Abstract

In order to broaden the indication of middle ear implant, research has been actively conducted on the reverse output method that stimulates the round window. However, it is very difficult to transmit the vibration output effectively because the indivisual anatomical difference of the round window niche is very large and also the visual field is not secured even by a skilled otolaryngologic surgeon. In this paper, we propose a new reverse stimulation method of middle ear implants that transmits energy to the inner ear by using air as a medium. This can compensate for the disadvantages of the conventional method of transmitting vibration energy and minimizes the energy transfer efficiency interference due to the combination of the excitation point and the output device. It was shown that forward and backward transfer characteristics were obtained by cadaveric experiments, and it was shown that it can overcome the acoustical impedance of high round window and transmit energy to inner ear. The receiver, which is the output device of the conventional hearing aids, can generate a constant volume velocity, so it can have a high output at a limited volume, such as a round window niche. So, suggested method can overcome the high acoustical impedance of the round window and deliver acoustic energy to the inner ear.

Keywords

References

  1. J.H. Cho, S.T. Woo, H.G. Lim, E.S. Jung, J.H. Lee, K.W. Seong, et al., "Proposition of a New Implantable Acoustic Sensor Based on Technology Evaluation of Fully Implantable Hearing Aids," Journal of Sensor Science and Technology, Vol. 23, No. 3, pp. 178-184, 2014. https://doi.org/10.5369/JSST.2014.23.3.178
  2. D.S. Haynes, J.A. Young, G.B. Wanna, and M.E. Glasscock III, "Middle Ear Implantable Hearing Devices: An Overview," Trends Amplification, Vol. 13, No. 3, pp. 206-214, 2009. https://doi.org/10.1177/1084713809346262
  3. H.J. Ahn, H.G Lim, M.N. Kim, and J.H Cho, "Design of External Coil System for Reducing Artifact of MR Image due to Implantable Aid," Journal of Korea Multimedia Society, Vol. 19, No. 2, pp. 375-385, 2016. https://doi.org/10.9717/kmms.2016.19.2.375
  4. T.J. Stewart and A. Belal, "Surgical Anatomy and Pathology of the Round Window," Clinical Otolaryngology, Vol. 6, No. 1, pp. 45-62, 1981. https://doi.org/10.1111/j.1365-2273.1981.tb01785.x
  5. Med-El GmbH, http://medel.com (accessed Jan., 22, 2018).
  6. Envoy Medical Co., http://esteemhearing.com (accessed Jan., 22, 2018).
  7. Cochlear Ltd., http://www.cochlear.com (accessed Jan., 22, 2018).
  8. V. Colletti, S.D. Soli, M. Carner, and L. Colletti, "Treatment of Mixed Hearing Losses via Implantation of a Vibratory Transducer on the RW," International Journal of Audiology, Vol. 45, No. 10, pp. 600-608, 2006. https://doi.org/10.1080/14992020600840903
  9. J.H. Spindel, P.R. Lambert, and R.A. Ruth, "The Basis for a Round Window Electromagnetic Implantable Hearing Aid," Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Vol. 13, No. 4, pp. 1891-1892, 1991.
  10. C.S. Stieger, J.J. Rosowski, and H.H. Nakajima, "Comparison of Forward (ear-canal) and Reverse (round-window) Sound Stimulation of the Cochlea," Hearing Research, Vol. 301, pp. 105-114, 2013. https://doi.org/10.1016/j.heares.2012.11.005
  11. Y. Shimizu, S. Puria, and R.L. Goode, "The Floating Mass Transducer on the Round Window Versus Attachment to an Ossicular Replacement Prosthesis," Otology and Neurotology, Vol. 32, No. 1, pp. 98-103, 2011. https://doi.org/10.1097/MAO.0b013e3181f7ad76
  12. S. Iwasaki, H. Suzuki, H, Moteki, M. Miyagawa, Y. Takumi, and S. Usami, "Experience with the Vibrant Soundbridge RW-Coupler for Round Window Vibroplasty with Tympanosclerosis," Acta Oto-Laryngology, Vol. 132, No. 6, pp. 676-682, 2012. https://doi.org/10.3109/00016489.2011.649492
  13. M. Toth, A. Alpar, L. Paronay, and I. Olah, "Development and Surgical Anatomy of the RW Niche," Annals of Anatomy, Vol. 188, No. 2, pp. 93-101, 2005. https://doi.org/10.1016/j.aanat.2005.09.006
  14. I. Todt, G. Rademacher, J. Wagner, P. Mittmann, D. Basta, and A. Ernst, "Radiological Control of the Floating Mass Transducer Attached to the Round Window," The Scientific World Journal, Vol. 2103, article ID 902945, 2013.
  15. K. O'Connor and S. Purial, "Middle Ear Cavity and Ear Canal Pressure-driven Stapes Velocity Responses in Human Cadaveric Temporal Bones," Journal of Acoustical Society of America, Vol. 129, No. 3, pp. 1517-1528, 2006.
  16. Standard Practice for Describing System Output of Implantable Middle Ear Hearing Devices, http://www.astm.org/Standards/F2504.htm (accessed Jan., 22, 2018).