol. 2010;13:167-171. https://doi.org/10.1053/j.tvir.2010.03.005
  7. Le Heron J, Padovani R, Smith I, Czarwinski R. Radiation protection of medical staff. Eur J Radiol. 2010;76:20-23. https://doi.org/10.1016/j.ejrad.2010.06.034
  8. Medical Radiation Protection in Japan, Chapter 4 of the Medical Law Enforcement Rules.2011.
  9. Ko JK, Park MH, Kim YM. Analysis of the Spatial Dose Rates during Dental Panoramic Radiography. Journal of Radiological Science and Technology. 2016:39(4):509-516. https://doi.org/10.17946/JRST.2016.39.4.04
  10. Jun YL, Cho PK, Han EK, et al. The Knowledge, Attitude and Behavior on the Radiation Safety Management for Dental Hygiene Major Students. Journal of Radiological Science and Technology. 2015;38(4):411-420. https://doi.org/10.17946/JRST.2015.38.4.11