Development of Indirect Dosimetry by Calculation Method in the Diagnostic X-ray Equipment

진단용엑스선촬영장치의 간접 선량 계산법 개발

  • Kim, Jung-Su (Depart. of Radiologic Technology, Chungbuk Health & Science University) ;
  • Kim, Sung-Hwan (Depart. of Radiological Science, College of Health Science, Cheongju University) ;
  • Jeon, Min-Cheol (Depart. of Radiological Technology, Daejeon Health Institute of Technology) ;
  • Ju, Won-Ha (Depart. of Radiology, Yeouido St. Mary's Hospital of The Catholic University) ;
  • Jeong, Min-Gyu (Depart. Of Radiology, ChungBuk National University Hospital) ;
  • Kim, Mi-Jeong (Ministry of Food and Drug Safety, Medical Device Research Division) ;
  • Lee, Seung-Youl (Ministry of Food and Drug Safety, Medical Device Research Division) ;
  • Lee, Tae-Hee (Ministry of Food and Drug Safety, Medical Device Research Division) ;
  • Seoung, Youl-Hun (Depart. of Radiological Science, College of Health Science, Cheongju University)
  • 김정수 (충북보건과학대학교 방사선과) ;
  • 김성환 (청주대학교 보건의료과학대학 방사선학과) ;
  • 전민철 (대전보건대학교 방사선과) ;
  • 주원하 (가톨릭대학교 여의도성모병원 영상의학과) ;
  • 정민규 (충북대학교병원 영상의학과) ;
  • 김미정 (식품의약품안전처 의료기기연구과) ;
  • 이승열 (식품의약품안전처 의료기기연구과) ;
  • 이태희 (식품의약품안전처 의료기기연구과) ;
  • 성열훈 (청주대학교 보건의료과학대학 방사선학과)
  • Received : 2018.10.29
  • Accepted : 2018.12.17
  • Published : 2018.12.31

Abstract

The purpose of this study was to develop the indirect dosimetry by calculation (IDC) method for diagnostic X-ray equipment. The experiments were performed with inverter type X-ray tubes: Toshiba (E7252X, Japan) and Varian (RAD-14, USA). For the development method, we first applied the standard quality of X-ray beam shown in the TRS457 document, and second, to produce the constants of trendline for the IDC, the total filtration on X-ray beam was subdivided. Third, in order to increase the precision, the energy region was divided into the high energy region and the low energy region and developed by the IDC. In order to verify the IDC, mean dose (mR) values were measured for three Toshiba X-ray tubes and three Varian X-ray tubes at clinical medical institutions and then compared with the IDC on the 2013. As a result, compared with the previous study, the accuracy of the IDC of this study were improved by 2.71% and 9.91% in Toshiba and Varian X-ray tubes, respectively.

Keywords

References

  1. Kimme-Smith C, Aberle DR, Sayre JW, Hart EM, Greaves SM, Brown K, et al. Effects of reduced exposure on computed radiography: comparison of nodule detection accuracy with conventional and asymmetric screen-film radiographs of a chest phantom. American Journal of Roentgenology, 1995;165(2):269-73. https://doi.org/10.2214/ajr.165.2.7618538
  2. Al Khalifah K, Brindhaban A. Comparison between conventional radiography and digital radiography for various kVp and mAs settings using a pelvic phantom. Radiography. 2004;10(2):119-25. https://doi.org/10.1016/j.radi.2004.02.006
  3. Van Metter R, Yorkston J. Applying a proposed definition for receptor dose to digital projection images. In: Medical imaging 2006: physics of medical imaging. San Diego, CA: SPIE; 2006, 6142.
  4. Seibert JA, Shelton DK, Moore EH. Computed radiography x-ray exposure trends. Acad Radiol, 1996;4:313-8.
  5. IEC 60601-2-54:2009/AMD1:2015, Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy; 2015.
  6. Hanan FA. Assessment of Entrance skin dose and effective dose of some routine Examinations using calculation technique. King Saud University; 2007.
  7. Suchart K, Montree T. An Estimation of X-Radiation Output using Mathematic Model. American Journal of Applied Sciences. 2011;8(9):923-6. https://doi.org/10.3844/ajassp.2011.923.926
  8. Sung DW. National Based Research of Patient Radiation National Based Research of Patient Radiation. The Report of Korea Food & Drug Administration; 2013.
  9. IEC 61267:2005, Medical diagnostic X-ray equipment - Radiations conditions for use in the determination of characteristics, Geneva.
  10. IEC 61674:2012, Medical electrical equipment - Dosimeters with ionization chambers and/or semiconductor detectors as used in X-ray diagnostic imaging. 2012. Oct. 24. 2018. Accessed at https://webstore.iec.ch/publication/5711
  11. Seoung YH. Evaluation of Accuracy and Reproducibility of Clinical General X-ray System by Safety Management Standard of Diagnostic Radiation Generators. J. Health. Med. Sci., Cheongju Univ. 2018;6(2):41-8.
  12. Duong VT, Ho QT, Bui DK. Establishing the standard X-ray beam qualities for calibration of dosimeters used in diagnostic radiology following IAEA-TRS457. No. VINATOM-AR-13. 2014.
  13. Suchart K, Montree T. An Estimation of X-Radiation Output using Mathematic Model. American Journal of Applied Sciences. 2011;8(9):923-6. https://doi.org/10.3844/ajassp.2011.923.926
  14. Roh YH, Kim JM, The Tendency of Medical Electrical Equipment - IEC 60601-2-54: Particular Requirements for the Basic Safety and Essential Performance of X-ray Equipment for Radiography and Radioscopy. J. Radiological Science & Technology. 2015;38(4):331-6. https://doi.org/10.17946/JRST.2015.38.4.01
  15. Technical White Paper: Monitoring and Tracking of Fluoroscopic Dose, CRCPD Publication #E-10-7, 2010.
  16. Mori T, et al. The Calculation and Measurement of an Entrances Surface Dose in Diagnostic X-ray. Committee of Radiation Protection (NDD group) in the Ibaragi branch of Japanese Society of Radiological Technology; 1984.
  17. Edmonds IR. Calculation of patient skin dose from diagnostic X-ray procedures. The British journal of radiology. 1984;57(680):733-4. https://doi.org/10.1259/0007-1285-57-680-733
  18. Tung CJ, Tsai HY. Evaluations of gonad and fetal doses for diagnostic radiology. Proc Nat Sci Counc Repub China B. 1999;23(3):107-13.
  19. Suzuki M, et al. The Expression for an Entrances Surface Dose of Patients in Diagnostic X-ray-NDD Method. Committee of Radiation Protection in the Ibaragi branch of Japanese Society of Radiological Technology; 1996.
  20. Kim SC, Kim CY, Ahn SM. Calculation Method of Entrance Skin Dose in X-ray Beam Quality Factor. Jour. of KoCon. 2010;10(2):258-67.
  21. Sung HJ, Han JB, Song JN, Choi NG. Entrance Surface Dose according to Dose Calculation: Head and Wrist. J. Radiological Science & Technology. 2016;39(3):305-12. https://doi.org/10.17946/JRST.2016.39.3.02
  22. Brosi P, Stuessi A, Verdun FR, et al. Copper filtration in pediatric digital X-ray imaging: its impact on image quality and dose. Radiol Phys Technol. 2011;4:148-55. https://doi.org/10.1007/s12194-011-0115-4
  23. Han JB, Choi NG, Sung HJ. Comparative Study of Radiation Exposure using Entrance Skin Dose Calculation Technique in Diagnostic X-Ray Radiography. Jour. of KoCon. 2011;11(12):357-63.