Assessment of Radiation Dose from Radioactive Wedge Filters during High-Energy X-Ray Therapy

  • Back, Geum-mun (Department of Radiation Oncology, Asan Medical Center) ;
  • Park, Sung Ho (Department of Neurosurgery, Ulsan University Hospital) ;
  • Kim, Tae-Hyung (Department of Radiological Science, Kangwon National University)
  • Received : 2017.03.17
  • Accepted : 2017.07.01
  • Published : 2017.06.30

Abstract

This paper evaluated the amount of radiation generated by wedge filters during radiation therapy using a high-energy linear accelerator, and the dose to the worker during wedge replacement. After 10-MV photon beam was irradiated with wedge filter, the wedge was removed from the linear accelerator, and the dose rate and energy spectrum were measured. The initial measurement was approximately 1 uSv/h, and the radiation level was reduced to 0.3 uSv/h after 6 min. The effective half-life derived from the dose rate measurement was approximately 3.5 min, and the influence of AI-28 was about 53%. From the energy spectrum measurements, a peak of 1,799 keV was measured for AI-28, while the peak for Co-58 was not measured in the control room. The peaks for Au-106 and Cd-105 were found only measurement was done without wedge removement from the linear accelerator. The additional doses received by the radiation worker during wedge replacement were estimated to be 0.08-0.4 mSv per year.

Keywords

References

  1. NCRP Report 79. Neutron contamination from medical electron accelerator, 1984.
  2. AAPM TG Report 27. Neutron measurements around high energy X-ray radiotherapy machines. 1986.
  3. Chilbani O, Ma C. Photoneutron dose calculations for high energy photon beams from Siemens and Varian LINACS. Med Phys 1990-2000, 2003;30. https://doi.org/10.1118/1.1590436
  4. Gudowska I, Brahme A, Andreo P, et al. Calcluation of absorbed dose and biological effect iveness from photoneutron reactions in bremsstrahlungbeam of end point 50 MeV. Phys in Med Biol. 1999;44:2099. https://doi.org/10.1088/0031-9155/44/9/301
  5. Roig M, Panettieri V, Ginjaume M, et al. Photonuclear isotope characterization of a Siemens KDS 18 MeV LINAC head. Phys in Med Biol. 2004;49: N242-6.
  6. Borchardt IM, Patterson JR, Beddoe AH. An investigation of photonuclear reactions in Cerrobend eutectic material with an 18 MV LINAC. Phys in Med Biol. 1991;36:649-53. https://doi.org/10.1088/0031-9155/36/5/008
  7. Pet rovic N, K rest ic-Vesovic J, Stojanovic D, et al. Contribution of activation products to occupational exposure following treatment using high energy photons in radiotherapy. Rad Prot Dos. 2011;143:109-12. https://doi.org/10.1093/rpd/ncq290
  8. Fischer HW, Tabot BE, Poppe B. Activation processes in a medical linear accelerator and spatial distribution of activation products. Phys Med Biol. 2006;51:N461. https://doi.org/10.1088/0031-9155/51/24/N02
  9. Fischer HW, Tabot B, Poppe B. Comparison of activation products and induced dose rates in diferent high-energy medical linear accelerators. Health Physics. 2008;94:272-8. https://doi.org/10.1097/01.HP.0000291945.15684.a7
  10. Israngkul-Na-Ayuthaya I, Suriyapee S, Pengvanich P. Evaluation of equivalent dose from neutrons and activation products from a 15 MV x-ray LINAC. J of Rad Research. 2015;56:919-26. https://doi.org/10.1093/jrr/rrv045