Patient dose in adult and pediatric dental panoramic radiography in Korea

성인과 소아에서 치과 파노라마방사선검사의 환자선량

  • Kim, Eun-Kyung (Dept. of Oral and Maxillofacial Radiology, Dankook University College of Dentistry) ;
  • Han, Won-Jeong (Dept. of Oral and Maxillofacial Radiology, Dankook University College of Dentistry) ;
  • Kim, Kyoung-A (Dept. of Oral and Maxillofacial Radiology, School of Dentistry, Chonbuk National University) ;
  • Lee, Wan (Dept. of Oral and Maxillofacial Radiology, College of Dentistry, Won-Kwang University) ;
  • Yoon, Suk-Ja (Dept. of Oral and Maxillofacial Radiology, School of Dentistry, Chonnam National University) ;
  • Hwang, Eui-Hwan (Dept. of Oral and Maxillofacial Radiology, School of Dentistry, Kyung Hee University) ;
  • Kim, Gyu-Tae (Dept. of Oral and Maxillofacial Radiology, School of Dentistry, Kyung Hee University) ;
  • Heo, Min-Suk (Dept. of Oral and Maxillofacial Radiology, School of Dentistry, Seoul National University) ;
  • An, Chang-Hyeon (Dept. of Oral and Maxillofacial Radiology, School of Dentistry, Kyungpook National University) ;
  • An, Seo-Young (Dept. of Oral and Maxillofacial Radiology, School of Dentistry, Kyungpook National University) ;
  • Han, Jin-Woo (Dept. of Oral and Maxillofacial Radiology, College of Dentistry, Gangneung-Wonju National University) ;
  • Jung, Yun-Hoa (Dept. of Oral and Maxillofacial Radiology, School of Dentistry, Pusan National University)
  • 김은경 (단국대학교 치과대학 영상치의학교실) ;
  • 한원정 (단국대학교 치과대학 영상치의학교실) ;
  • 김경아 (전북대학교 치의학전문대학원 영상치의학교실) ;
  • 이완 (원광대학교 치과대학 영상치의학교실) ;
  • 윤숙자 (전남대학교 치의학전문대학원 구강악안면방사선학교실) ;
  • 황의환 (경희대학교 치의학전문대학원 영상치의학교실) ;
  • 김규태 (경희대학교 치의학전문대학원 영상치의학교실) ;
  • 허민석 (서울대학교 치의학대학원 영상치의학교실) ;
  • 안창현 (경북대학교 치의학전문대학원 영상치의학교실) ;
  • 안서영 (경북대학교 치의학전문대학원 영상치의학교실) ;
  • 한진우 (강릉원주대학교 치과대학 구강악안면방사선학교실) ;
  • 정연화 (부산대학교 치의학전문대학원 영상치의학교실)
  • Received : 2017.05.11
  • Accepted : 2017.06.22
  • Published : 2017.08.01

Abstract

Objective: To suggest diagnostic reference levels (DRLs) for dental panoramic radiography in adults and children through the nationwide survey in Korea. Materials and Methods: Two hundred twelve dental institutions on a national basis were visited. The radiographic examination protocols were surveyed and their patient doses at the clinical exposure setting for adult and children (5- and 10-year old) were measured at 244 panoramic radiographic equipment. The measured DAP were analyzed and compared according to age group, the size of hospital, the type of radiographic system, the installation duration of equipment. Results: The mean exposure parameters were 70.1 kV, 9.2 mA, 14.4 second for adult and 66.6 kV, 7.9 mA, 13.8 second for 10-year old child and 65.5 kV, 7.3 mA, 13.7 second for 5-year old child. The mean and third quartile patient DAPs were $138.3mGy\;cm^2$ and $151.0mGy\;cm^2$ for adult, $99.5mGy\;cm^2$ and $104.8mGy\;cm^2$ for 10-year old child and $89.5mGy\;cm^2$ and $95.5mGy\;cm^2$ for 5-year old child. The mean patient dose at the university dental hospital was lower than that at the dental clinic (p<0.05). The mean patient dose of direct digital radiography type was higher than that of film-based type. However, the difference did not show statistically significance. Conclusion: DRLs for dental panoramic radiography in adult, 10- and 5-year old child were suggested to be $151mGy\;cm^2$, $105mGy\;cm^2$, and $96mGy\;cm^2$ in Korea based on this nationwide survey.

