Abstract
This study is to suggest a method to evaluate the detector performance using change of absorbed dose and histogram according to sensitivity change of Auto Exposure Control(AEC). The experiment site is skull, abdomen pelvis and the accuracy of the detector was evaluated by measuring the absorbed dose of the detector sensitivity S200, S400, S800, S1000. Also the dynamic range of the detector was evaluated through the histogram analysis. As a result, the absorbed dose decreased gradually as the sensitivity was set higher from S200 to S1000. And through the sensitivity histogram analysis, as the sensitivity of the skull is set higher, the amount of information at both ends of the histogram is lost. Abdomen and pelvis areas showed underflow phenomena in which the amount of information in the first part of the histogram was lost as the sensitivity was set higher. In conclusion, the detector accurately implemented the sensitivity change, but the dynamic range of the image due to the sensitivity change of the AEC due to the deterioration of the detector performance can not be realized properly and it was found that the evaluation through the absorbed dose and the histogram is useful when evaluating the performance of the detector.
본 연구는 자동노출제어장치의 감도변화에 따른 흡수선량의 변화와 히스토그램을 이용하여 검출기 성능평가 방법을 제안하고자 하였다. 실험방법은 두개부, 복부, 골반부위 별로 검출기 감도 S200, S400, S800, S1000의 흡수선량을 측정하여 검출기의 정확성을 평가하였다. 또한 영상의 히스토그램 분석을 통하여 검출기의 감도별 동적범위를 평가하였다. 그 결과 부위 별 흡수선량은 감도가 S200에서 S1000으로 높게 설정될수록 흡수선량이 단계적으로 감소하였다. 감도 별 히스토그램을 분석한 결과 두개부는 감도가 높게 설정될수록 히스토그램의 양단의 정보량이 상실되는 결과를 보였다. 복부와 골반부는 감도가 높게 설정될수록 히스토그램의 첫부분의 정보량이 상실되는 언더플로우 현상을 보였다. 결론적으로 검출기가 감도의 변화는 정확하게 구현해 내었으나 검출기 성능저하로 인한 자동노출제어장치의 감도변화에 따른 영상의 동적범위는 제대로 구현해내지 못한다는 것을 알 수 있었고 검출기의 성능평가 시 흡수선량과 히스토그램을 통한 평가가 유용하다는 것을 알 수 있었다.