There was a shortage of research reports on sterilization criterion and contamination of ultrasonic probes. Therefore, in this study, we were going to provide a basic study to measure the level of microbial contamination in ultrasonic probes and to investigate the radiographer's awareness of infection. After the scan, samples were collected from the rubber part of the probe by opening a sterile swab (Transport Medium AM608-1S) for medical bacteria collection with the remaining gel removed with a paper towel. Also, the collected samples of bacteria were grown for seven days and then the laboratory was analyzed. Among the total 29 types of microorganisms, Micrococcus luteus 21(26%), Moraxella species 16(20%), Coagulase negative staphylococcus 8(10%), Bacillus species 5(7%), Bicillus circulans 3(5%), Acinetobacter lwoffii 2(2%), and 1 other Candida parapsilosis (1%) a number of bacteria and fungus, was detected. In a disinfectant experiment using LuciPac Pen on the Lumitester PD-30s, we cultured the rubber part of the probe two to three times to measure the bacteria. Bacteria decreased to 97% with Aquanax (alkaline reduced water 100%), 99% with Klarion wash (0.01% sodium hydroxide), 94% with Klarion disinfection (0.01% nitrous acid water), Sterilization was best with Klarion wash (0.01% sodium hydroxide). Therefore, guidelines for cleaning and disinfection of ultrasonic probes was required, and further development of probe-only disinfectants is required.

Keywords

References

  1. Trick WE, Vernon MO, Hayes RA, Nathan C, Rice TW, Peterson BJ, et al. Impact of ring wearing on hand contamination and comparison of hand hygiene agents in a hospital. Clinical infectious diseases. 2003;36(11):1383-90. https://doi.org/10.1086/374852
  2. Foxman B. Epidemiology of urinary tract infections: incidence, morbidity, and economic costs. The American journal of medicine. 2002;113(1):5-13. https://doi.org/10.1016/S0002-9343(02)01054-9
  3. Wagenlehner FME, Naber KG. Antibiotics and resistance of uropathogens. EAU Update Series. 2004;2(3):125-35. https://doi.org/10.1016/j.euus.2004.06.003
  4. Jeong SY, Choi JH, Kim EK, Kim SM, Son HJ, Cho NH, et al. Actual disinfection and sterilization control in Korean healthcare facilities. J korean acad fundam Nurs. 2014;21(4):392-402. https://doi.org/10.7739/jkafn.2014.21.4.392
  5. Lee CB, Lee YS, Lee WH, Cho CC, Yoon HY, Lee YM, et al. Investigation into the Actual State of Sanitary Management and Recognition Degree and Infection Level of Ultrasonographic Probes. Journal of radiological science and technology. 2004;27:51-8.
  6. Abdullah BJJ, Yusof MM, Khoo BH. Physical methods of reducing the transmission of nosocomial infections via ultrasound and probe. Clinical radiology. 1998;53(3):212-4. https://doi.org/10.1016/S0009-9260(98)80103-7
  7. Kim SC. Bacteriological monitoring of radiology room apparatus in the department of radiological technology and contamination on hands of radiological technologists. Journal of the Korean Society of Radiological Technology. 2008;31(4):329-35.
  8. Shin SG, Lee HY. Bacteriological research for the contamination of digital portable radiography. Journal of Radiological Science and Technology. 2017;40(1):7-12. https://doi.org/10.17946/JRST.2017.40.1.02
  9. Choi SG, Song WH, Kweon DC. Bacteriological research for the contamination of equipment in chest radiography. Journal of Radiological Science and Technology. 2015;38(4):395-401. https://doi.org/10.17946/JRST.2015.38.4.09
  10. Yoon J, Kim HJ. A study on effective disinfection methods of medical ultrasound probe resident floras. Journal of the Korea academia-industrial cooperation society. 2018;19(1):346-54. https://doi.org/10.5762/KAIS.2018.19.1.346
  11. Weist K, Wendt C, Petersen LR, Versmold H, RRden H. An outbreak of pyodermas among neonates caused by ultrasound gel contaminated with methicillin-susceptible Staphylococcus aureus. Infection Control & Hospital Epidemiology. 2000;21(12):761-4. https://doi.org/10.1086/501729
  12. Gaillot O, MaruRjouls C, Abachin E, Lecuru F, Arlet G, Simonet M, Berche P. Nosocomial outbreak of Klebsiella pneumoniae producing SHV-5 extended-spectrum ${\beta}$-lactamase, originating from a contaminated ultrasonography coupling gel. Journal of clinical microbiology. 1998;36(5):1357-60.
  13. Hayashi S, Koibuchi H, Taniguchi N, Hirai Y. Evaluation of procedures for decontaminating ultrasound probes. Journal of Medical Ultrasonics. 2012;39(1):11-4. https://doi.org/10.1007/s10396-011-0332-9
  14. Cogen AL, Nizet V, Gallo RL. Skin microbiota: a source of disease or defence?. British Journal of Dermatology. 2008;158(3):442-55. https://doi.org/10.1111/j.1365-2133.2008.08437.x
  15. Bailey EM, Constance TD, Albrecht LM, Rybak MJ. Coagulase-negative staphylococci: incidence, pathogenicity, and treatment in the 1990s. DICP. 1990;24(7-8):714-20. https://doi.org/10.1177/106002809002400715
  16. Dalal A, Urban C, Segal-Maurer S. Endocarditis due to Corynebacterium amycolatum. Journal of medical microbiology. 2008;57(10):1299-302. https://doi.org/10.1099/jmm.0.2008/003343-0
  17. Taiwo SS. Methicillin resistance in Staphylococcus aureus: a review of the molecular epidemiology, clinical significance and laboratory detection methods. West African journal of medicine. 2009;28(5):281-90.
  18. Koibuchi H, Kotani K, Taniguchi N. Ultrasound probes as a possible vector of bacterial transmission. Medical ultrasonography. 2013;15(1):41-4. https://doi.org/10.11152/mu.2013.2066.151.hk1upp2
  19. Koibuchi H, Fujii Y, Kotani K, Konno K, Matsunaga H, Miyamoto M, Taniguchi N. Degradation of ultrasound probes caused by disinfection with alcohol. Journal of Medical Ultrasonics. 2011;38(2):97-100. https://doi.org/10.1007/s10396-010-0296-1