DOI QR코드

DOI QR Code

Effects of grooved abutment on stability of implant abutment screw

Grooved abutment가 임플란트 지대주 연결나사의 안정성에 미치는 영향

  • Sim, Il-Gwang (Department of Prosthodontics, The Graduate School, Yonsei University) ;
  • Yang, Seung-Won (Department of Prosthodontics, The Graduate School, Yonsei University) ;
  • Shim, June-Sung (Department of Prosthodontics, Yonsei University College of Dentistry) ;
  • Kim, Jee-Hwan (Department of Prosthodontics, Yonsei University College of Dentistry)
  • 심일광 (연세대학교 대학원 치과보철학교실) ;
  • 양승원 (연세대학교 대학원 치과보철학교실) ;
  • 심준성 (연세대학교 치과대학 치과보철학교실) ;
  • 김지환 (연세대학교 치과대학 치과보철학교실)
  • Received : 2016.06.13
  • Accepted : 2016.09.02
  • Published : 2016.10.31

Abstract

Purpose: The aim of this study was to investigate the effects of grooved abutments on abutment screw loosening. Materials and methods: This study was conducted to evaluate the abutment screw loosening after 6 months for 50 patients (51 implants) treated at the department of Prosthodontics in Yonsei University Dental Hospital from March, 2015 to July, 2015. A control group with non-grooved abutment consists of 30 implants, and an experimental group with grooved abutment consists of 21 implants. Astra, Straumann, Implantium, Osstem system were used in the study. The abutments with loose screws cases after a period of 6 months has been investigated, with two kinds of measurements: 1) measuring the additional rotational angle on abutment during placement with the same force, 2) measuring the PTV on bucco-cervical area of implant crown. All data collected has been analyzed by normality test followed by Mann-Whitney test using SPSS program. Results: No complications were reported after 6 months for the 51 implants. Abutment screw loose and crown fracture have not been seen in the study groups. The data collected from the two measurements showed no significant differences between the two groups with P-value 0.576 (average= control group: $7.35^{\circ}$, experimental group: $4.75^{\circ}$) for the additional rotational angle measurement and with P-value 0.767 for PTV. Conclusion: There are no significant differences between the grooved and non-grooved abutment in screw stability. However, further studies with long-term followups and larger group of patients is needed in order to investigate the effects of grooved abutment on screw stability.

목적: 본 연구의 목적은 grooved abutment가 지대주 나사풀림 현상에 미치는 영향을 확인하는 것이다. 재료 및 방법: 2015년 3월부터 7월까지 연세대학교 치과대학병원 보철과에 내원한 50명, 51개 임플란트에 대해서 임플란트 보철 장착 후 6개월 점검 시 지대주 연결나사의 안정성을 평가하였다. 대조군은 groove가 없는 지대주를 사용한 30개 임플란트이었고, 실험군은 groove가 있는 지대주를 사용한 21개 임플란트이었으며, Astra, Straumann, Implantium, Osstem 시스템이 사용되었다. 6개월 점검 시에 지대주 연결나사의 풀림여부를 조사하였고, 동일한 힘으로 재체결 시 지대주 연결나사의 추가 회전각을 측정하였으며, 보철물 협측하방 부위에서 PTV를 측정하였다. SPSS 프로그램을 이용하여 정규성 분석 후 Mann-Whitney 분석을 시행하였다. 결과: 51개 임플란트에 대해 보철 장착 후 6개월 점검을 시행한 바, 나사 풀림이나 보철물 파절 등의 특이할 만한 합병증은 나타나지 않았다. 6개월 후 지대주 연결나사의 추가회전각은, 실험군 평균은 $4.75^{\circ}$이었고 대조군 평균은 $7.35^{\circ}$이었으며, 통계적으로 유의한 차이는 없었다 (P = .576). PTV에서도 유의한 차이는 없었다 (P = .767). 결론: 지대주의 groove 여부가 지대주 연결나사의 안정성에 미치는 영향은 유의한 차이가 없으나, 이에 대해 좀더 장기적이고 다수의 증례를 이용한 추가적인 연구가 필요해 보인다.