Keywords

Acknowledgement

Supported by : 식품의약품안전처

References

  1. White SC, Pharoah MJ. Oral radiology: Principles and interpretation. 7th ed. St. Louis: Elsevier Inc. 2014. p. 29-40.
  2. European Commission. Radiation protection No. 172. Evidence based guidelines on cone beam CT for dental and maxillofacial radiology. [Internet]. Luxemburg: 2012 [cited 2017 February 28]. Available from: https://ec.europa.eu/energy/sites/ener/files/documents/172.pdf
  3. ICRP. Diagnostic reference levels in medical imaging: review and additional advice. A web module produced by Committee 3 of the International Commission on Radiological Protection. ICRP; 2001 [cited 2017 February 28], available from http://www.icrp.org/docs/DRL_for_web.pdf
  4. ICRP. The 2007 recommendations of the International Commission on Radiological Protection. ICRP Publication 103. Ann ICRP 2007; 37(2-4).
  5. European Commission. Radiation protection No. 109. Guidance on Diagnostic Reference Levels (DRLs) for Medical Exposures. [Internet]. Luxemburg: 1999 [cited 2017 February 28]. Available from: https://ec.europa.eu/energy/sites/ener/files/documents/109_en.pdf
  6. Hart D, Hillier MC, Wall BF. HPA-RPD-029. Doses to patients from radiographic and fluoroscopi x-ray imaging procedures in the UK - 2005 review. [Internet]. Chilton: HPARPD; 2007 [cited 2017 February 28]. Available from https://www.gov.uk/government/uploads/system/uploads/attachment_data/file/431134/HPA-RPD-029.pdf
  7. Hart D, Hillier MC, Shrimpton PC. HPA-CRCE-034. Dose to patients from radiographic and fluoroscopic x-ray imaging procedures in the UK - 2010 review. [Internet]. Chilton: HPARPD 2012. [cited 2017 February 28]. Available from https://www.gov.uk/government/uploads/system/uploads/attachment_data/file/342780/HPA-CRCE-034_Doses_to_patients_from_radiographic_and_fluoroscopic_x_ray_imaging_procedures_2010.pdf
  8. Poppe B, Looe HK, Pfaffenberger A, Chofor N, Eenboom F, Sering M, Ruhmann A, Poplawski A, Willborn K. Dose-area product measurements in panoramic dental radiology. Radiat Prot Dosimetry 2007; 123: 131-134. https://doi.org/10.1093/rpd/ncl090
  9. Tierris CE, Yakoumakis EN, Bramis GN, Georgiou E. Dose area product reference levels in dental panoramic radiology. Radiat Prot Dosimetry 2004;111: 283-287. https://doi.org/10.1093/rpd/nch341
  10. Kim EK. Development of diagnostic reference level in dental x-ray examination in Korea. Final report. Chungcheongbuk-do: Korean Ministry of Food and Drug Safety. 2009 Nov. Report no. 09142Radiology510.
  11. Holroyd JR. HPA-CRCE-043. Trends in dental radiography equipment and patient dose in the UK and Republic of Ireland. [Internet]. Chilton: HPA CRCE; 2013 [cited 2017 February 28]. Available from https://www.gov.uk/government/uploads/system/uploads/attachment_data/file/340127/HPA-CRCE-043_for_website.pdf
  12. FANC. Federaal Agentschap voor Nucleaire Controle. Acceptance criteria dentistry Degree of December 12, 2008 on the acceptance criteria for x-ray machines for diagnostic use in dentistry. Article 43. Equipment, measurement conditions and criteria for the measurement of the dose. [Internet]. Brussels: 2008 [cited 2017 February 28]. Available from http://www.jurion.fanc.fgov.be/jurdb-consult/consultatieLink? wettekstId=7705&appLang=nl&wettekstLang=nl
  13. NCRP Report No. 172. Reference levels and achievable doses in medical and dental imaging: Recommendations for the United States. [Internet]. Bethesda: 2012. [cited 2017 February 28]. Available from http://www.ncrppublications.org/Reports/172