DOI QR코드

DOI QR Code

Evaluation and Design of Ultrasonic Vibrator for Dental Surgery

치과용 골 수술기의 초음파 진동자 설계 및 평가

  • Park, Ki-Moon (School of Mechanical Engineering, Yeungnam University) ;
  • Kim, Jung-Hyun (School of Mechanical and Automotive Engineering, Kyungil University) ;
  • Ko, Tae Jo (School of Mechanical Engineering, Yeungnam University)
  • 박기문 (영남대학교 기계공학부) ;
  • 김정현 (경일대학교 기계자동차학부) ;
  • 고태조 (영남대학교 기계공학부)
  • Received : 2016.03.03
  • Accepted : 2016.04.18
  • Published : 2016.06.30

Abstract

A dental ultrasonic surgical instrument, commercially known as a scaler, is a high-value-added advanced technology that is used for tartar removal, implant operations, and gum and jaw bone surgery. In this study, the piezoelectric phenomenon for making linear motion associated with input electrical signals was studied, and the behavior of the ultrasonic vibrator was investigated by using the commercially available finite element program ANSYS(R) for the purpose of designing dental surgery tools. Modal analysis was carried out, and the optimal frequency range was calculated from the analyzed results. The ultrasonic vibrator was then redesigned based on the calculated optimal frequency range. The performance of the system was tested, and consequently, the proposed methodology was proven useful in vibrator design.

Keywords

References

  1. Lee, J. H., Oh, J. M., Hong, Y., Kim, S., Paik, J., Lee, Y. J., and Lee, S. D., "Design and Evaluation of Ultrasonic Bone Surgical Instruments for Dental Applicaion," Journal of the Korean Institute of Electrical and Electronic Material Engineers, Vol. 25, No. 12, pp. 990-995, 2012. https://doi.org/10.4313/JKEM.2012.25.12.990
  2. Sa, M. W., Shim, H. R., Ko, T. J., Lee, J. M., Kim, J. Y., "A Study about Experimental Evaluation of an Ultrasonic Surgery Unit for Bone-cutting," Journal of the Korean Society of Manufacturing Process Engineers, Vol. 15, No. 1, pp. 1-7, 2016. https://doi.org/10.14775/ksmpe.2016.15.1.001
  3. Lee, B. G., Kim, K. L., and Kim, K. E., "Design of Ultrasonic Vibration Tool Horn for Micromachining using FEM," Transactions of the Korean Society of Machine Tool Engineers, Vol. 17, No. 6, pp. 63-70, 2008.
  4. Park, M. H., Kim, T. Y. and Kim, M. H., "Design and FEM Analysis of Langevin Type Ultrasonic Vibrator," KSMTE Conference, pp. 525-528, 2000.
  5. Park, K. M. and Ko, T. J., "Design of Ultrasonic Vibrator for Dental Surgery using FEA," Proc. of the 6th International conference on Manufacturing, Machine Design and Tribology, No. 15-204, pp. 86-87, 2015.
  6. Moreno, E., Acevedo, P., Fuentes, M., Sotomayor, A., Borrto, L., Villafuerte, M. E., and Leija, L., "Design and Construction of a Bolt-Clamped Langevin Transducer," International Conference on Electrical and Electronics Engineering, pp. 393-395, 2005.
  7. Li, C. P. and Ko, T. J., "The Effects of Ultrasonic Vibration on Surface Finish and Tool Wear in End-milling Machining," 2012 International Conference on Manipulation, Manufacturing and Measurement on the Nanoscale (3M-NANO), pp. 330-334, 2012.
  8. Zeng, W. M., Li, Z. C., Pei, Z. J., and Treadwell, C., "Experimental observation of tool wear in rotary ultrasonic machining of advanced ceramics," International Journal of Machine Tools and Manufacture, Vol. 45, pp. 1468-1473, 2005. https://doi.org/10.1016/j.ijmachtools.2005.01.031
  9. Kadivar, M. A., Akbari, J., Yousefi, R., Rahi, A., and Nick, M. G., "Investigating the effects of vibration method on ultrasonic-assisted drilling of Al/SiCp metal matrix composites," Robotics and Computer-Integrated Manufacturing, Vol. 30, pp. 344-350, 2014. https://doi.org/10.1016/j.rcim.2013.10.001
  10. Kurniawan, R. and Ko, T. J., "A New Tool Holder Design with two-Dimensional Motion for Fabricating Micro-Dimple and Groove Patterns," International Journal of Precision Engineering and Manufacturing, Vol. 15, No. 6, pp. 1165-1171, 2014. https://doi.org/10.1007/s12541-014-0452-4