Keywords

References

  1. Wittneben JG, Buser D, Salvi GE, Bargin W, Hicklin S, Bragger U. Complication and failure rates with implant-supported fixed dental prostheses and single crowns: a 10-year retrospective study. Clin Implant Dent Relat Res 2014;16:356-64. https://doi.org/10.1111/cid.12066
  2. Jung RE, Pjetursson BE, Glauser R, Zembic A, Zwahlen M, Lang NP. A systematic review of the 5-year survival and complication rates of implant-supported single crowns. Clin Oral Implants Res 2008;19:119-30. https://doi.org/10.1111/j.1600-0501.2007.01453.x
  3. Dierens M, De Bruyn H, Kisch J, Nilner K, Cosyn J, Vandeweghe S. Prosthetic survival and complication rate of single implant treatment in the periodontally healthy patient after 16 to 22 years of follow-up. Clin Implant Dent Relat Res 2016;18:117-28. https://doi.org/10.1111/cid.12266
  4. Cha HS, Kim YS, Jeon JH, Lee JH. Cumulative survival rate and complication rates of single-tooth implant; focused on the coronal fracture of fixture in the internal connection implant. J Oral Rehabil 2013;40:595-602. https://doi.org/10.1111/joor.12065
  5. Pjetursson BE, Bragger U, Lang NP, Zwahlen M. Comparison of survival and complication rates of tooth-supported fixed dental prostheses (FDPs) and implant-supported FDPs and single crowns (SCs). Clin Oral Implants Res 2007;18:97-113. https://doi.org/10.1111/j.1600-0501.2007.01439.x
  6. Kim ES, Shin SY. Influence of the implant abutment types and the dynamic loading on initial screw loosening. J Adv Prosthodont 2013;5:21-8. https://doi.org/10.4047/jap.2013.5.1.21
  7. Gehrke P, Dhom G, Brunner J, Wolf D, Degidi M, Piattelli A. Zirconium implant abutments: fracture strength and influence of cyclic loading on retaining-screw loosening. Quintessence Int 2006;37:19-26.
  8. Kitagawa T, Tanimoto Y, Odaki M, Nemoto K, Aida M. Influence of implant/abutment joint designs on abutment screw loosening in a dental implant system. J Biomed Mater Res B Appl Biomater 2005;75:457-63.
  9. Kim SK, Lee JB, Koak JY, Heo SJ, Lee KR, Cho LR, Lee SS. An abutment screw loosening study of a Diamond Like Carboncoated CP titanium implant. J Oral Rehabil 2005;32:346-50. https://doi.org/10.1111/j.1365-2842.2004.01475.x
  10. Bakaeen LG, Winkler S, Neff PA. The effect of implant diameter, restoration design, and occlusal table variations on screw loosening of posterior single-tooth implant restorations. J Oral Implantol 2001;27:63-72. https://doi.org/10.1563/1548-1336(2001)027<0063:TEOIDR>2.3.CO;2
  11. Dixon DL, Breeding LC, Sadler JP, McKay ML. Comparison of screw loosening, rotation, and deflection among three implant designs. J Prosthet Dent 1995;74:270-8. https://doi.org/10.1016/S0022-3913(05)80134-9
  12. Siamos G, Winkler S, Boberick KG. Relationship between implant preload and screw loosening on implant-supported prostheses. J Oral Implantol 2002;28:67-73. https://doi.org/10.1563/1548-1336(2002)028<0067:TRBIPA>2.3.CO;2
  13. Jorneus L, Jemt T, Carlsson L. Loads and designs of screw joints for single crowns supported by osseointegrated implants. Int J Oral Maxillofac Implants 1992;7:353-9.
  14. Zhou JF, Gu BQ, Chen Y. An improved design of spiral groove mechanical seal. Chinese J Chem Eng 2007;15:499-506. https://doi.org/10.1016/S1004-9541(07)60115-3
  15. Byrne D, Jacobs S, O'Connell B, Houston F, Claffey N. Preloads generated with repeated tightening in three types of screws used in dental implant assemblies. J Prosthodont 2006;15:164-71. https://doi.org/10.1111/j.1532-849X.2006.00096.x
  16. Coelho PG, Sudack P, Suzuki M, Kurtz KS, Romanos GE, Silva NR. In vitro evaluation of the implant abutment connection sealing capability of different implant systems. J Oral Rehabil 2008;35:917-24. https://doi.org/10.1111/j.1365-2842.2008.01886.x